The Keithley 617 is a high-performance programmable electrometer designed for highly sensitive measurements of voltage, current, resistance, and charge. Featuring a 4½-digit LED display, an integrated isolated 100V voltage source, and standard IEEE-488 (GPIB) connectivity, the Keithley 617 delivers the exceptional input impedance and low input bias current required for sensitive research and calibration environments.

Specifications

General

ParameterValue
InterfacesIEEE-488 (GPIB)
Display4½-digit LED with 2-digit exponent
Warm-up Time1 hour to rated accuracy
Standard AccessoriesModel 6011 triaxial input cable

Physical

ParameterValue
Power Supply90–125 V AC / 180–250 V AC, 50–60 Hz, 35 VA max
Operating Temperature0°C to 50°C
Dimensions (W×H×D)216 mm × 127 mm × 359 mm
Weight2.9 kg

General Measurement Capabilities

ParameterValue
Measurement FunctionsDC Voltage, DC Current, Resistance (constant-current and constant-voltage), Charge
Display TypeLED, 4½-digit mantissa plus 2-digit exponent
Ranging ModesAutomatic and Manual
Data Store Buffer100 readings
CalibrationDigital calibration from front panel or IEEE-488 bus
Zero CorrectFront panel zero correct control
Baseline SuppressionOne-button suppression from front panel or IEEE-488 bus
TriggeringOne-shot front panel or bus control

DC Voltage Measurements (Voltmeter)

ParameterValue
Voltage Measurement Range200 mV to 200 V
200mV Range Resolution10 µV
2V Range Resolution100 µV
20V Range Resolution1 mV
200V Range Resolution10 mV
Input Impedance>200 TΩ
Input Bias Current<5 fA at 23°C
Common Mode Rejection Ratio (CMRR)>120 dB
Normal Mode Rejection Ratio (NMRR)>60 dB at 50 Hz or 60 Hz

DC Current Measurements (Ammeter)

ParameterValue
Current Measurement Range2 pA to 20 mA
2pA Range Resolution100 aA
20pA Range Resolution1 fA
200pA Range Resolution10 fA
2nA Range Resolution100 fA
20nA Range Resolution1 pA
200nA Range Resolution10 pA
2µA Range Resolution100 pA
20µA Range Resolution1 nA
200µA Range Resolution10 nA
2mA Range Resolution100 nA
20mA Range Resolution1 µA
Input Voltage Burden<1 mV (<3 mV on 20 mA range)

Resistance Measurements (Ohmmeter)

ParameterValue
Resistance Measurement Range2 kΩ to 200 GΩ (up to 10^16 Ω with V-source)
2kΩ Range Resolution100 mΩ
20kΩ Range Resolution1 Ω
200kΩ Range Resolution10 Ω
2MΩ Range Resolution100 Ω
20MΩ Range Resolution1 kΩ
200MΩ Range Resolution10 kΩ
2GΩ Range Resolution100 kΩ
20GΩ Range Resolution1 MΩ
200GΩ Range Resolution10 MΩ
Maximum Open Circuit Voltage250 V DC
Resistance Measurement MethodsConstant current and constant voltage

Charge Measurements (Coulombmeter)

ParameterValue
Charge Measurement Range200 pC to 20 nC
200pC Range Resolution10 fC
2nC Range Resolution100 fC
20nC Range Resolution1 pC

Internal Voltage Source (V-Source)

ParameterValue
V-Source Output Voltage Range-102 V to +102 V
V-Source Output Resolution5 mV steps on 10V range, 50 mV steps on 100V range
V-Source ConfigurationIsolated from electrometer section

Analog & Auxiliary Outputs

ParameterValue
Analog OutputsPreamplifier output and 2V full range output on rear panel

System Interfaces & Connectivity

ParameterValue
Guard ConfigurationSelectable driven cable guard
Input Connector Type2-slot female triaxial

Power and Environmental Specifications

ParameterValue
Power Consumption35 VA maximum
Line Voltage Selection90–125 V AC or 180–250 V AC (rear panel switch)
Line Frequency50 Hz to 60 Hz
Operating Humidity Range<70% relative humidity up to 35°C
Storage Temperature Range-25°C to 65°C

Voltage Source

ParameterValue
Voltage Source Max Current2 mA (short circuit protected)
Voltage Source Accuracy (10V Range)±(0.15% + 10 mV)
Voltage Source Accuracy (100V Range)±(0.15% + 100 mV)
Voltage Source Temperature Coefficient±(0.005% + 100 µV)/°C on 10V range, ±(0.005% + 1 mV)/°C on 100V range
Voltage Source Output Resistance< 0.1 Ω

Protection

ParameterValue
Overload Protection (Volts)250 V peak, DC to 60 Hz continuous
Overload Protection (Amps)100 V peak on 2 pA to 200 nA ranges, 250 V peak on 2 µA to 2 mA ranges
Overload Protection (Ohms)250 V peak, DC to 60 Hz continuous
Overload Protection (Coulombs)100 V peak, DC to 60 Hz continuous

Outputs

ParameterValue
Preamp Output Voltage1 V for full scale input (inverting)
Preamp Output Impedance1 kΩ
2V Analog Output Voltage2 V for full scale display (non-inverting)
2V Analog Output Impedance10 kΩ

Interfaces

ParameterValue
GPIB Interface FunctionsSH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1

Performance

ParameterValue
Maximum Reading Rate18 readings per second (via GPIB)

Voltage Measurement

ParameterValue
Voltage Measurement Accuracy (200mV Range)±(0.15% of reading + 4 counts), 1 year (18°C-28°C)
Voltage Measurement Accuracy (2V to 200V Ranges)±(0.05% of reading + 1 count), 1 year (18°C-28°C)
Input Capacitance< 20 pF

Current Measurement

ParameterValue
Current Measurement Accuracy (2pA Range)±(1.6% of reading + 6 counts), 1 year (18°C-28°C)
Current Measurement Accuracy (2nA to 200nA Ranges)±(0.25% of reading + 5 counts), 1 year (18°C-28°C)
Current Measurement Accuracy (2µA to 20mA Ranges)±(0.15% of reading + 5 counts), 1 year (18°C-28°C)

Resistance Measurement

ParameterValue
Resistance Measurement Accuracy (2kΩ Range)±(0.2% of reading + 4 counts), 1 year (18°C-28°C)
Resistance Measurement Accuracy (20kΩ to 20MΩ Ranges)±(0.2% of reading + 1 count), 1 year (18°C-28°C)
Resistance Measurement Accuracy (20GΩ Range)±(1.5% of reading + 1 count)
Resistance Measurement Accuracy (200GΩ Range)±(2.0% of reading + 1 count)

Charge Measurement

ParameterValue
Charge Measurement Accuracy (200pC Range)±(0.4% of reading + 4 counts), 1 year (18°C-28°C)
Charge Measurement Accuracy (2nC to 20nC Ranges)±(0.4% of reading + 1 count), 1 year (18°C-28°C)

Rear Panel Connections

ParameterValue
GPIB Address ConfigurationSelectable via rear panel DIP switches
External Trigger InputBNC connector, TTL-compatible, active low
Meter Complete OutputBNC connector, TTL-compatible, active low pulse
Voltage Source Output ConnectorSafety BNC (red)

Technology

Solid-State Electrometer

Applications

Low-level current, high resistance, voltage, and charge measurements