The Keithley 2000 is a high-performance 6.5-digit digital multimeter that combines broad measurement ranges with exceptional speed and accuracy. Featuring a built-in 1024-reading memory buffer and an expansion slot for optional 20-channel multiplexer cards, it is an ideal choice for both precise benchtop troubleshooting and high-throughput automated measurement applications.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232
DisplayVacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V, 50 Hz / 60 Hz / 400 Hz, 22 VA max
Operating Temperature0 to 50
Dimensions (W×H×D)213 x 89 x 370 mm
Weight2.9

General Characteristics

ParameterValue
Maximum Resolution (Digits)6.5
Reading Memory (Buffer Size)1024 readings
Supported Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, 2-Wire Resistance, 4-Wire Resistance, Frequency, Period, Temperature, Diode, Continuity
Warm-up Time1 hour to rated accuracy

DC Voltage

ParameterValue
Measurement Ranges100 mV, 1 V, 10 V, 100 V, 1000 V
Maximum Resolution100 nV
Accuracy (90 Days)0.002% (basic)
Maximum Input Voltage1000 V DC
Input Impedance (100 mV to 10 V ranges)>10 GΩ
Input Impedance (100 V to 1000 V ranges)10 MΩ ±1%
Input Bias Current< 30 pA at 25°C

AC Voltage

ParameterValue
Frequency Range3 Hz to 300 kHz
Measurement TechniqueTrue RMS

DC Current

ParameterValue
Maximum Input Current3 A

Resistance

ParameterValue
Measurement Types (2-Wire / 4-Wire)2-Wire, 4-Wire

Frequency and Period

ParameterValue
Period Range333 ms to 2 µs

Temperature Measurement

ParameterValue
Supported Thermocouple TypesJ, K, T (Requires external CJC or scanner card with CJC)

System Speeds and Measurement Rates

ParameterValue
Maximum Readings per Second (To Memory)2000 rdgs/s (at 4.5 digits)
Maximum Readings per Second (To Interface)50 rdgs/s (at 6.5 digits via IEEE-488)

Math and Triggering

ParameterValue
Mathematical OperationsREL, Min/Max/Average/StdDev, dB, dBm, Limit Test, %, mX+b

Interfaces and Expansion

ParameterValue
Emulation ModesSCPI, Keithley 199, Fluke 8840A, Fluke 8842A

Environmental and Compliance

ParameterValue
Storage Temperature Range-40 °C to 70 °C

Physical Characteristics

ParameterValue
Rack Mount CompatibilityHalf-rack, 2U

Performance

ParameterValue
Common Mode Rejection Ratio (CMRR)140 dB (DC, 1 kΩ unbalance in LO lead)
Normal Mode Rejection Ratio (NMRR)60 dB (for 1 NPLC or greater)

Diode Test

ParameterValue
Diode Test Current1 mA

Continuity

ParameterValue
Continuity ThresholdAdjustable from 1 Ω to 1000 Ω

Power

ParameterValue
Power Consumption22 VA maximum

Environmental

ParameterValue
Operating Humidity80% RH up to 35°C, 70% RH up to 50°C (non-condensing)
Maximum Operating Altitude2000 m
Pollution Degree2

Measurement Configuration

ParameterValue
Maximum Common Mode Voltage500V peak
Integration Time Settings0.01, 0.1, 1, 10 Power Line Cycles (PLC)

Voltage Measurement

ParameterValue
AC Input Impedance1 MΩ ±2% shunted by < 100 pF
Crest Factor (AC)3:1 at full scale

Current Measurement

ParameterValue
Input Protection (Current)3A, 250V fast-blow fuse

Technology

Digital Multimeter (DMM)

Applications

Benchtop & System Test