The Keithley 2002 is an 8.5-digit high-performance digital multimeter engineered for applications demanding exceptional resolution, accuracy, and high throughput. Built around a 28-bit A/D converter, the Model 2002 offers unmatched measurement integrity with a broad suite of built-in functions, an emulation mode for the HP 3458A, and support for an optional internal 10-channel scanner card.

Specifications

General

ParameterValue
Compatible Scanner CardsKeithley 2001-SCAN, 2001-TCSCAN

Physical

ParameterValue
Operating Temperature0°C to 50°C
Rack Mount Height2U
Rack Mount WidthHalf-rack

DC Voltage

ParameterValue
DCV Measurement Ranges200 mV, 2 V, 20 V, 200 V, 1000 V
DCV Maximum Resolution1 nV (on 200 mV range)
DCV Input Resistance>100 GΩ on 200 mV, 2 V, 20 V ranges; 10 MΩ ±1% on 200 V, 1000 V ranges

AC Voltage

ParameterValue
ACV Measurement Ranges200 mV, 2 V, 20 V, 200 V, 750 V
ACV Measurement TechniqueTrue RMS, AC-coupled or AC+DC-coupled
ACV Frequency Range1 Hz to 2 MHz
ACV Maximum Crest Factor5:1 at full scale
ACV Coupling OptionsAC-coupled or AC+DC-coupled

DC Current

ParameterValue
DCI Measurement Ranges200 µA, 2 mA, 20 mA, 200 mA, 2 A

AC Current

ParameterValue
ACI Measurement Ranges200 µA, 2 mA, 20 mA, 200 mA, 2 A

Frequency and Period

ParameterValue
Frequency Range1 Hz to 15 MHz
Period Range67 ns to 1 s

Interfaces and Connectivity

ParameterValue
GPIB InterfaceIEEE-488.2 compatible
SCPI CompatibilityFully compatible

Power and Physical Specifications

ParameterValue
Warm-up Time4 hours to rated accuracy

Environmental and Standards

ParameterValue
Storage Temperature Range-40°C to 70°C

Power

ParameterValue
Line Frequency Range45 Hz to 66 Hz, and 360 Hz to 440 Hz (automatically sensed)

System/General

ParameterValue
Maximum Guard to Earth Voltage500 V peak
Maximum LO to Earth Voltage500 V peak

Technology

Analog-to-digital integration

Applications

High-precision laboratory calibration, component testing, and automated test equipment (ATE) environments