The Keithley 2182A Nanovoltmeter is a high-performance, two-channel instrument optimized for stable, low-noise voltage measurements and for characterizing low-resistance materials and devices. Featuring an outstanding 1 nV sensitivity, integrated delta mode coordination with Keithley current sources, and extremely low noise, it is a industry standard for precision metrology and superconductivity research.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488.2), RS-232, Trigger Link
Display2-line Vacuum Fluorescent Display (VFD)
Warm-up Time2.5 hours to rated accuracy

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V ±10%, 45 Hz to 66 Hz & 360 Hz to 440 Hz, 22 VA
Operating Temperature0°C to 50°C
Dimensions (W×H×D)213 mm × 89 mm × 370 mm
Weight5 kg

System Capabilities and Functions

ParameterValue
Number of Input Channels2 (Channel 1: Nanovolt/Voltage, Channel 2: Voltage/Temperature)
Delta Mode SupportYes, coordinated low-resistance measurements with Keithley 6220/6221 current sources
Internal Buffer Capacity1024 readings
Math FunctionsMX+B, Percentage, Ratio, Difference
Filter SettingsAnalog, Digital (Averaging and Advanced)

DC Voltage Performance

ParameterValue
Channel 1 Voltage Ranges10 mV, 100 mV, 1 V, 10 V, 100 V
Channel 2 Voltage Ranges100 mV, 1 V, 10 V
Best Voltage Resolution1 nV
Channel 1 Input Impedance>10 GΩ (10 mV to 10 V ranges), 10 MΩ ± 1% (100 V range)
Channel 2 Input Impedance>10 GΩ (all ranges)
Maximum Input Voltage (Channel 1)120 V peak (continuous)
Maximum Input Voltage (Channel 2)15 V peak (continuous)

Noise and Rejection

ParameterValue
Channel-to-Channel Isolation>10^10 Ω in parallel with <150 pF
Common Mode Rejection Ratio (CMRR)>140 dB (DC, 50 Hz or 60 Hz with 1 kΩ unbalance in LO lead)
Normal Mode Rejection Ratio (NMRR)>60 dB (50 Hz or 60 Hz, ±0.1%)
Peak-to-Peak Noise (0.1Hz to 10Hz)6 nV p-p (typical at 1 PLC), 1.5 nV p-p (typical at 5 PLC)

Measurement Speeds and Integration

ParameterValue
Integration Time Options (NPLC)0.01 PLC to 50 PLC (50 Hz), 0.01 PLC to 60 PLC (60 Hz)
Maximum Reading Rate to Buffer2000 readings/second (at 0.01 PLC, 16-bit resolution)

Temperature Measurement

ParameterValue
Supported ThermocouplesJ, K, T, E, R, S, B, N
Temperature Measurement Resolution0.001 °C
Reference Junction Compensation Accuracy±0.5 °C

Input and Output Interfaces

ParameterValue
Primary Input Connector TypeLow-thermal LEMO connector (front panel)
Analog Output Voltage Range±1.2 V full scale
Analog Output Resolution12-bit
Analog Output Impedance1 kΩ nominal
Digital I/O Port Configuration4-bit open-collector outputs, 1 trigger input, 1 end-of-measurement output
Trigger Link Connector8-pin micro-DIN

Physical Characteristics

ParameterValue
Rackmount SupportYes, 2U half-rack

Environmental and Safety Compliance

ParameterValue
Storage Temperature Range-40 °C to 70 °C
Operating Humidity RangeUp to 80% RH at up to 35°C (non-condensing)

Power Supply

ParameterValue
Line Frequency45 Hz to 66 Hz and 360 Hz to 440 Hz (automatically sensed)
Power Consumption22 VA

Measurement Performance

ParameterValue
Maximum Common Mode Voltage350 V peak

Analog Output

ParameterValue
Analog Output Accuracy±(0.15% of output + 1 mV)
Analog Output Settling Time2.5 ms (nominal)

Triggering

ParameterValue
Trigger InputTTL compatible, minimum 2 µs pulse width
Trigger OutputTTL compatible, ~15 µs pulse width

Interfaces

ParameterValue
Command Language SupportSCPI (Standard Commands for Programmable Instruments)

Delta & Pulse Modes

ParameterValue
Delta Mode CompatibilityKeithley 6220 and 6221 current sources
Pulse Mode CompatibilityKeithley 6221 current source

A/D Converter

ParameterValue
A/D Converter ArchitectureMulti-slope integrating

Digital I/O

ParameterValue
Digital I/O Max Sink Current100 mA

Environmental

ParameterValue
Operating Altitude Limit2000 m (6562 ft) above sea level
Shock and Vibration SpecificationMIL-T-28800E Type III, Class 5

Compliance & Safety

ParameterValue
Safety ConformityConforms to European Union Low Voltage Directive, IEC 61010-1
EMC ConformityConforms to European Union EMC Directive, EN 61326-1

Technology

Precision Nanovoltmeter

Applications

Low-voltage metrology, low-resistance material characterization, superconductivity research, temperature measurements