The Keithley 2302 is a single-channel high-speed battery simulator designed specifically for testing portable, battery-operated devices such as cellular phones, transceivers, and medical electronics. It mimics the transient response and variable output resistance of a real battery, enabling precise assessment of device behavior under realistic operating conditions.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488) |
| Display | Vacuum Fluorescent Display (VFD), 2-line |
| Warm-up Time | 1 hour |
| Setup Storage | 5 user-programmable locations plus factory default |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V / 120 V / 220 V / 240 V AC ±10%, 50 Hz / 60 Hz, 150 VA max |
| Operating Temperature | 0°C to 50°C |
| Dimensions (W×H×D) | 213 mm × 89 mm × 411 mm |
| Weight | 3.2 kg |
Simulation & Output Specifications
| Parameter | Value |
|---|---|
| Maximum Voltage Output | 15 V DC |
| Minimum Voltage Output | 0 V DC |
| Voltage Programming Resolution | 1 mV |
| Voltage Programming Accuracy | ±(0.05% + 3 mV) |
| Maximum Current Output | 3 A DC continuous (5 A peak for pulse loads) |
| Maximum Current Sinking | 3 A DC continuous |
| Output Impedance Range | 0 Ω to 1 Ω |
| Output Impedance Resolution | 10 mΩ |
| Transient Recovery Time | < 40 µs to within 100 mV of prior level |
Measurement & Readback
| Parameter | Value |
|---|---|
| Voltage Readback Resolution | 1 mV |
| Voltage Readback Accuracy | ±(0.05% + 3 mV) |
| Current Readback Ranges | 5 A, 5 mA |
| Dynamic Current Measurement Modes | Peak, average, and background current |
| Pulse Current Measurement Integration Time Range | 33.3 µs to 833 ms |
System Protection & Safety
| Parameter | Value |
|---|---|
| Isolation Voltage (Output to Chassis) | 150 V DC |
Connectivity & Control Interfaces
| Parameter | Value |
|---|---|
| Remote Sense Terminals | Yes (4-wire sense) |
| Remote Sense Maximum Lead Drop | 1 V per lead |
| SCPI Command Set Compatibility | Yes |
Physical & Interface Characteristics
| Parameter | Value |
|---|---|
| Form Factor | Benchtop or half-rack (2U height) |
| Cooling Mechanism | Internal fan forced-air cooling |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -20°C to 70°C |
| Operating Humidity Range | 0% to 70% RH non-condensing (up to 35°C) |
| Safety Standards | Conforms to European Union Directive 73/23/EEC (EN 61010-1) |
| EMC Standards | Conforms to European Union Directive 89/336/EEC (EN 61326-1) |
Performance
| Parameter | Value |
|---|---|
| Voltage Line Regulation | 0.01% + 2 mV |
| Voltage Load Regulation | 0.01% + 2 mV |
| Voltage Ripple and Noise (RMS) | <1 mV RMS |
| Voltage Ripple and Noise (Peak-to-Peak) | <3 mV (20 Hz to 20 MHz) |
| Current Ripple and Noise (RMS) | <3 mA RMS |
| Current Readback Accuracy (5 A Range) | ±(0.2% + 200 µA) |
| Current Readback Accuracy (5 mA Range) | ±(0.2% + 1 µA) |
Current Output
| Parameter | Value |
|---|---|
| Current Limit Range | 10 mA to 5 A |
| Current Limit Programming Accuracy | ±(0.2% + 5 mA) |
| Current Limit Programming Resolution | 1.25 mA |
Analog Outputs
| Parameter | Value |
|---|---|
| Current Monitor Output Scaling (5 A Range) | 0 to 3 V (0.6 V/A) |
| Current Monitor Output Scaling (5 mA Range) | 0 to 3 V (600 V/A) |
Interfaces
| Parameter | Value |
|---|---|
| Trigger Link Interface | 8-pin micro-DIN connector |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 2000 m |
Output Specifications
| Parameter | Value |
|---|---|
| Number of Channels | 1 |
Digital Voltmeter (DVM)
| Parameter | Value |
|---|---|
| DVM Input Voltage Range | -5 V to +30 V DC |
| DVM Input Impedance | 2 GΩ typical |
| DVM Readback Accuracy | ±(0.05% + 3 mV) |
| DVM Readback Resolution | 1 mV |
Power Specifications
| Parameter | Value |
|---|---|
| Mains Frequency | 50 Hz / 60 Hz (automatically sensed) |
| Power Consumption | 150 VA maximum |
Technology
Linear Battery Simulation
Applications
Battery simulation and testing for low-power wireless and portable devices
Looking for the Keithley 2302?
Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.