The Keithley 2612B is a dual-channel System SourceMeter (SMU) instrument designed for high-throughput production testing and precise benchtop characterization. It integrates a stable DC power source, a true current source, a 6.5-digit multimeter, an electronic load, and a pulse generator into a single compact, half-rack 2U unit.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), LAN (LXI Class C), USB 2.0 (Device), USB 2.0 (Host), RS-232, TSP-Link |
| Display | Dual-line vacuum fluorescent display (VFD) |
| Cooling | Forced air, temperature-controlled fans |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V to 240 V AC (±10%), 50 Hz to 60 Hz, 240 VA maximum |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 213 mm × 89 mm × 460 mm |
| Weight | 4.75 kg |
Source Specifications
| Parameter | Value |
|---|---|
| Number of Channels | 2 |
| Maximum Voltage Output | 200 V |
| Maximum DC Current Output | 1.5 A |
| Maximum Pulse Current Output | 10 A |
| Maximum Source Power per Channel | 30.3 W |
| Voltage Source Ranges | 200 mV, 2 V, 20 V, 200 V |
| Current Source Ranges | 100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 1.5 A, 10 A |
| Voltage Source Programming Resolution | 5 µV (on 200 mV range) |
| Current Source Programming Resolution | 2 pA (on 100 nA range) |
| Basic Voltage Source Accuracy | 0.02% |
| Basic Current Source Accuracy | 0.06% |
| Voltage Limit (Compliance) Range | 20 mV to 210 V |
| Current Limit (Compliance) Range | 10 nA to 1.575 A (up to 10.5 A for pulse) |
Measurement Specifications
| Parameter | Value |
|---|---|
| Voltage Measurement Ranges | 200 mV, 2 V, 20 V, 200 V |
| Current Measurement Ranges | 100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 1.5 A, 10 A |
| Voltage Measurement Resolution | 100 nV (6.5-digit, on 200 mV range) |
| Current Measurement Resolution | 100 fA (6.5-digit, on 100 nA range) |
| Basic Voltage Measurement Accuracy | 0.015% (1-year) |
| Basic Current Measurement Accuracy | 0.02% |
| Input Impedance (Voltage Meter) | >10 GΩ |
| Measurement Resolution | 6.5 digits |
Pulse Mode Specifications
| Parameter | Value |
|---|---|
| Minimum Pulse Width | 150 µs |
| Pulse Current Amplitude | Up to 10 A |
| Pulse Voltage Amplitude | Up to 200 V |
System & Dynamic Performance
| Parameter | Value |
|---|---|
| Maximum Sweep Rate | 20,000 points per second |
| Maximum Reading Rate to Internal Memory | 20,000 readings per second |
| Maximum Reading Rate to GPIB/USB/LAN | 20,000 readings per second |
| High Capacitance Mode | Supported (up to 50 µF max load capacitance) |
I/O and Connectivity
| Parameter | Value |
|---|---|
| Rear Panel Terminals | Phoenix-type screw terminal connectors |
| Sensing Modes | 2-wire local or 4-wire remote |
| Guard Terminals | Active Guard per channel |
| GPIB Interface | IEEE-488.2 |
| USB Front Panel Port | USB 2.0 Host (Type A) |
| USB Rear Panel Port | USB 2.0 Device (Type B) |
| TSP-Link Interface | Proprietary expansion bus, 2 RJ-45 ports |
| Digital I/O Port | 25-pin D-sub, 14 digital I/O lines |
| Interlock Connector | Rear-panel safety interlock |
Software and Control
| Parameter | Value |
|---|---|
| Command Set Compatibility | TSP (Test Script Processor), SCPI Emulation (Keithley 2400) |
| Scripting Language | Lua-based TSP scripting |
| Embedded Web Interface | LXI Web Console (Virtual Front Panel / TSP Express) |
| Software Drivers Supported | IVI-COM, IVI-C, LabVIEW, MATLAB |
Power and Operating Environment
| Parameter | Value |
|---|---|
| AC Line Voltage Range | 100 V to 240 V AC (±10%) |
| AC Line Frequency | 50 Hz or 60 Hz (auto-sensing) |
| Power Consumption | 240 VA maximum |
| Operating Temperature Range | 0 °C to 50 °C |
| Storage Temperature Range | -25 °C to 65 °C |
| Operating Humidity Range | 0% to 60% RH, up to 35 °C (derate 3% RH/°C, 35 °C to 50 °C) |
| Altitude Limits | Maximum 2000 m (operating) |
| Warm-Up Period | 2 hours (to specified accuracy) |
Physical Characteristics
| Parameter | Value |
|---|---|
| Form Factor | Bench or 2U rackmount (half-rack) |
| Rack Unit Height | 2U |
| Chassis Width | 213 mm |
| Chassis Height | 89 mm |
| Chassis Depth | 460 mm |
Compliance & Certifications
| Parameter | Value |
|---|---|
| Electromagnetic Compatibility | Conforms to European Union EMC Directive (EN 61326-1) |
| Safety Compliance | Conforms to European Union Low Voltage Directive (EN 61010-1) |
| Calibration Interval | 1 year |
System Specifications
| Parameter | Value |
|---|---|
| Maximum Common Mode Voltage | 250 V DC |
| TSP-Link Expansion Limit | Up to 32 nodes (up to 64 SMU channels) |
Performance
| Parameter | Value |
|---|---|
| Maximum Load Capacitance (Normal Mode) | 10 nF |
| Source Overrange | 101% of range |
| Measure Overrange | 102% of range |
Interfaces
| Parameter | Value |
|---|---|
| Digital I/O Lines | 14 |
| Digital I/O Power Supply Output | 5 V DC, 500 mA maximum |
Dynamic Performance
| Parameter | Value |
|---|---|
| Voltage Source Settling Time | < 50 µs (typical, to 0.1% of final value, 20V range) |
| Current Source Settling Time | < 80 µs (typical, to 0.1% of final value, 1mA range) |
Measurement Speed
| Parameter | Value |
|---|---|
| A/D Converter Integration Time Range | 0.001 PLC to 25 PLC |
Regulation
| Parameter | Value |
|---|---|
| Voltage Source Line Regulation | 0.01% of range |
| Voltage Source Load Regulation | 0.01% of range + 100 µV (for any load change) |
| Current Source Line Regulation | 0.01% of range |
| Current Source Load Regulation | 0.01% of range + 100 pA (for any change in output voltage) |
Noise Performance
| Parameter | Value |
|---|---|
| Voltage Source Noise (0.1 Hz to 10 Hz) | < 15 µV p-p (typical, 200 mV range) |
| Current Source Noise (0.1 Hz to 10 Hz) | < 5 pA p-p (typical, 100 nA range) |
Guard & Sensing
| Parameter | Value |
|---|---|
| Maximum Guard Offset Voltage | 150 µV |
| Maximum Voltage Between Guard and LO | 3 V |
| Sense Input Impedance | > 10 GΩ |
| Maximum Sense Lead Resistance Voltage Drop | 3 V per lead (for rated accuracy) |
System Buffer
| Parameter | Value |
|---|---|
| Memory Capacity | > 1,000,000 readings per channel |
System Expansion
| Parameter | Value |
|---|---|
| Maximum TSP-Link Cable Length | 3 meters (9.8 ft) |
Technology
Source Measure Unit (SMU)
Applications
Semiconductor, optoelectronic, and electronic component characterization and production testing