The HP 41421B is a 1-slot Medium Power Source/Monitor Unit (MPSMU) module designed for the HP 4142B Modular DC Source/Monitor mainframe. It offers precision DC voltage/current sourcing and measurement capabilities up to ±100 V and ±100 mA (max 2 W), and supports pulsed measurements with high resolution.

Specifications

General

ParameterValue
InterfacesNone (internal mainframe interface; HP 4142B mainframe provides GPIB)

Physical

ParameterValue
Power SupplyPower provided by HP 4142B mainframe backplane
Operating Temperature5 to 40
Weight1.0

Module Overview

ParameterValue
Module TypeMedium Power Source/Monitor Unit (MPSMU)
Compatible MainframesHP 4142B Modular DC Source/Monitor
Required Mainframe Slot Count1 slot
Number of Channels1
Maximum Output Power2 W
Maximum Output Voltage±100 V
Maximum Output Current±100 mA
Primary Operating ModesV source (constant or pulse) and I monitor, I source (constant or pulse) and V monitor
Special Measurement ModesV source (constant only) and V monitor, I source (constant only) and I monitor (in High speed spot, Analog search, and Quasi-pulsed spot measurements)
Compliance FeatureLimits output voltage (V compliance) or current (I compliance) to prevent device damage
Compliance ResolutionSame resolution as the output voltage or current within the maximum output range

Voltage Source Specifications

ParameterValue
Voltage Source Range (2 V)Maximum output ±100 mA
Voltage Source Range (20 V)Maximum output ±100 mA
Voltage Source Range (40 V)Maximum output ±50 mA
Voltage Source Range (100 V)Maximum output ±20 mA

Current Source Specifications

ParameterValue
Current Source Range (1 nA to 10 mA)Maximum output ±100 V
Current Source Range (100 mA)Maximum output ±20 V

Timing & Pulse Mode Specifications

ParameterValue
Pulse Width Range1 ms to 50 ms
Pulse Width Resolution100 µs
Pulse Period Range10 ms to 500 ms
Pulse Period Resolution100 µs
Maximum Pulse Duty Cycle50%
Pulse Current Polarity RestrictionPulse base current and pulse current must have the same polarity

Front Panel Connectors

ParameterValue
Force ConnectorTriaxial
Sense ConnectorTriaxial

Voltage Source Resolution

ParameterValue
Voltage Source Resolution (2 V Range)100 µV
Voltage Source Resolution (20 V Range)1 mV
Voltage Source Resolution (40 V Range)2 mV
Voltage Source Resolution (100 V Range)5 mV

Current Source Resolution

ParameterValue
Current Source Resolution (1 nA Range)100 fA
Current Source Resolution (10 nA Range)1 pA
Current Source Resolution (100 nA Range)10 pA
Current Source Resolution (1 µA Range)100 pA
Current Source Resolution (10 µA Range)1 nA
Current Source Resolution (100 µA Range)10 nA
Current Source Resolution (1 mA Range)100 nA
Current Source Resolution (10 mA Range)1 µA
Current Source Resolution (100 mA Range)10 µA

Voltage Measurement Resolution

ParameterValue
Voltage Measurement Resolution (2 V Range)20 µV (High-Resolution ADC)
Voltage Measurement Resolution (20 V Range)200 µV (High-Resolution ADC)
Voltage Measurement Resolution (40 V Range)400 µV (High-Resolution ADC)
Voltage Measurement Resolution (100 V Range)1 mV (High-Resolution ADC)

Current Measurement Resolution

ParameterValue
Current Measurement Resolution (1 nA Range)20 fA (High-Resolution ADC)
Current Measurement Resolution (10 nA Range)100 fA (High-Resolution ADC)
Current Measurement Resolution (100 nA Range)1 pA (High-Resolution ADC)
Current Measurement Resolution (1 µA Range)10 pA (High-Resolution ADC)
Current Measurement Resolution (10 µA Range)100 pA (High-Resolution ADC)
Current Measurement Resolution (100 µA Range)1 nA (High-Resolution ADC)
Current Measurement Resolution (1 mA Range)10 nA (High-Resolution ADC)
Current Measurement Resolution (10 mA Range)100 nA (High-Resolution ADC)
Current Measurement Resolution (100 mA Range)1 µA (High-Resolution ADC)

Environmental

ParameterValue
Operating Humidity5% to 80% RH (non-condensing)
Storage Temperature-40 °C to +70 °C
Storage Humidity5% to 90% RH (non-condensing)

Accuracy (Voltage Source)

ParameterValue
Voltage Source Accuracy (2 V Range)±(0.05% of setting + 400 µV)
Voltage Source Accuracy (20 V Range)±(0.05% of setting + 3 mV)
Voltage Source Accuracy (40 V Range)±(0.05% of setting + 6 mV)
Voltage Source Accuracy (100 V Range)±(0.05% of setting + 15 mV)

Accuracy (Voltage Measurement)

ParameterValue
Voltage Measurement Accuracy (2 V Range)±(0.05% of reading + 400 µV)
Voltage Measurement Accuracy (20 V Range)±(0.05% of reading + 3 mV)
Voltage Measurement Accuracy (40 V Range)±(0.05% of reading + 6 mV)
Voltage Measurement Accuracy (100 V Range)±(0.05% of reading + 15 mV)

Accuracy (Current Source)

ParameterValue
Current Source Accuracy (1 nA Range)±(0.5% of setting + 3 pA + 10 fA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (10 nA Range)±(0.25% of setting + 15 pA + 100 fA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (100 nA Range)±(0.25% of setting + 100 pA + 1 pA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (1 µA Range)±(0.2% of setting + 1 nA + 10 pA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (10 µA Range)±(0.2% of setting + 10 nA + 100 pA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (100 µA Range)±(0.2% of setting + 100 nA + 1 nA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (1 mA Range)±(0.2% of setting + 1 µA + 10 nA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (10 mA Range)±(0.2% of setting + 10 µA + 100 nA × Vo) [Vo: output voltage in Volts]
Current Source Accuracy (100 mA Range)±(0.2% of setting + 100 µA + 1 µA × Vo) [Vo: output voltage in Volts]

Accuracy (Current Measurement)

ParameterValue
Current Measurement Accuracy (1 nA Range)±(0.5% of reading + 3 pA + 10 fA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (10 nA Range)±(0.25% of reading + 15 pA + 100 fA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (100 nA Range)±(0.25% of reading + 100 pA + 1 pA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (1 µA Range)±(0.2% of reading + 1 nA + 10 pA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (10 µA Range)±(0.2% of reading + 10 nA + 100 pA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (100 µA Range)±(0.2% of reading + 100 nA + 1 nA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (1 mA Range)±(0.2% of reading + 1 µA + 10 nA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (10 mA Range)±(0.2% of reading + 10 µA + 100 nA × Vm) [Vm: measurement voltage in Volts]
Current Measurement Accuracy (100 mA Range)±(0.2% of reading + 100 µA + 1 µA × Vm) [Vm: measurement voltage in Volts]

Guard Specifications

ParameterValue
Guard Offset Voltage±2 mV max
Maximum Guard Capacitance100 pF

Safety & Limits

ParameterValue
Maximum Common Mode Voltage±42 V

Technology

Source/Monitor Unit (SMU)

Applications

Semiconductor parameter analysis and DC characterization