The Keithley 2425 is a 100W, 3A SourceMeter Unit (SMU) designed for high-power precision electronic testing, providing four-quadrant voltage and current sourcing and measurement in a compact half-rack instrument.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488), RS-232, Digital I/O |
| Display | 12-character alphanumeric vacuum fluorescent (VFD), 2 lines |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V to 240 V AC, 50 Hz/60 Hz, 250 VA maximum |
| Operating Temperature | 0°C to 50°C |
| Dimensions (W×H×D) | 213 mm × 89 mm × 370 mm |
| Weight | 4.1 kg |
Voltage Source Specifications
| Parameter | Value |
|---|---|
| Voltage Source Ranges | 200 mV, 2 V, 20 V, 100 V |
| Voltage Source 200mV Range Resolution | 5 µV |
| Voltage Source 2V Range Resolution | 50 µV |
| Voltage Source 20V Range Resolution | 500 µV |
| Voltage Source 100V Range Resolution | 5 mV |
| Voltage Source 200mV Range Accuracy | ±(0.02% of rdg + 600 µV) |
| Voltage Source 2V Range Accuracy | ±(0.02% of rdg + 600 µV) |
| Voltage Source 20V Range Accuracy | ±(0.02% of rdg + 2.4 mV) |
| Voltage Source 100V Range Accuracy | ±(0.02% of rdg + 12 mV) |
| Voltage Limit Compliance Range | 200 mV to 105 V |
Current Source Specifications
| Parameter | Value |
|---|---|
| Current Source Ranges | 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A |
| Current Source 10µA Range Resolution | 500 pA |
| Current Source 100µA Range Resolution | 5 nA |
| Current Source 1mA Range Resolution | 50 nA |
| Current Source 10mA Range Resolution | 500 nA |
| Current Source 100mA Range Resolution | 5 µA |
| Current Source 1A Range Resolution | 50 µA |
| Current Source 3A Range Resolution | 150 µA |
| Current Source 10µA Range Accuracy | ±(0.033% of rdg + 6 nA) |
| Current Source 100µA Range Accuracy | ±(0.033% of rdg + 60 nA) |
| Current Source 1mA Range Accuracy | ±(0.033% of rdg + 600 nA) |
| Current Source 10mA Range Accuracy | ±(0.033% of rdg + 6 µA) |
| Current Source 100mA Range Accuracy | ±(0.033% of rdg + 60 µA) |
| Current Source 1A Range Accuracy | ±(0.067% of rdg + 900 µA) |
| Current Source 3A Range Accuracy | ±(0.067% of rdg + 2.7 mA) |
| Current Limit Compliance Range | 10 µA to 3.15 A |
Voltage Measurement Specifications
| Parameter | Value |
|---|---|
| Voltage Measure Ranges | 200 mV, 2 V, 20 V, 100 V |
| Voltage Measure 200mV Range Resolution | 1 µV |
| Voltage Measure 2V Range Resolution | 10 µV |
| Voltage Measure 20V Range Resolution | 100 µV |
| Voltage Measure 100V Range Resolution | 1 mV |
| Voltage Measure 200mV Range Accuracy | ±(0.012% of rdg + 300 µV) |
| Voltage Measure 2V Range Accuracy | ±(0.012% of rdg + 300 µV) |
| Voltage Measure 20V Range Accuracy | ±(0.015% of rdg + 1.5 mV) |
| Voltage Measure 100V Range Accuracy | ±(0.015% of rdg + 10 mV) |
| Voltage Measure Input Impedance | >10 GΩ |
Current Measurement Specifications
| Parameter | Value |
|---|---|
| Current Measure Ranges | 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A |
| Current Measure 10µA Range Resolution | 100 pA |
| Current Measure 100µA Range Resolution | 1 nA |
| Current Measure 1mA Range Resolution | 10 nA |
| Current Measure 10mA Range Resolution | 100 nA |
| Current Measure 100mA Range Resolution | 1 µA |
| Current Measure 1A Range Resolution | 10 µA |
| Current Measure 3A Range Resolution | 100 µA |
| Current Measure 10µA Range Accuracy | ±(0.027% of rdg + 700 pA) |
| Current Measure 100µA Range Accuracy | ±(0.025% of rdg + 6 nA) |
| Current Measure 1mA Range Accuracy | ±(0.027% of rdg + 60 nA) |
| Current Measure 10mA Range Accuracy | ±(0.025% of rdg + 600 nA) |
| Current Measure 100mA Range Accuracy | ±(0.027% of rdg + 6 µA) |
| Current Measure 1A Range Accuracy | ±(0.05% of rdg + 600 µA) |
| Current Measure 3A Range Accuracy | ±(0.05% of rdg + 1.8 mA) |
Operating Ranges & Power Envelope
| Parameter | Value |
|---|---|
| Maximum Output Power | 100 W (source), 40 W (sink) |
| Quadrant Operation Limits | 4-Quadrant (source or sink capability) |
| Maximum Common Mode Voltage | 250 V DC |
| Guard Output Resistance | <0.1 Ω |
| Guard Offset Voltage | <150 µV |
System Speed & Memory
| Parameter | Value |
|---|---|
| Reading Buffer Capacity | 5,000 readings |
| Maximum Pass/Fail Test Rate | 500 µs per test (2,000 tests/s) |
Connectivity & Interfaces
| Parameter | Value |
|---|---|
| GPIB Interface | IEEE-488.1 and IEEE-488.2 compliant |
| RS-232 Interface | DB-9 male, 3-wire or 5-wire |
| Digital I/O Port | 4-bit output, 1-bit input, DB-9 female component handler interface |
| Sensing Configuration | 2-wire local sense or 4-wire remote sense |
| Output Connection Type | Front-panel safety banana jacks, rear-panel screw terminal barrier strip |
| Trigger Link Port | 8-pin micro-DIN |
Software & Programming
| Parameter | Value |
|---|---|
| Programming Languages / Command Sets | SCPI (Standard Commands for Programmable Instruments) |
| Driver Support | LabVIEW, LabWindows/CVI, IVI-C |
Power & Environmental
| Parameter | Value |
|---|---|
| AC Input Voltage Range | 100 V to 240 V AC (±10%) |
| AC Input Frequency Range | 50 Hz / 60 Hz |
| Storage Temperature Range | -40°C to 70°C |
| Operating Humidity Range | Maximum 70% RH at 35°C (non-condensing) |
| Cooling Mechanism | Forced air cooling (internal fan) |
Physical & Safety Specs
| Parameter | Value |
|---|---|
| Calibration Interval | 1 Year |
| Overvoltage Protection | 2V, 20V, 40V, 60V, 80V, 100V, NONE |
Resistance Measurement
| Parameter | Value |
|---|---|
| Resistance Measure Ranges | 2.0 Ω to 20 MΩ |
| Max Lead Resistance (4-wire) | 10% of range per lead (20 Ω to 200 Ω ranges), 1 kΩ per lead (2 kΩ to 20 MΩ ranges) |
Performance & Speed
| Parameter | Value |
|---|---|
| Measurement Speed (0.01 PLC) | 1750 readings/s (to memory) |
| Measurement Speed (0.1 PLC) | 740 readings/s (to memory) |
| Measurement Speed (1.0 PLC) | 44 readings/s (to memory) |
| NMRR | >60 dB (at 50 Hz or 60 Hz with 1 PLC or greater) |
| CMRR | 120 dB (at DC, 50 Hz, or 60 Hz) |
Sweep Capabilities
| Parameter | Value |
|---|---|
| Sweep Types | Linear, Log, Custom, Source Memory |
| Source Memory List Size | 100 points |
General & Environmental
| Parameter | Value |
|---|---|
| Warm-up Time | 1 hour |
| Maximum Altitude | 2000 m |
| Power Consumption | 250 VA |
Guard Specifications
| Parameter | Value |
|---|---|
| Maximum Guard Output Current | 2 mA |
Sensing Specifications
| Parameter | Value |
|---|---|
| Maximum Remote Sense Lead Drop | 3 V (maximum voltage difference between Source and Sense HI or LO) |
Interfaces
| Parameter | Value |
|---|---|
| Digital I/O Lines | 4 |
Memory & Storage
| Parameter | Value |
|---|---|
| Saved Setups Capacity | 3 user-defined setups |
Programming & Control
| Parameter | Value |
|---|---|
| SCPI Version Compliance | SCPI-1996.0 |
Physical Specifications
| Parameter | Value |
|---|---|
| Rackmount Dimensions | 2U height, half-rack width |
Performance
| Parameter | Value |
|---|---|
| Transient Response Time | < 250 µs (to recover to within 0.1% after 50% load change) |