The Keithley Model 2700 (Integra Series) is a high-performance, 6.5-digit (22-bit) digital multimeter, data acquisition, and logging system. It features two expansion slots to accommodate up to 80 differential measurement channels, making it a highly versatile and cost-effective solution for multi-channel test applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), RS-232 |
| Display | 12-character alphanumeric VFD with 7-character secondary display |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V / 120 V / 220 V / 240 V ±10%, 45 Hz to 66 Hz and 360 Hz to 440 Hz, <28 VA |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 213 mm × 89 mm × 370 mm |
| Weight | 3.7 |
DC Voltage
| Parameter | Value |
|---|---|
| DC Voltage Ranges | 100 mV, 1 V, 10 V, 100 V, 1000 V |
| DC Voltage Resolution | 100 nV (on 100 mV range) |
| DC Voltage Input Impedance | >10 GΩ on 100 mV, 1 V, 10 V ranges; 10 MΩ ±1% on 100 V, 1000 V ranges |
| DC Voltage Best 1-Year Accuracy | ±(0.003% of reading + 0.0005% of range) |
| DC Voltage Normal Mode Rejection Ratio (NMRR) | >60 dB at 50 Hz or 60 Hz |
| DC Voltage Common Mode Rejection Ratio (CMRR) | >140 dB at DC, 50 Hz or 60 Hz (with 1 kΩ unbalance) |
| Maximum DC Input Voltage | 1000 V peak |
| DC Voltage Input Bias Current | <75 pA at 23 °C |
AC Voltage
| Parameter | Value |
|---|---|
| AC Voltage Ranges | 100 mV, 1 V, 10 V, 100 V, 750 V |
| AC Voltage Resolution | 100 nV (on 100 mV range) |
| AC Voltage Frequency Range | 3 Hz to 300 kHz |
| AC Voltage Input Impedance | 1 MΩ ±2% in parallel with <100 pF |
| AC Voltage Best 1-Year Accuracy | ±(0.06% of reading + 0.03% of range) at 10 Hz - 20 kHz |
| Maximum AC Input Voltage | 750 V RMS, 1000 V peak |
| AC Voltage Crest Factor | 5:1 at full scale |
| AC Voltage Bandwidth Filters | 3 Hz - 300 kHz, 30 Hz - 300 kHz, 300 Hz - 300 kHz |
DC Current
| Parameter | Value |
|---|---|
| DC Current Ranges | 20 mA, 100 mA, 1 A, 3 A |
| DC Current Best Resolution | 10 nA (on 20 mA range) |
| DC Current Best 1-Year Accuracy | ±(0.05% of reading + 0.005% of range) |
| DC Current Shunt Resistance | 0.1 Ω for 1 A and 3 A ranges, 5.1 Ω for 20 mA and 100 mA ranges |
| DC Current Burden Voltage | <0.3 V on 20 mA and 1 A ranges, <1.0 V on 100 mA and 3 A ranges |
| DC Current Overload Protection | 3 A, 250 V fuse |
AC Current
| Parameter | Value |
|---|---|
| AC Current Ranges | 1 A, 3 A |
| AC Current Best Resolution | 1 µA (on 1 A range) |
| AC Current Frequency Range | 3 Hz to 5 kHz |
| AC Current Shunt Resistance | 0.1 Ω |
| AC Current Burden Voltage | <0.3 V on 1 A range, <1.0 V on 3 A range |
| AC Current Best 1-Year Accuracy | ±(0.1% of reading + 0.04% of range) at 45 Hz - 2 kHz |
| AC Current Overload Protection | 3 A, 250 V fuse |
Resistance
| Parameter | Value |
|---|---|
| Resistance Ranges | 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ (2-wire and 4-wire) |
| Resistance Best Resolution | 100 µΩ (on 100 Ω range) |
| Resistance Best 1-Year Accuracy | ±(0.01% of reading + 0.001% of range) |
| Resistance Source Current | 1 mA (on 100 Ω and 1 kΩ ranges) |
| Resistance Open Circuit Voltage | <6.9 V |
| Offset Compensation | Selectable on 100 Ω, 1 kΩ, and 10 kΩ ranges |
| Maximum 4-Wire Lead Resistance | 10% of range per lead (for 100 Ω and 1 kΩ), 1 kΩ per lead (for other ranges) |
Temperature Measurement
| Parameter | Value |
|---|---|
| Supported Thermocouples | J, K, T, E, R, S, B, N |
| Thermocouple Resolution | 0.001 °C |
| Supported RTDs | PT100, D100, F100, User-defined (2-wire, 3-wire, or 4-wire) |
| Supported Thermistors | 2.2 kΩ, 5 kΩ, 10 kΩ |
| Cold Junction Compensation Accuracy | ±1.0 °C (using internal sensor on 7700 card) |
Frequency and Period
| Parameter | Value |
|---|---|
| Frequency Range | 3 Hz to 500 kHz |
| Period Range | 2 µs to 333 ms |
| Input Signal Amplitude Range | 100 mV RMS to 750 V RMS |
| Frequency Gate Times | 10 ms, 100 ms, 1 s |
| Frequency Best 1-Year Accuracy | ±0.01% of reading |
Data Acquisition and Scanning
| Parameter | Value |
|---|---|
| Max Scanning Speed | Up to 180 channels/s (for DCV) |
| Memory Buffer Capacity | 55,000 readings |
| Triggering Sources | Manual, External Trigger, GPIB, RS-232, Trigger Link, Immediate |
| Trigger Delay Range | 0 to 99 hours (1 ms resolution) |
| Channel-to-Channel Crosstalk | <-40 dB at 10 MHz (module dependent) |
| Max Voltage Between Channels | 300 V DC or RMS (module dependent) |
Expansion Slots and Modules
| Parameter | Value |
|---|---|
| Number of Expansion Slots | 2 slots (rear panel) |
| Compatible Modules | 7700, 7701, 7702, 7703, 7705, 7706, 7707, 7708, 7709, 7710, 7711, 7712 |
| Max System Channels | 80 differential channels (with two 40-channel multiplexer cards) |
Connectivity and Interfaces
| Parameter | Value |
|---|---|
| Digital I/O Lines | 4 outputs (limit/alarm, TTL compatible), 1 input (trigger) |
| Trigger Link Connection | 8-pin micro-DIN |
| Command Languages | SCPI (Standard Commands for Programmable Instruments) |
Environmental and Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to 70 °C |
| Operating Humidity Range | Up to 80% RH at 35 °C (non-condensing) |
| Warm-up Time | 2 hours to rated accuracy |
| Safety Compliance | Conforms to European Union Low Voltage Directive (EN61010-1) |
| EMC Compliance | Conforms to European Union EMC Directive (EN61326-1) |
Power
| Parameter | Value |
|---|---|
| Power Consumption | 28 VA |
| Power Line Frequency | 45 Hz to 66 Hz and 360 Hz to 440 Hz (automatically sensed) |
System & Memory
| Parameter | Value |
|---|---|
| User Setup Storage | 5 setups (1 factory default, 4 user-defined) in non-volatile memory |
| Real-Time Clock | Lithium battery-backed, 10-year typical life |
Measurement & Math
| Parameter | Value |
|---|---|
| Math Functions | mX+b, Percent (%), Reciprocal (1/X), dB, dBm, Limit Testing |
| dBm Reference Resistance | 1 Ω to 9999 Ω |
Diode Test
| Parameter | Value |
|---|---|
| Diode Test Current | 1 mA |
| Diode Test Voltage Range | 3 V |
Continuity Test
| Parameter | Value |
|---|---|
| Continuity Threshold Range | 1 Ω to 1000 Ω |
| Continuity Test Current | 1 mA |
Measurement Rates
| Parameter | Value |
|---|---|
| Max Reading Rate to Internal Memory | 2000 rdgs/s |
| Max Reading Rate to GPIB | 375 rdgs/s |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Maximum Common Mode Voltage | 500 V peak |
| Digital Output Sink Current | 100 mA (max. per output) |
| Digital Input Type | TTL compatible |
| Input Connection Locations | Front and rear selectable |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 2000 m |
Frequency Measurement
| Parameter | Value |
|---|---|
| Frequency Input Impedance | 1 MΩ ±2% in parallel with <150 pF |
Technology
Digital Multimeter and Data Acquisition
Applications
Multi-channel temperature profiling, precision data logging, manufacturing test, and automated device characterization