The Keysight HP 41420A is a High Power Source/Monitor Unit (HPSMU) plug-in module designed for the HP 4142B Modular DC Source/Monitor. It is capable of forcing and measuring up to +/-200 V or +/-1 A with a maximum power output of 14 W. It features robust pulsed and constant measurement capabilities with advanced compliance limits.

Specifications

Physical

ParameterValue
Operating Temperature5 °C to 40 °C
Weight2.2 kg
Mainframe Slots Occupied2 slots

General

ParameterValue
Module TypeHigh Power Source/Monitor Unit (HPSMU)
Compatible MainframeHP 4142B Modular DC Source/Monitor

Output Specifications

ParameterValue
Maximum Output Voltage+/-200 V
Maximum Output Current+/-1 A
Maximum Output Power14 W
Maximum Current (2 V range)+/-1 A
Maximum Current (20 V range)+/-1 A, +/-700 mA
Maximum Current (40 V range)+/-350 mA
Maximum Current (100 V range)+/-125 mA
Maximum Current (200 V range)+/-50 mA
Maximum Voltage (1 nA to 100 mA current ranges)10 nA to 100 mA current ranges
Maximum Voltage (1 A current range)+/-200 V, +/-100 V, +/-40 V, +/-20 V, +/-14 V

Operating Modes & Advanced Features

ParameterValue
Standard Operating ModesV source / I monitor, I source / V monitor
Advanced Operating ModesV source / V monitor, I source / I monitor (High speed spot, Analog search, Quasi-pulsed spot)
Compliance LimitationsPrevents damage by specifying I compliance for V source or V compliance for I source
Compliance ResolutionSame resolution as the output voltage or current

Pulse Parameters

ParameterValue
Pulse Width Range1 ms to 50 ms
Pulse Width Resolution100 µs
Pulse Period Range10 ms to 500 ms
Pulse Period Resolution100 µs
Maximum Pulse Duty Cycle50%
Pulse Measurement Range (2 V output range)10 nA to 1 A
Pulse Measurement Range (20 V to 200 V output range)100 µA to 1 A
Pulse Output Range (10 nA to 10 µA measurement range)2 V
Pulse Output Range (100 µA to 1 A measurement range)2 V to 200 V

Electrical Specifications

ParameterValue
Voltage Source Ranges±2 V, ±20 V, ±40 V, ±100 V, ±200 V
Current Source Ranges±10 nA, ±100 nA, ±1 µA, ±10 µA, ±100 µA, ±1 mA, ±10 mA, ±100 mA, ±1 A
Voltage Measurement Ranges±2 V, ±20 V, ±40 V, ±100 V, ±200 V
Current Measurement Ranges±10 nA, ±100 nA, ±1 µA, ±10 µA, ±100 µA, ±1 mA, ±10 mA, ±100 mA, ±1 A

Resolution

ParameterValue
Minimum Current Source Resolution20 pA
Minimum Voltage Source Resolution100 µV
Minimum Current Measurement Resolution100 fA (using High-Resolution ADC)
Minimum Voltage Measurement Resolution20 µV (using High-Resolution ADC)

Connectivity

ParameterValue
Output TerminalsDual triaxial connectors (Force and Sense)
Sensing Configuration4-wire Kelvin connection

Performance

ParameterValue
Basic Voltage Measurement Accuracy±0.05%
Basic Current Measurement Accuracy±0.2%
Integration Time SettingsShort / Medium / Long
Maximum Load Capacitance1000 pF
Maximum Guard Capacitance1000 pF

Technology

Modular Source/Monitor Unit (SMU)

Applications

Semiconductor and electronic component DC parametric testing