The Keithley 2520 Pulsed Laser Diode Test System is an integrated, high-speed test solution designed for production testing and LIV characterization of laser diodes. It combines high-accuracy pulsed current sourcing with three synchronized measurement channels for voltage and optical power.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232, Trigger Link, Digital I/O
DisplayVacuum Fluorescent Display (VFD), 2 lines x 20 characters
Buffer Storage Capacity1000 points per channel
Cooling MethodInternal fan
Warm-up Time1 hour

Physical

ParameterValue
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz, 120 VA max
Operating Temperature0 °C to 50 °C
Dimensions (W×H×D)213 mm x 89 mm x 370 mm
Weight4.7 kg
Rackmount Height2U (half-rack width)

Source Channel

ParameterValue
Laser Diode Drive Current Ranges500 mA, 5 A
Maximum Drive Current5.0 A
Source Voltage Compliance10 V
Pulse Width Range500 ns to 10 ms
Pulse Period Range20 µs to 200 ms
Pulse Duty CycleUp to 4.0%
Drive Current Resolution (500 mA Range)10 µA
Drive Current Resolution (5 A Range)100 µA
Drive Current Accuracy (500 mA Range)±(0.3% + 500 µA)
Drive Current Accuracy (5 A Range)±(0.3% + 5 mA)
Pulse Rise/Fall Time< 60 ns into 1 Ω load (typical)
Pulse Width Programming Resolution10 ns
Pulse Period Programming Resolution10 µs

Voltage Measurement Channel

ParameterValue
Voltage Measurement Ranges5 V, 10 V
Voltage Resolution (5 V Range)150 µV
Voltage Resolution (10 V Range)300 µV
Voltage Measurement Accuracy (5 V Range)±(0.3% + 3 mV)
Voltage Measurement Accuracy (10 V Range)±(0.3% + 6 mV)

Photodiode Measurement Channels

ParameterValue
Photodiode Current Ranges10 mA, 20 mA, 50 mA, 100 mA
PD Current Accuracy (10 mA Range)±(0.3% + 30 µA)
PD Current Accuracy (20 mA Range)±(0.3% + 60 µA)
PD Current Accuracy (50 mA Range)±(0.3% + 150 µA)
PD Current Accuracy (100 mA Range)±(0.3% + 300 µA)

Triggering and Synchronization

ParameterValue
Trigger Link Port8-pin micro-DIN
Digital I/O Port4-bit output, 1-bit input

Analog Outputs

ParameterValue
Laser Diode Current Monitor (IMON)0 to 5 V output
Laser Diode Voltage Monitor (VMON)0 to 5 V output
Photodiode 1 Current Monitor (PD1MON)0 to 5 V output
Photodiode 2 Current Monitor (PD2MON)0 to 5 V output

Environmental

ParameterValue
Operating HumidityUp to 70% R.H. non-condensing (at 35 °C)
Storage Temperature Range-25 °C to 65 °C
Maximum Operating Altitude2000 m

Photodiode Current Measurement

ParameterValue
PD Current Resolution (10 mA Range)100 nA
PD Current Resolution (20 mA Range)200 nA
PD Current Resolution (50 mA Range)500 nA
PD Current Resolution (100 mA Range)1 µA

Pulse Characteristics

ParameterValue
Pulse Width Programming Accuracy±(1% of program + 20 ns)
Pulse Period Programming Accuracy±(1% of program + 200 ns)
Pulse Jitter< 2 ns (typical, sample-to-sample)
Pulse Overshoot / Undershoot< 5% (< 10% on 5 A range) typical

Voltage Measurement

ParameterValue
Voltage Measurement Input Impedance> 10 GΩ

Monitor Outputs

ParameterValue
Analog Monitor Output Impedance50 Ω

Triggering

ParameterValue
Trigger Input to Pulse Output Delay400 ns (typical)
Trigger Output Pulse Width2 µs (typical)

Photodiode Bias Sources

ParameterValue
Photodiode Bias Voltage Range0 to 20 V (reverse bias)
Photodiode Bias Voltage Resolution10 mV
Photodiode Bias Voltage Accuracy±(0.5% of setting + 10 mV)
Photodiode Bias Voltage Maximum Current100 mA

A/D Converters & Sampling

ParameterValue
Number of A/D Converters3 (14-bit)
Maximum Sampling Rate10 MHz (100 ns sample interval)
Measurement Sample Interval Resolution100 ns

Analog Monitor Outputs

ParameterValue
Analog Monitor Output Voltage Range0 to 5 V (into 50 Ω load)
Analog Monitor Output Channels4 (IMON, VMON, PD1MON, PD2MON)

Protection & Interlocks

ParameterValue
External Hardware Shutdown Response Time< 1 µs
External Hardware Shutdown InputTTL (active low)

Interfaces & Connectivity

ParameterValue
Remote Test Head Cable Length2 m (6.6 ft)

Technology

Pulsed Laser Diode Testing

Applications

LIV characterization, production testing of laser diodes, and high-speed optical power measurement

Looking for the Keithley 2520?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us