The Keithley 2401 is a low-voltage SourceMeter (SMU) instrument designed for high-precision voltage and current sourcing and measurement. Part of the industry-standard Keithley 2400 series, it is optimized for low-voltage applications such as testing photovoltaic cells, semiconductor devices, and other components requiring precise 4-quadrant operation up to 20V and 1A.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488), RS-232, Digital I/O, Trigger Link |
| Display | 5×7 dot matrix vacuum fluorescent display (VFD) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V to 240 V AC (±10%), 50 Hz or 60 Hz, 190 VA maximum |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 213 mm × 89 mm × 370 mm |
| Weight | 3.21 kg |
Voltage Source
| Parameter | Value |
|---|---|
| Maximum Output Voltage | 21.0000 V |
| Voltage Ranges | 200 mV, 2 V, 20 V |
| Programming Resolution (200mV range) | 5 µV |
| Programming Resolution (2V range) | 50 µV |
| Programming Resolution (20V range) | 500 µV |
| Programming Accuracy (200mV range) | ±(0.02% + 600 µV) |
| Programming Accuracy (2V range) | ±(0.02% + 600 µV) |
| Programming Accuracy (20V range) | ±(0.02% + 2.4 mV) |
| Temperature Coefficient (Voltage Source) | ±(0.15 × accuracy specification) / °C (0°C–18°C & 28°C–50°C) |
| Noise (0.1Hz to 10Hz) | 30 µV p-p (typical on 20V range) |
| Line Regulation (Voltage) | 0.01% of range |
| Load Regulation (Voltage) | 0.01% of range + 100 µV |
Current Limit and Source
| Parameter | Value |
|---|---|
| Maximum Output Current | 1.05 A |
| Current Ranges | 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A |
| Programming Resolution (1µA range) | 50 pA |
| Programming Resolution (10µA range) | 500 pA |
| Programming Resolution (100µA range) | 5 nA |
| Programming Resolution (1mA range) | 50 nA |
| Programming Resolution (10mA range) | 500 nA |
| Programming Resolution (100mA range) | 5 µA |
| Programming Accuracy (1µA range) | ±(0.035% + 600 pA) |
| Programming Accuracy (10µA range) | ±(0.033% + 2 nA) |
| Programming Accuracy (100µA range) | ±(0.031% + 20 nA) |
| Programming Accuracy (1mA range) | ±(0.034% + 200 nA) |
| Programming Accuracy (10mA range) | ±(0.034% + 2 µA) |
| Programming Accuracy (100mA range) | ±(0.034% + 20 µA) |
| Temperature Coefficient (Current Limit) | ±(0.15 × accuracy specification) / °C (0°C–18°C & 28°C–50°C) |
| Line Regulation (Current) | 0.01% of range |
| Load Regulation (Current) | 0.01% of range + 100 pA |
Measurement
| Parameter | Value |
|---|---|
| Voltage Measurement Ranges | 200 mV, 2 V, 20 V |
| Voltage Measurement Resolution (200mV range) | 100 nV |
| Voltage Measurement Resolution (2V range) | 1 µV |
| Voltage Measurement Resolution (20V range) | 10 µV |
| Voltage Measurement Accuracy (200mV range) | ±(0.012% + 300 µV) |
| Voltage Measurement Accuracy (2V range) | ±(0.012% + 300 µV) |
| Voltage Measurement Accuracy (20V range) | ±(0.015% + 1.5 mV) |
| Voltage Measurement Input Impedance | > 10 GΩ |
| Current Measurement Ranges | 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A |
| Current Measurement Resolution (1µA range) | 10 pA |
| Current Measurement Resolution (10µA range) | 100 pA |
| Current Measurement Resolution (100µA range) | 1 nA |
| Current Measurement Resolution (1mA range) | 10 nA |
| Current Measurement Resolution (10mA range) | 100 nA |
| Current Measurement Resolution (100mA range) | 1 µA |
| Current Measurement Accuracy (1µA range) | ±(0.029% + 300 pA) |
| Current Measurement Accuracy (10µA range) | ±(0.027% + 700 pA) |
| Current Measurement Accuracy (100µA range) | ±(0.025% + 6 nA) |
| Current Measurement Accuracy (1mA range) | ±(0.027% + 60 nA) |
| Current Measurement Accuracy (10mA range) | ±(0.035% + 600 nA) |
| Current Measurement Accuracy (100mA range) | ±(0.055% + 6 µA) |
| Temperature Coefficient (Measurement) | ±(0.15 × accuracy specification) / °C (0°C–18°C & 28°C–50°C) |
| Integration Time Range | 0.01 PLC to 10 PLC |
Output and Connection Characteristics
| Parameter | Value |
|---|---|
| Output Connector Type | Front panel banana jacks, rear panel screw terminal block |
| Remote Sensing (4-Wire Sense) | Supported (up to 1V drop per lead) |
| Maximum Sense Lead Resistance | 1 MΩ for rated accuracy |
| Maximum Common Mode Voltage | 40 V DC |
System, Control and Connectivity
| Parameter | Value |
|---|---|
| Command Language | SCPI (Standard Commands for Programmable Instruments) |
| Data Buffer/Memory Capacity | 5000 readings |
| Maximum Reading Rate (to memory) | 2081 readings/s (at 0.01 NPLC) |
| Maximum Reading Rate (to GPIB) | 1750 readings/s (at 0.01 NPLC) |
| Digital I/O Port | 4-bit I/O port, DB-9 connector |
Power Requirements
| Parameter | Value |
|---|---|
| Input Frequency Range | 47 Hz to 63 Hz (automatic sensing) |
Physical and Environmental Specifications
| Parameter | Value |
|---|---|
| Storage Temperature Range | -25 °C to 65 °C |
| Operating Humidity Range | Maximum 70% RH at up to 35 °C (non-condensing) |
| Warm-up Time | 1 hour to rated specifications |
| Cooling Method | Forced air cooling (internal fan) |
Output Characteristics
| Parameter | Value |
|---|---|
| Maximum Output Power | 20 W |
| Overvoltage Protection (OVP) Range | 2 V to 21 V |
Resistance Measurement
| Parameter | Value |
|---|---|
| Resistance Measurement Ranges | 2.00000 Ω to 200.000 MΩ |
System Performance
| Parameter | Value |
|---|---|
| Common Mode Rejection Ratio (CMRR) | > 120 dB (DC, 50 Hz, 60 Hz) |
| Normal Mode Rejection Ratio (NMRR) | > 60 dB (50 Hz, 60 Hz) at 1 PLC or greater |
| Source Delay Range | 0 s to 9999.999 s |
Power and Safety
| Parameter | Value |
|---|---|
| Maximum Power Consumption | 190 VA |
| Safety Compliance | Conforms to European Union Directive 73/23/EEC, EN 61010-1 |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Trigger Link Connector | 8-pin micro DIN |
| Active Guard Terminal | Yes, available on rear panel |
Guard
| Parameter | Value |
|---|---|
| Guard Offset Voltage | < 150 µV |
| Guard Output Impedance | < 0.1 Ω |
| Maximum Guard Current | 50 mA |
Source Specifications
| Parameter | Value |
|---|---|
| Voltage Source Settling Time | < 250 µs to 0.1% of final value |
| Current Source Settling Time | < 250 µs to 0.1% of final value |
| Voltage Source Overshoot | < 0.1% typical (full scale step, resistive load) |
| Current Source Overshoot | < 0.1% typical (full scale step, resistive load) |
| Sweep Type | Linear, Log, Custom |
Memory
| Parameter | Value |
|---|---|
| Source Memory | 100-point |
Display
| Parameter | Value |
|---|---|
| Display Resolution | 5.5 digits |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 2000 m |
Technology
Source Measure Unit (SMU)
Applications
Precision DC sourcing and measurement of low-voltage semiconductor and electronic components