The Keysight (Agilent HP) 41422A is a High Current Source/Monitor Unit (HCSMU) module designed for use in the HP 4142B Modular DC Source/Monitor mainframe. This dual-slot module offers high-current force and measurement capabilities up to 1 A DC and up to 2 A in pulse mode, making it well-suited for high-power semiconductor device characterization.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB / HP-IB (via HP 4142B mainframe) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered directly from HP 4142B mainframe backplane |
| Operating Temperature | 5 °C to 40 °C |
| Weight | 2.1 |
Module Configuration
| Parameter | Value |
|---|---|
| Module Type | High Current Source/Monitor Unit (HCSMU) |
| Mainframe Slot Requirements | 2 slots (occupies 2 slots in HP 4142B mainframe) |
| Compatible Mainframe | HP 4142B Modular DC Source/Monitor |
System and Channel Configuration
| Parameter | Value |
|---|---|
| Output Channels | 1 Channel |
| Sensing Modes | Kelvin Connection (4-wire Remote Sensing) |
| Output Terminals | Force (Triaxial), Sense (Triaxial), Common Force (Triaxial), Common Sense (Triaxial) |
Voltage Source Specifications
| Parameter | Value |
|---|---|
| Voltage Output Ranges | ±2 V, ±10 V |
| Voltage Programming Resolution (2 V Range) | 100 µV |
| Voltage Programming Resolution (10 V Range) | 500 µV |
| Voltage Programming Accuracy (2 V Range) | ±(0.06% + 400 µV) |
| Voltage Programming Accuracy (10 V Range) | ±(0.06% + 2 mV) |
| Voltage Compliance Range | 100 µV to 10 V |
Current Source Specifications
| Parameter | Value |
|---|---|
| Current Output Ranges (DC) | 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A |
| Current Output Ranges (Pulse) | 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 10 A |
| Current Programming Resolution (10 µA Range) | 1 nA |
| Current Programming Resolution (100 µA Range) | 10 nA |
| Current Programming Resolution (1 mA Range) | 100 nA |
| Current Programming Resolution (10 mA Range) | 1 µA |
| Current Programming Resolution (100 mA Range) | 10 µA |
| Current Programming Resolution (1 A Range) | 100 µA |
| Current Programming Accuracy (10 µA Range) | ±(0.2% + 10 nA) |
| Current Programming Accuracy (100 µA Range) | ±(0.2% + 100 nA) |
| Current Programming Accuracy (1 mA Range) | ±(0.2% + 1 µA) |
| Current Programming Accuracy (10 mA Range) | ±(0.2% + 10 µA) |
| Current Programming Accuracy (100 mA Range) | ±(0.2% + 100 µA) |
| Current Programming Accuracy (1 A Range) | ±(0.4% + 1 mA) |
| Current Compliance Range (DC) | 10 nA to 1 A |
| Current Compliance Range (Pulse) | 10 nA to 10 A |
Measurement Specifications
| Parameter | Value |
|---|---|
| Voltage Measurement Ranges | 2 V, 10 V |
| Voltage Measurement Resolution (2 V Range) | 100 µV |
| Voltage Measurement Resolution (10 V Range) | 500 µV |
| Voltage Measurement Accuracy (2 V Range) | ±(0.05% + 200 µV) |
| Voltage Measurement Accuracy (10 V Range) | ±(0.05% + 1 mV) |
| Current Measurement Ranges (DC) | 10 mA, 100 mA, 1 A |
| Current Measurement Ranges (Pulse) | 10 mA, 100 mA, 1 A, 10 A |
| Current Measurement Accuracy (10 mA Range) | ±(0.2% + 5 µA) |
| Current Measurement Accuracy (100 mA Range) | ±(0.2% + 50 µA) |
| Current Measurement Accuracy (1 A Range) | ±(0.4% + 500 µA) |
Dynamic and Pulse Characteristics
| Parameter | Value |
|---|---|
| Minimum Pulse Width | 100 µs |
| Maximum Pulse Width | 10 ms |
Output Capabilities
| Parameter | Value |
|---|---|
| Maximum Peak Current | 10 A (pulse mode only) |
| Maximum DC Current | 1 A |
| Maximum Peak Power | 100 W |
| Maximum DC Power | 10 W |
| Maximum Output Voltage | 10 V |
Pulse Timing
| Parameter | Value |
|---|---|
| Pulse Width Resolution | 10 µs |
| Pulse Period Range | 1 ms to 100 ms |
| Pulse Period Resolution | 100 µs |
| Maximum Duty Cycle (10 A Range) | 1% |
Measurement Capabilities
| Parameter | Value |
|---|---|
| Sensing Method | Kelvin (4-wire) connection |
Programming Accuracy
| Parameter | Value |
|---|---|
| Current Programming Accuracy (10 A Pulse Range) | ±(0.4% of setting + 22 mA) |
Programming Resolution
| Parameter | Value |
|---|---|
| Current Programming Resolution (10 A Pulse Range) | 2 mA |
Measurement Accuracy
| Parameter | Value |
|---|---|
| Current Measurement Accuracy (10 A Pulse Range) | ±(0.4% of reading + 17 mA) |
Technology
Modular Source Measure Unit (SMU)
Applications
High-current semiconductor device testing and parameter analysis