- Home
- Categories
- Test equipment
- Meters
- Power
- 8502A - new version
The Giga-tronics 8502A (8500A Series) is a dual-channel peak power meter designed for high-performance peak and CW microwave power measurements. This updated version features a modernized display and improved system processing while maintaining full compatibility with Giga-tronics' extensive line of peak and CW power sensors. It is ideal for analyzing pulsed radar, electronic warfare, and telecommunication signals.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2) |
| Display | Color LCD graphic display |
| Warm-up Time | 30 minutes |
| Saved Setups Capacity | 10 complete instrument setups |
| Self-Test Capability | Automatic power-on self-test of RAM, ROM, and circuitry |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 VAC ±10%, 48 to 440 Hz |
| Operating Temperature | 0 to +50 °C |
| Dimensions (W×H×D) | 426 mm × 133 mm × 457 mm |
| Weight | 13.6 kg |
Measurement Performance
| Parameter | Value |
|---|---|
| Frequency Range | 0.01 to 50 GHz (sensor dependent) |
| Peak Power Measurement Range | -20 to +20 dBm |
| CW Power Measurement Range | -70 to +47 dBm (sensor dependent) |
| Maximum Rise Time | < 15 ns (with 85001 sensor) |
| Maximum Fall Time | < 15 ns (with 85001 sensor) |
| Minimum Pulse Width | 100 ns |
| Instrumentation Linearity | ±0.1 dB (or ±2.3%) |
| Dynamic Range (Peak) | 40 dB |
| Dynamic Range (CW) | 90 dB (sensor dependent) |
Channel and Sensor Specifications
| Parameter | Value |
|---|---|
| Number of Channels | 2 |
| Sensor Compatibility | Giga-tronics 8500 Series Peak Power Sensors & 8000 Series CW Sensors |
| Sensor Input Impedance | 50 ohms (nominal) |
| Sensor Calibration Method | EEPROM-programmed factory calibration factors applied automatically |
Internal Calibration Source
| Parameter | Value |
|---|---|
| Calibrator Frequency | 1.0 GHz |
| Calibrator Power Output Level | 0 dBm (1.0 mW) |
| Calibrator Output Connector Type | Type N (female) |
| Calibrator VSWR | < 1.15 at 1.0 GHz |
Triggering and Timebase
| Parameter | Value |
|---|---|
| Trigger Modes | Internal, External, Free Run |
| Trigger Sources | Channel A, Channel B, External Trigger Input |
| Trigger Level Range | -20 to +20 dBm (internal) |
| External Trigger Input Level | TTL compatible |
| Timebase Range | 10 ns/division to 10 ms/division |
| Timebase Accuracy | ±0.01% |
Display and User Interface
| Parameter | Value |
|---|---|
| Auto-Measurement Parameters | Rise time, fall time, pulse width, pulse period, PRF, duty cycle, delay, peak power, average power |
| Measurement Units | dBm, dB, W, mW, µW, nW, pW |
System Interfaces and Connectivity
| Parameter | Value |
|---|---|
| Analog Recorder Output | Rear panel BNC connector (0 to 10 V scale) |
| External Trigger Input Connector | Rear panel BNC (female) |
Power and Electrical
| Parameter | Value |
|---|---|
| Power Consumption | 120 VA maximum |
Physical and Environmental
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 to +75 °C |
| Relative Humidity | Up to 95% non-condensing |
| Rack Height Unit | 3U (standard 19-inch rack compatible) |
Triggering
| Parameter | Value |
|---|---|
| Trigger Delay Range | 100 ns to 100 ms |
| Trigger Delay Resolution | 10 ns |
Calibration
| Parameter | Value |
|---|---|
| Calibrator Frequency Accuracy | ±1.0% |
| Calibrator Power Accuracy | ±0.06 dB (±1.4%) at 0 dBm |
| Calibration Factor Range | 50.0% to 150.0% |
| Calibration Factor Resolution | 0.1% |
Power Requirements
| Parameter | Value |
|---|---|
| AC Input Voltage Range | 90 to 260 VAC (auto-ranging) |
| AC Input Frequency Range | 47 to 440 Hz |
Performance
| Parameter | Value |
|---|---|
| Display Resolution (Logarithmic) | 0.001 dB to 1 dB (selectable) |
| Display Resolution (Linear) | 1 to 4 digits (selectable) |
| Averaging Range | 1 to 1024 (selectable in binary steps) |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 4,572 m (15,000 ft) |
| Maximum Storage Altitude | 15,240 m (50,000 ft) |
Calibrator
| Parameter | Value |
|---|---|
| Calibrator Source Impedance | 50 Ω |
Inputs & Outputs
| Parameter | Value |
|---|---|
| VGA Output | 15-pin D-sub |
Technology
RF & Microwave Diode Detection
Applications
Pulsed RF signal analysis, radar testing, and CW power measurement