The Keithley DMM7510 is a premium 7½-digit graphical sampling digital multimeter that integrates a high-accuracy instrument with a high-speed 18-bit digitizer and a 5-inch capacitive touchscreen display. Designed for precision laboratory measurements and rapid transient waveform analysis, it represents the state of the art in high-performance test instrumentation.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB, LAN (LXI Core), USB Device (Type B), USB Host (Type A), Digital I/O (TSP-Link) |
| Display | 5-inch color TFT touchscreen with WVGA (800 × 480) resolution |
| Real-Time Clock | Yes, with battery backup |
| Front/Rear Input Selector Switch | Manual front-panel toggle |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V to 240 V AC ±10%, 50 Hz/60 Hz/400 Hz (auto-sensing), 55 VA maximum |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 224 mm × 106 mm × 410 mm (with bumpers and handle) / 213 mm × 88 mm × 360 mm (without bumpers and handle) |
| Weight | 4.08 kg (without bumpers) / 4.36 kg (with bumpers) |
DC Voltage Specifications
| Parameter | Value |
|---|---|
| DCV Measurement Ranges | 100 mV, 1 V, 10 V, 100 V, 1000 V |
| DCV Best Resolution | 10 nV |
| DCV Input Impedance (100mV to 10V) | >10 GΩ or 10 MΩ selectable |
| DCV Input Impedance (100V to 1000V) | 10 MΩ ±1% |
| DCV Accuracy (90-Day, 10 V range) | ±(11 ppm of reading + 1.5 ppm of range) |
| DCV Accuracy (1-Year, 10 V range) | ±(14 ppm of reading + 1.5 ppm of range) |
| DCV Temp Coefficient (0°C–18°C & 28°C–50°C) | ±(0.15 ppm of reading + 0.08 ppm of range) per °C (10 V range) |
| Common Mode Rejection Ratio (CMRR) | 140 dB (DC, 50 Hz/60 Hz, 1 kΩ unbalance in LO lead) |
| Normal Mode Rejection Ratio (NMRR) | 60 dB (at 50 Hz/60 Hz, with line sync enabled) |
AC Voltage Specifications
| Parameter | Value |
|---|---|
| ACV Measurement Ranges | 100 mV, 1 V, 10 V, 100 V, 700 V |
| ACV Best Resolution | 100 nV |
| ACV Frequency Range | 3 Hz to 300 kHz |
| ACV Accuracy (1-Year, 10 Hz - 20 kHz) | ±(0.06% of reading + 0.01% of range) |
| ACV Accuracy (1-Year, 20 kHz - 100 kHz) | ±(0.12% of reading + 0.05% of range) on 10 V range |
| ACV Input Impedance | 1 MΩ ±1% in parallel with <100 pF |
| ACV Maximum Crest Factor | 5 at full scale |
| AC Common Mode Rejection Ratio | 70 dB (for 50 Hz/60 Hz, 1 kΩ unbalance in LO lead) |
DC Current Specifications
| Parameter | Value |
|---|---|
| DCI Measurement Ranges | 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A |
| DCI Best Resolution | 1 pA |
| DCI Accuracy (90-Day, 10 mA range) | ±(100 ppm of reading + 20 ppm of range) |
| DCI Accuracy (1-Year, 10 mA range) | ±(150 ppm of reading + 25 ppm of range) |
| DCI Max Burden Voltage (1 A) | <300 mV |
| DCI Max Burden Voltage (3 A) | <1.5 V |
| DCI Max Burden Voltage (10 A) | <0.5 V |
AC Current Specifications
| Parameter | Value |
|---|---|
| ACI Measurement Ranges | 105 mA, 10 mA, 100 mA, 1 A, 3 A |
| ACI Best Resolution | 1 nA |
| ACI Frequency Range | 3 Hz to 10 kHz |
| ACI Accuracy (1-Year, 10 Hz - 10 kHz) | ±(0.10% of reading + 0.04% of range) on 100 mA range |
Resistance Specifications
| Parameter | Value |
|---|---|
| Resistance Measurement Ranges | 10 Ω, 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 1 GΩ |
| Resistance Best Resolution | 1 µΩ |
| Resistance Accuracy (1-Year, 10 kΩ range) | ±(24 ppm of reading + 3 ppm of range) |
| Maximum Test Current | 10 mA |
| Offset Compensation | Selectable on 10 Ω to 100 kΩ ranges |
| Dry Circuit Mode Support | Yes (20 mV clamp, available on 100 Ω, 1 kΩ, and 10 kΩ ranges) |
Digitizer Specifications
| Parameter | Value |
|---|---|
| Digitizer Resolution | 18-bit |
| Maximum Sampling Rate | 1 MS/s |
| Analog Bandwidth (-3dB) | 1 MHz |
| Aperture Range | 1 µs to 1 s |
| Digitizer Memory | Up to 27.5 million readings (compact), 11 million readings (standard) |
Temperature Measurements
| Parameter | Value |
|---|---|
| Supported RTD Types | PT100, D100, F100, PT385, PT3916, and user-defined |
| Supported Thermocouple Types | J, K, T, E, R, S, B, N |
| Supported Thermistor Types | 2.2 kΩ, 5 kΩ, and 10 kΩ |
Triggering and Synchronization
| Parameter | Value |
|---|---|
| Trigger Sources | Hardware, Software (Bus), Timer, Analog Threshold, External Trigger, TSP-Link |
| External Trigger Input Level | TTL compatible (nominal 0 V to 5 V, Vil < 0.8 V, Vih > 2.0 V) |
| External Trigger Output Level | TTL compatible |
| Digital I/O Ports | 6 dual-purpose digital I/O lines (input or output, with programmable limit comparisons) |
System and Interface Specifications
| Parameter | Value |
|---|---|
| LXI Compliance Version | LXI version 1.4 Core 2011 |
| Programming Languages / Command Sets | SCPI, TSP (Test Script Processor), Keithley 2001 emulation |
| Internal Reading Buffer Capacity | Up to 27.5 million readings (compact), 11 million readings (standard) |
| USB Host Port Functionality | Saves screen captures, instrument settings, and reading data to USB flash drives |
Math and Analytics Functions
| Parameter | Value |
|---|---|
| Built-in Math Operations | mX+b, Percent, Reciprocal (1/X), Ratio, Delta |
| Graphing and Trend Plotting | Waveform display, swipe-to-zoom, trend charts, histogram |
| Limit Testing Configuration | Dual limits with high/low pass/fail indication and digital I/O port mapping |
Power and Environmental Specifications
| Parameter | Value |
|---|---|
| Line Voltage Input Ranges | 100 V / 120 V / 220 V / 240 V AC ±10% (auto-sensing) |
| Line Frequency Range | 50 Hz, 60 Hz, and 400 Hz (auto-sensing) |
| Maximum Power Consumption | 55 VA |
| Maximum Operating Altitude | 2000 meters |
| Relative Humidity Limits | Up to 80% R.H. @ 35°C, non-condensing |
| Warm-up Time | 90 minutes |
Physical and Safety Specifications
| Parameter | Value |
|---|---|
| Safety Compliance Certifications | UL 61010-1, CAN/CSA-C22.2 No. 61010-1, EN 61010-1 |
| Electromagnetic Compatibility (EMC) Standards | Conforms to European Union EMC Directive, EN 61326-1 |
| Warranty Period | 3 years |
Capacitance
| Parameter | Value |
|---|---|
| Capacitance Measurement Ranges | Not supported |
| Capacitance Best Resolution | 0.1 pF |
| Capacitance Accuracy (1-Year) | 0.5% of reading + 0.5% of range |
Frequency & Period
| Parameter | Value |
|---|---|
| Frequency Measurement Range | 3 Hz to 1 MHz |
| Period Measurement Range | 333 ms to 1 µs |
| Frequency Accuracy (1-Year) | 0.01% of reading |
Diode & Continuity
| Parameter | Value |
|---|---|
| Diode Test Voltage Range | 10 V |
| Diode Test Current Options | 10 µA, 100 µA, 1 mA |
| Continuity Threshold Range | 1 Ω to 1000 Ω |
Temperature Measurement
| Parameter | Value |
|---|---|
| RTD Temperature Measurement Range | -200 °C to +850 °C |
| Thermocouple Temperature Measurement Range | -200 °C to +1820 °C (sensor type dependent) |
| Thermistor Temperature Measurement Range | -80 °C to +150 °C |
Digitizer
| Parameter | Value |
|---|---|
| Digitizer Voltage Ranges | 100 mV, 1 V, 10 V, 100 V, 1000 V |
| Digitizer Current Ranges | 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A |
Input Protection & Safety
| Parameter | Value |
|---|---|
| Maximum Common Mode Voltage | 500 V peak (LO to chassis ground) |
| Voltage Input Overload Protection | 1000 V peak (continuous) |
| Current Input Protection | 11 A, 1000 V fast-acting fuse |
Resolution & Display
| Parameter | Value |
|---|---|
| Digits of Resolution | 3.5 to 7.5 digits |
| Display Resolution | 800 × 480 pixels |
| Touchscreen Technology | Capacitive multi-touch |
DC Voltage
| Parameter | Value |
|---|---|
| DCV Input Bias Current | < 30 pA @ 23°C |
Protection
| Parameter | Value |
|---|---|
| Sense HI to Sense LO Overload Protection | 200 V peak |
| Sense LO to Input LO Overload Protection | 150 V peak |
| Front Current Input Fuse | 3 A, 1000 V (fast-acting, user-replaceable) |
| Rear Current Input Fuse | 11 A, 1000 V (fast-acting) |
Measurement Speed
| Parameter | Value |
|---|---|
| NPLC Range | 0.0005 to 12 PLC (@ 60 Hz) / 0.0005 to 15 PLC (@ 50 Hz) |
| Maximum Reading Rate (4.5 Digits) | 26,100 rdgs/s (at 0.0005 PLC to memory) |
| Maximum Reading Rate (7.5 Digits) | 60 rdgs/s (at 1 PLC @ 60 Hz to memory) |
Connectivity & Integration
| Parameter | Value |
|---|---|
| TSP-Link Support | Yes |
Environmental
| Parameter | Value |
|---|---|
| Chassis Vibration Standard | MIL-PRF-28800F Class 3 Random |
Technology
Solid State Analog-to-Digital Conversion with Multi-slope and SAR Architectures
Applications
Precision DC/AC metrology, high-speed transient analysis, components testing, and automated system integration