The Keithley 2601B is a high-performance, single-channel System SourceMeter Source Measure Unit (SMU) designed to tightly integrate precision power sourcing and highly accurate measurement capabilities. Belonging to Keithley's industry-standard Series 2600B family, it features a dual-line vacuum fluorescent display, a fast sweep rate up to 20,000 readings per second, and embedded Test Script Processor (TSP) technology to automate complex test sequences directly on the instrument.

Specifications

General

ParameterValue
InterfacesGPIB, LAN (LXI Class C), USB 2.0 (Type B), RS-232, TSP-Link, Digital I/O (25-pin)
DisplayDual-line Vacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz, 240 VA maximum
Operating Temperature0 °C to 50 °C
Dimensions (W×H×D)213 mm × 89 mm × 460 mm
Weight4.75

Source Specifications

ParameterValue
Number of SMU Channels1
Maximum Output Power40 W (DC), 200 W (Pulse)
Maximum DC Voltage40 V
Maximum DC Current3 A @ 6 V, 1 A @ 40 V
Maximum Pulse Current10 A @ 20 V
Voltage Source Ranges100 mV, 1 V, 6 V, 40 V
Current Source Ranges100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A (Pulse only)
Voltage Source Programming Resolution5 µV (100 mV range), 50 µV (1 V range), 500 µV (6 V range), 2 mV (40 V range)
Voltage Programming Accuracy±(0.02% + 12 mV) on 40 V range
Current Programming Accuracy±(0.05% + 1.8 mA) on 1 A range
Voltage Compliance Range101 mV to 40.4 V
Current Compliance Range101 nA to 3.03 A
Voltage Source Load Regulation0.01% of range + 1 mV (0% to 100% load change)
Voltage Source Line Regulation0.01% of range (10% line change)
Current Source Load Regulation0.01% of range + 100 pA (100 nA to 1 A ranges, 0% to 100% load change)
Current Source Line Regulation0.01% of range (10% line change)
Transient Response Time< 110 µs (to recover to within 0.1% of limit for a 10% to 90% load change)
Voltage Source Settling Time< 50 µs typical (to within 0.1% of final value, 6 V range, 100 mA load)
Current Source Settling Time< 80 µs typical (to within 0.1% of final value, 1 mA range, 10 V compliance)
Voltage Source Overshoot< 0.1% typical (full scale step, resistive load)
Current Source Overshoot< 0.1% typical (full scale step, resistive load, 10 µA to 3 A ranges)

Measurement Specifications

ParameterValue
Voltage Measurement Ranges100 mV, 1 V, 6 V, 40 V
Current Measurement Ranges100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A (Pulse only)
Voltage Measurement Accuracy±(0.015% + 8 mV) on 40 V range
Current Measurement Accuracy±(0.03% + 1.5 mA) on 1 A range
Maximum Voltage Measurement Resolution100 nV
Maximum Current Measurement Resolution100 fA
Input Impedance>10 GΩ (on <6 V ranges)
Sensing Method2-wire, 4-wire (Remote Sense)

Pulse Characteristics

ParameterValue
Minimum Pulse Width100 µs
Maximum Pulse Duty Cycle10% (at 10 A range)

Dynamic and Sweep Performance

ParameterValue
Maximum Sweep Rate20,000 readings per second (to memory)
Reading Buffer Capacity>140,000 readings

Connectivity and Interfaces

ParameterValue
LXI ComplianceLXI Core 2011
Digital I/O Port Configuration25-pin D-sub, 14 configurable digital I/O lines
TSP-LinkYes, for multi-channel expansion

Software and Programming

ParameterValue
Command Set SupportTSP (Test Script Processor), SCPI (2400 emulation)
Programming LanguageLua (embedded in TSP)

Power and Environmental

ParameterValue
Operating Humidity Range0 °C to 35 °C, 70% R.H.; 35 °C to 50 °C, derate 3% R.H./°C
Storage Temperature Range-25 °C to 65 °C
Warm-Up Time2 hours

Physical and Safety

ParameterValue
Form Factor2U half-rack
Calibration Interval1 year

Electrical Specifications

ParameterValue
Maximum Common Mode Voltage250 V DC (LO to chassis)
Maximum Sense Lead Impedance3 V max voltage drop per lead to maintain accuracy
Sense Input Impedance> 10 GΩ
Guard Offset Voltage< 300 µV
Source Over-range Capability101% of source range
Measure Over-range Capability102% of measurement range

A/D Converter

ParameterValue
ADC Integration Time Range0.001 PLC to 10 PLC

Power Requirements

ParameterValue
Maximum Power Consumption240 VA
Line Frequency Range50 Hz / 60 Hz (auto-sensing)
Line Voltage Range100 V to 240 V AC (±10% auto-sensing)

General & Mechanical

ParameterValue
Cooling MethodForced air, variable speed

Compliance & Safety

ParameterValue
Product Safety StandardsUL 61010-1:2004, CAN/CSA C22.2 No. 61010-1-04, EN 61010-1:2001
EMC ComplianceConforms to European Union Directive 2004/108/EEC, EN 61326-1

System Performance

ParameterValue
Maximum Single Reading Rate20,000 readings/s (to internal memory, 0.001 PLC)
Trigger Latency< 4 µs
Trigger Jitter< 0.5 µs
Common Mode Isolation Impedance> 1 GΩ, < 450 pF typical

Interfaces & Connectivity

ParameterValue
USB Interface TypeUSB 2.0 High Speed (Type-B port for device control, Type-A port for USB flash drives)
Ethernet Port DetailsRJ-45 connector, 10Base-T/100Base-TX
TSP-Link Interface DetailsDual RJ-45 connectors for multi-chassis system expansion (up to 32 nodes)
Internal 5V Power Supply Limit500 mA maximum cumulative limit

Technology

Source Measure Unit (SMU)

Applications

Precision DC/Pulse I-V characterization, semiconductor test, component testing, and production automation