The HP 41421B is a 1-slot Medium Power Source/Monitor Unit (MPSMU) module designed for the HP 4142B Modular DC Source/Monitor mainframe. It offers precision DC voltage/current sourcing and measurement capabilities up to ±100 V and ±100 mA (max 2 W), and supports pulsed measurements with high resolution.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | None (internal mainframe interface; HP 4142B mainframe provides GPIB) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Power provided by HP 4142B mainframe backplane |
| Operating Temperature | 5 to 40 |
| Weight | 1.0 |
Module Overview
| Parameter | Value |
|---|---|
| Module Type | Medium Power Source/Monitor Unit (MPSMU) |
| Compatible Mainframes | HP 4142B Modular DC Source/Monitor |
| Required Mainframe Slot Count | 1 slot |
| Number of Channels | 1 |
| Maximum Output Power | 2 W |
| Maximum Output Voltage | ±100 V |
| Maximum Output Current | ±100 mA |
| Primary Operating Modes | V source (constant or pulse) and I monitor, I source (constant or pulse) and V monitor |
| Special Measurement Modes | V source (constant only) and V monitor, I source (constant only) and I monitor (in High speed spot, Analog search, and Quasi-pulsed spot measurements) |
| Compliance Feature | Limits output voltage (V compliance) or current (I compliance) to prevent device damage |
| Compliance Resolution | Same resolution as the output voltage or current within the maximum output range |
Voltage Source Specifications
| Parameter | Value |
|---|---|
| Voltage Source Range (2 V) | Maximum output ±100 mA |
| Voltage Source Range (20 V) | Maximum output ±100 mA |
| Voltage Source Range (40 V) | Maximum output ±50 mA |
| Voltage Source Range (100 V) | Maximum output ±20 mA |
Current Source Specifications
| Parameter | Value |
|---|---|
| Current Source Range (1 nA to 10 mA) | Maximum output ±100 V |
| Current Source Range (100 mA) | Maximum output ±20 V |
Timing & Pulse Mode Specifications
| Parameter | Value |
|---|---|
| Pulse Width Range | 1 ms to 50 ms |
| Pulse Width Resolution | 100 µs |
| Pulse Period Range | 10 ms to 500 ms |
| Pulse Period Resolution | 100 µs |
| Maximum Pulse Duty Cycle | 50% |
| Pulse Current Polarity Restriction | Pulse base current and pulse current must have the same polarity |
Front Panel Connectors
| Parameter | Value |
|---|---|
| Force Connector | Triaxial |
| Sense Connector | Triaxial |
Voltage Source Resolution
| Parameter | Value |
|---|---|
| Voltage Source Resolution (2 V Range) | 100 µV |
| Voltage Source Resolution (20 V Range) | 1 mV |
| Voltage Source Resolution (40 V Range) | 2 mV |
| Voltage Source Resolution (100 V Range) | 5 mV |
Current Source Resolution
| Parameter | Value |
|---|---|
| Current Source Resolution (1 nA Range) | 100 fA |
| Current Source Resolution (10 nA Range) | 1 pA |
| Current Source Resolution (100 nA Range) | 10 pA |
| Current Source Resolution (1 µA Range) | 100 pA |
| Current Source Resolution (10 µA Range) | 1 nA |
| Current Source Resolution (100 µA Range) | 10 nA |
| Current Source Resolution (1 mA Range) | 100 nA |
| Current Source Resolution (10 mA Range) | 1 µA |
| Current Source Resolution (100 mA Range) | 10 µA |
Voltage Measurement Resolution
| Parameter | Value |
|---|---|
| Voltage Measurement Resolution (2 V Range) | 20 µV (High-Resolution ADC) |
| Voltage Measurement Resolution (20 V Range) | 200 µV (High-Resolution ADC) |
| Voltage Measurement Resolution (40 V Range) | 400 µV (High-Resolution ADC) |
| Voltage Measurement Resolution (100 V Range) | 1 mV (High-Resolution ADC) |
Current Measurement Resolution
| Parameter | Value |
|---|---|
| Current Measurement Resolution (1 nA Range) | 20 fA (High-Resolution ADC) |
| Current Measurement Resolution (10 nA Range) | 100 fA (High-Resolution ADC) |
| Current Measurement Resolution (100 nA Range) | 1 pA (High-Resolution ADC) |
| Current Measurement Resolution (1 µA Range) | 10 pA (High-Resolution ADC) |
| Current Measurement Resolution (10 µA Range) | 100 pA (High-Resolution ADC) |
| Current Measurement Resolution (100 µA Range) | 1 nA (High-Resolution ADC) |
| Current Measurement Resolution (1 mA Range) | 10 nA (High-Resolution ADC) |
| Current Measurement Resolution (10 mA Range) | 100 nA (High-Resolution ADC) |
| Current Measurement Resolution (100 mA Range) | 1 µA (High-Resolution ADC) |
Environmental
| Parameter | Value |
|---|---|
| Operating Humidity | 5% to 80% RH (non-condensing) |
| Storage Temperature | -40 °C to +70 °C |
| Storage Humidity | 5% to 90% RH (non-condensing) |
Accuracy (Voltage Source)
| Parameter | Value |
|---|---|
| Voltage Source Accuracy (2 V Range) | ±(0.05% of setting + 400 µV) |
| Voltage Source Accuracy (20 V Range) | ±(0.05% of setting + 3 mV) |
| Voltage Source Accuracy (40 V Range) | ±(0.05% of setting + 6 mV) |
| Voltage Source Accuracy (100 V Range) | ±(0.05% of setting + 15 mV) |
Accuracy (Voltage Measurement)
| Parameter | Value |
|---|---|
| Voltage Measurement Accuracy (2 V Range) | ±(0.05% of reading + 400 µV) |
| Voltage Measurement Accuracy (20 V Range) | ±(0.05% of reading + 3 mV) |
| Voltage Measurement Accuracy (40 V Range) | ±(0.05% of reading + 6 mV) |
| Voltage Measurement Accuracy (100 V Range) | ±(0.05% of reading + 15 mV) |
Accuracy (Current Source)
| Parameter | Value |
|---|---|
| Current Source Accuracy (1 nA Range) | ±(0.5% of setting + 3 pA + 10 fA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (10 nA Range) | ±(0.25% of setting + 15 pA + 100 fA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (100 nA Range) | ±(0.25% of setting + 100 pA + 1 pA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (1 µA Range) | ±(0.2% of setting + 1 nA + 10 pA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (10 µA Range) | ±(0.2% of setting + 10 nA + 100 pA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (100 µA Range) | ±(0.2% of setting + 100 nA + 1 nA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (1 mA Range) | ±(0.2% of setting + 1 µA + 10 nA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (10 mA Range) | ±(0.2% of setting + 10 µA + 100 nA × Vo) [Vo: output voltage in Volts] |
| Current Source Accuracy (100 mA Range) | ±(0.2% of setting + 100 µA + 1 µA × Vo) [Vo: output voltage in Volts] |
Accuracy (Current Measurement)
| Parameter | Value |
|---|---|
| Current Measurement Accuracy (1 nA Range) | ±(0.5% of reading + 3 pA + 10 fA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (10 nA Range) | ±(0.25% of reading + 15 pA + 100 fA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (100 nA Range) | ±(0.25% of reading + 100 pA + 1 pA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (1 µA Range) | ±(0.2% of reading + 1 nA + 10 pA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (10 µA Range) | ±(0.2% of reading + 10 nA + 100 pA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (100 µA Range) | ±(0.2% of reading + 100 nA + 1 nA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (1 mA Range) | ±(0.2% of reading + 1 µA + 10 nA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (10 mA Range) | ±(0.2% of reading + 10 µA + 100 nA × Vm) [Vm: measurement voltage in Volts] |
| Current Measurement Accuracy (100 mA Range) | ±(0.2% of reading + 100 µA + 1 µA × Vm) [Vm: measurement voltage in Volts] |
Guard Specifications
| Parameter | Value |
|---|---|
| Guard Offset Voltage | ±2 mV max |
| Maximum Guard Capacitance | 100 pF |
Safety & Limits
| Parameter | Value |
|---|---|
| Maximum Common Mode Voltage | ±42 V |
Technology
Source/Monitor Unit (SMU)
Applications
Semiconductor parameter analysis and DC characterization