The Keithley Model 2420 High Voltage SourceMeter is a high-performance, 60W source measure unit (SMU) designed for high current or medium power applications. Combining a highly stable DC power source and a precision 5½-digit multimeter, it enables tightly coupled precision voltage and current sourcing and measurement. The 2420 offers 4-quadrant source and sink capabilities, high repeatability, and low noise, making it highly suitable for testing resistors, thermistors, solar cells, batteries, and high-current diodes.

Specifications

General

ParameterValue
InterfacesGPIB (SCPI), RS-232, Keithley Trigger Link, Programmable Digital I/O (DIO)
DisplayDouble-line Vacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz, 190 VA maximum
Operating Temperature0 °C to 50 °C
Dimensions (W×H×D)213 mm × 89 mm × 370 mm
Weight4.1
CoolingForced air, internal fan

Resolution and Display

ParameterValue
Digits of Resolution5½ digits

DC Voltage

ParameterValue
DCV Ranges200 mV, 2 V, 20 V, 60 V
DCV Best 1-Year Accuracy0.012% basic measure accuracy
DCV Measurement Resolution±1 µV
Voltage Programming Resolution (200 mV range)5 µV
Voltage Programming Resolution (2 V range)50 µV
Voltage Programming Resolution (20 V range)500 µV
Voltage Programming Resolution (60 V range)1.5 mV
Voltage Sourcing/Measuring Limit±60 V

DC Current

ParameterValue
DCI Ranges10 µA to 3 A
Current Sourcing/Measuring Limit±3 A

System Speed and Performance

ParameterValue
Maximum Output Power60 W
Source and Sink Operation4-quadrant operation
Sensing Configurations2- and 4-wire remote V-source and measure sensing
Transient Response Time<30 µs
Contact Check FunctionOnly available on 2420-C

Power and Environmental

ParameterValue
DC Floating VoltageUp to ±250 V DC maximum from chassis ground

Triggering and Math Functions

ParameterValue
Pass/Fail ComparatorYes, for fast sorting and binning

Interfaces and Connectivity

ParameterValue
Digital I/O PortProgrammable DIO port for automation/handler/prober control
Supported SoftwareKeithley LabTracer 2.0 I-V curve tracing application software

Safety & Electrical

ParameterValue
Common Mode Voltage250 V DC maximum (LO to chassis)

Performance

ParameterValue
Warm-up Time1 hour to rated accuracy
Overrange Limit105% of range (source and measure)

Memory & Logging

ParameterValue
Reading Buffer Size5,000 readings (volatile)

Control & Software

ParameterValue
Command LanguageSCPI-1996.0

Connectivity

ParameterValue
Trigger Link8-pin micro-DIN connector

Protection & Limits

ParameterValue
Overvoltage Protection Range2 V to 66 V

Measurement Settings

ParameterValue
Max Sense Lead Resistance1 MΩ
Input Impedance>10 GΩ
Common Mode Impedance>10 GΩ in parallel with <290 pF

Power & Safety

ParameterValue
Power Consumption190 VA max.

Environmental

ParameterValue
Operating Humidity70% RH up to 35 °C (derated 3% RH/°C from 35 °C to 50 °C)
Operating Altitude0 to 2000 m

Compliance

ParameterValue
Safety ComplianceConforms to European Union Directive 73/23/EEC, EN 61010-1
EMC ComplianceConforms to European Union Directive 89/336/EEC, EN 55011, EN 50082-1
VibrationMIL-PRF-28800F Class 3

Technology

Source Measure Unit (SMU)

Applications

Precision DC sourcing and measurement for high-current component testing

Looking for the Keithley 2420?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us