The Keithley 2303 is a 45W High-Speed Battery Simulating DC Power Supply designed for testing battery-operated, wireless communication devices. It is optimized to provide outstanding voltage stability during fast pulse load changes, such as those found in cellular devices. It features precision readback of load currents (including short pulses), is capable of sinking current to simulate a discharged battery, and offers a continuous output of 15V at 3A or 9V at 5A.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488) |
| Display | Dual-line Vacuum Fluorescent Display (VFD) |
| Standard Warranty | 1 year |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100V / 120V / 220V / 240V AC (switch-selectable), 50/60 Hz, 150 VA max |
| Operating Temperature | 0 to 50 °C |
| Dimensions (W×H×D) | 213 mm × 89 mm × 411 mm |
| Weight | 5.4 kg |
DC Output Specifications
| Parameter | Value |
|---|---|
| Output Voltage Range | 0 to 15 V DC |
| Output Current Range | 0 to 3 A DC |
| Maximum Output Power | 45 W |
| Sinking Current Range | 0 to 2 A DC |
| Voltage Programming Accuracy | ±(0.05% + 10 mV) |
| Voltage Programming Resolution | 5 mV |
| Current Limit Programming Range | 0 to 3.1 A |
| Current Limit Programming Accuracy | ±(0.16% + 5 mA) |
| Current Limit Programming Resolution | 1.25 mA |
| Voltage Line Regulation | ±(0.01% + 2 mV) |
| Voltage Load Regulation | ±(0.01% + 2 mV) |
| Current Line Regulation | ±(0.01% + 1 mA) |
| Current Load Regulation | ±(0.01% + 1 mA) |
| Output Ripple and Noise (Peak-to-Peak) | <3 mV p-p (20 Hz to 20 MHz) |
| Output Ripple and Noise (RMS) | <1 mV RMS (20 Hz to 20 MHz) |
Transient Response
| Parameter | Value |
|---|---|
| Transient Voltage Recovery Time | <40 µs |
| Transient Voltage Recovery Level | to within 50 mV of original level |
| Transient Voltage Drop | <75 mV |
| Bandwidth Selection | High and Low settings |
Current Measurement (Readback)
| Parameter | Value |
|---|---|
| Current Readback Ranges | 5 A, 5 mA, 500 µA |
| Current Readback Accuracy (5A Range) | ±(0.2% + 400 µA) |
| Current Readback Accuracy (5mA Range) | ±(0.2% + 1 µA) |
| Current Readback Accuracy (500µA Range) | ±(0.2% + 100 nA) |
| Current Readback Resolution (5mA Range) | 0.1 µA |
| Current Readback Resolution (500µA Range) | 10 nA |
| DC Current Sensitivity | 10 nA |
| Pulse Current Measurement Modes | Pulse Peak, Average, and Baseline |
| Pulse Current Trigger Level Range | 5 mA to 5 A (in 5 mA steps) |
| Pulse Current Integration Time Range | 33.3 µs to 833 ms |
Current Measurement (Readback
| Parameter | Value |
|---|---|
| Current Readback Resolution (5A Range) | 100 µA |
Voltage Measurement (Readback)
| Parameter | Value |
|---|---|
| Voltage Readback Range | 0 to 15 V DC |
| Voltage Readback Accuracy | ±(0.05% + 3 mV) |
| Voltage Readback Resolution | 1 mV |
| Remote Sense Voltage Range | Up to 15 V DC |
| Maximum Remote Sense Line Drop Compensation | 1 V per lead |
Digital and Control Interfaces
| Parameter | Value |
|---|---|
| GPIB Interface | IEEE-488.2 conforming |
| Programming Language | SCPI (Standard Commands for Programmable Instruments) |
| Trigger Delay Range | 0 to 100 ms |
Protection and Safety
| Parameter | Value |
|---|---|
| Overvoltage Protection (OVP) Range | 0 to 15 V (limit settings) |
| Output Enable/Disable Control | Yes, front panel button and SCPI controllable |
Power Requirements
| Parameter | Value |
|---|---|
| Input AC Voltage Range | 100V / 120V / 220V / 240V AC (switch-selectable) |
| Input Frequency Range | 50 Hz / 60 Hz |
| Maximum Power Consumption | 150 VA |
Physical Specifications
| Parameter | Value |
|---|---|
| Width | 213 mm (8.4 in) |
| Height | 89 mm (3.5 in) |
| Depth | 411 mm (16.2 in) |
| Net Weight | 3.2 kg (7.1 lbs) |
| Shipping Weight | 5.4 kg (12 lbs) |
| Cooling Method | Forced air (internal fan) |
| Rack Mounting Configurations | Single or dual rack mount (2U high, half-rack width) |
Environmental and Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -20 °C to 70 °C |
| Operating Humidity Range | 0% to 70% RH, non-condensing (up to 35 °C) |
| Recommended Calibration Cycle | 1 year |
Performance & Accuracy
| Parameter | Value |
|---|---|
| Pulse Current Measurement Accuracy (5A Range, >1ms) | ±(0.2% + 10mA) |
| Temperature Coefficient (Voltage Programming) | ±(0.005% + 2mV)/°C |
| Temperature Coefficient (Voltage Readback) | ±(0.005% + 2mV)/°C |
| Temperature Coefficient (Current Limit Programming) | ±(0.05% + 1mA)/°C |
| Temperature Coefficient (Current Readback, 5A Range) | ±(0.05% + 1mA)/°C |
| Temperature Coefficient (Current Readback, 5mA Range) | ±(0.05% + 1µA)/°C |
| Output Impedance | 0.01 Ω (typical at DC) |
| Overvoltage Protection Accuracy | ±(0.5% + 200mV) |
| Overvoltage Protection Response Time | <150 µs (typical) |
Remote Sensing
| Parameter | Value |
|---|---|
| Maximum Sense Lead Resistance | 0.5 Ω (per lead for rated accuracy) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Output Connector | 8-pin rear-panel quick-disconnect screw terminal block |
| Trigger Link Connector | 8-pin micro-DIN |
Power & Safety
| Parameter | Value |
|---|---|
| Isolation Voltage (Output to Chassis) | 150 VDC maximum |
| Warm-up Time | 1 hour |
| Mains Fuse (100V/120V) | 2.0 A, 250 V, slow-blow (5x20 mm) |
| Mains Fuse (220V/240V) | 1.0 A, 250 V, slow-blow (5x20 mm) |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 2000 m |
Output Specifications
| Parameter | Value |
|---|---|
| Number of Channels | 1 |
Measurement
| Parameter | Value |
|---|---|
| A/D Converter Resolution | 16-bit |
Compliance
| Parameter | Value |
|---|---|
| Safety Compliance | Conforms to European Union Directive 73/23/EEC (EN 61010-1) |
| EMC Compliance | Conforms to European Union Directive 89/336/EEC (EN 55011, EN 50082-1) |
Technology
Battery Simulating DC Power Supply
Applications
Battery simulation and power consumption testing for wireless and portable battery-powered devices