The Keithley 2400 is a standard 20 W SourceMeter instrument that tightly couples a highly stable DC power source with a true instrument-grade 5½-digit multimeter. Operating as a compact, single-channel DC parametric tester, it functions as a voltage source, current source, voltage meter, current meter, and ohmmeter. With 4-quadrant operation, it is capable of sourcing and sinking power, making it ideal for testing components like diodes, resistors, active circuit protection devices, and portable battery-powered electronics.

Specifications

General

ParameterValue
InterfacesIEEE-488 (GPIB), RS-232, Keithley Trigger Link, Programmable Digital I/O
DisplayVacuum Fluorescent Display (VFD)
Warm-up Time1 hour to rated accuracy
Calibration Interval1 year
Maximum Common Mode Voltage250 V DC

Physical

ParameterValue
Operating Temperature0 to 50
Dimensions (W×H×D)213 × 89 × 370
Weight3.21
CoolingConvection

System Specifications

ParameterValue
Maximum Output Power20 W
Quadrant Operation4-quadrant (Source and sink)
Instrument FunctionsVoltage Source, Current Source, Voltage Meter, Current Meter, Ohmmeter

Voltage Source Specifications

ParameterValue
Voltage Source Ranges±200 mV, ±2 V, ±20 V, ±200 V
Minimum Source Voltage Resolution5 µV

Current Source Specifications

ParameterValue
Current Source Ranges±1 µA, ±10 µA, ±100 µA, ±1 mA, ±10 mA, ±100 mA, ±1 A

Voltage Measurement Specifications

ParameterValue
Voltage Measurement Ranges±200 mV, ±2 V, ±20 V, ±200 V
Minimum Measurement Voltage Resolution1 µV

Current Measurement Specifications

ParameterValue
Current Measurement Ranges±1 µA, ±10 µA, ±100 µA, ±1 mA, ±10 mA, ±100 mA, ±1 A
Minimum Measurement Current Resolution10 pA

Measurement Specifications

ParameterValue
Basic Measurement Accuracy0.012%
Measurement Resolution5½ digits

Resistance Measurement Specifications

ParameterValue
Remote Sensing Capability2-, 4-, and 6-wire configurations

System Speeds and Timing

ParameterValue
Maximum Reading Rate1700 readings/second at 4½ digits (via GPIB)

Math and Memory Capabilities

ParameterValue
Pass/Fail ComparatorBuilt-in for fast sorting/binning
Data Buffer Size5000 readings

Interfaces and Connectivity

ParameterValue
Supported SoftwareKeithley LabTracer 2.0 I-V curve tracing software
Command SetSCPI

Physical and Mechanical

ParameterValue
Form FactorHalf-rack, 2U high

Environmental and Compliance

ParameterValue
Storage Temperature-25 °C to 65 °C

Measurement

ParameterValue
Input Impedance> 10 GΩ (Voltage Measurement)
Common Mode Rejection Ratio (CMRR)> 120 dB (at DC, 50/60 Hz)
Normal Mode Rejection Ratio (NMRR)> 60 dB (at 50/60 Hz)

Instrument Functions

ParameterValue
Sweep TypesLinear, Logarithmic, Custom, Source-Memory

Connectivity

ParameterValue
Trigger Link8-pin Micro-DIN
Digital I/ODB-9 female (Component Handler Interface)

Power

ParameterValue
Line Frequency50 Hz to 60 Hz
Power Consumption210 VA maximum

Performance

ParameterValue
Overrange105% of range, source and measure
Source Delay0 to 9999.999 seconds
Compliance LimitsBipolar, user-programmable voltage or current limit

Compliance

ParameterValue
Safety ComplianceEN 61010-1
EMC ComplianceEN 61326-1

Environmental

ParameterValue
Maximum Operating Altitude2000 m
Pollution Degree2
Operating EnvironmentIndoor use only

Technology

Source Measure Unit (SMU)

Applications

DC parametric testing, IDDQ testing, battery-powered device testing, component testing