The Keysight HP 4192A is a high-performance LF Impedance Analyzer designed for both network analysis and impedance analysis on telecommunication filters, audio/video electronic circuits, and basic components. Operating over a 5 Hz to 13 MHz frequency range, it offers mHz-level frequency resolution, a variable test signal, and integrated DC bias. The 4192A supports both floating and grounded device testing with automated sweep capabilities and HP-IB (GPIB) programmability.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | HP-IB (GPIB) |
| Display | Dual 4 1/2-digit 7-segment LED displays with parameter indicators |
| Warm-up Time | 30 minutes |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 V ±10% (switchable), 48 to 66 Hz, ≤150 VA |
| Operating Temperature | 0 to 55 °C |
| Dimensions (W×H×D) | 425.5 mm × 235 mm × 612 mm |
| Weight | 19 kg |
Measurement Parameters & Functions
| Parameter | Value |
|---|---|
| Primary Parameters | |Z|, |Y|, Gain-Phase (relative amplitude, phase, group delay) |
| Secondary Parameters | Theta, R, X, G, B, L, C, D, Q, Delta, Delta% |
| Ranging Modes | Auto, Manual |
| Device Measurement Compatibility | Floating or grounded devices |
| Zero Compensation | Automated zero adjustment (Open/Short) |
| Self-Test Capability | Built-in automatic self-test |
Test Signal (Stimulus) Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 5 Hz to 13 MHz |
| Frequency Resolution (5 Hz to 10 kHz) | 1 mHz |
| Frequency Resolution (10 kHz to 100 kHz) | 10 mHz |
| Frequency Resolution (100 kHz to 1 MHz) | 0.1 Hz |
| Frequency Resolution (1 MHz to 13 MHz) | 1 Hz |
| AC Test Signal Level Range | 5 mV to 1.1 Vrms |
| AC Test Signal Level Resolution (5 mV to 100 mV) | 1 mV |
| AC Test Signal Level Resolution (100 mV to 1.1 V) | 5 mV |
DC Bias Source
| Parameter | Value |
|---|---|
| Internal DC Bias Range | -35.00 V to +35.00 V |
| Internal DC Bias Resolution (0.00 V to ±9.99 V) | 10 mV |
| Internal DC Bias Resolution (±10.0 V to ±35.0 V) | 50 mV |
| External DC Bias Input Range | Up to ±45 V |
Measurement Ranges
| Parameter | Value |
|---|---|
| Impedance Range (|Z|, R, X) | 0.1 mOhm to 1.2999 MOhm |
| Admittance Range (|Y|, G, B) | 1 nS to 1.2999 S |
| Capacitance Range (C) | 0.1 fF to 129.99 mF |
| Inductance Range (L) | 0.01 nH to 1299.9 H |
| Dissipation Factor Range (D) | 0.0001 to 19.999 |
| Quality Factor Range (Q) | 0.01 to 10000 |
| Phase Angle Range (Theta) | -180.00° to +180.00° |
Measurement Speed & Sweep Capabilities
| Parameter | Value |
|---|---|
| Sweep Parameters | Frequency, DC Bias |
| Sweep Modes | Single, Continuous, Manual |
Front Panel & Display
| Parameter | Value |
|---|---|
| Measurement Terminals | 4-terminal pair configuration BNC |
Rear Panel Interfaces & Control
| Parameter | Value |
|---|---|
| Recorder Outputs | X, Y1, Y2 analog recorder outputs (rear panel) |
Performance
| Parameter | Value |
|---|---|
| Basic Measurement Accuracy | ±0.1% |
| Frequency Accuracy | ±0.005% |
| Measurement Time (Normal Mode) | 187 ms (at ≥ 10 kHz) |
| Measurement Time (Fast Mode) | 87 ms (at ≥ 10 kHz) |
System & Control
| Parameter | Value |
|---|---|
| Trigger Modes | Internal, Manual, External, HP-IB |
| HP-IB Interface Functions | SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0 |
Electrical
| Parameter | Value |
|---|---|
| Power Consumption | ≤ 150 VA |
Environmental
| Parameter | Value |
|---|---|
| Operating Humidity | ≤ 95% RH at 40 °C |
DC Bias
| Parameter | Value |
|---|---|
| Internal DC Bias Current Capacity | 40 mA max. |
| External DC Bias Maximum Input Voltage | ±35 V |
| External DC Bias Maximum Current | 100 mA max. |
Reference Oscillator
| Parameter | Value |
|---|---|
| Option 001 Reference Aging Rate | ±5 × 10^-8 per day |
| Option 001 Reference Temp Stability | ±1 × 10^-8 (0 °C to 55 °C) |
Rear Panel Connections
| Parameter | Value |
|---|---|
| DC Bias Monitor Output | 1/10 of actual DC bias voltage (output impedance 10 kΩ) |
| External Trigger Input | BNC, TTL compatible, active on negative-going transition (pulse width ≥ 2 µs) |
| External Reference Input | BNC, 1 MHz, 2 MHz, or 5 MHz (≥ 0 dBm into 50 Ω) |
| Reference Frequency Output | BNC, 1 MHz (≥ 0 dBm into 50 Ω) |
Measurement Functions
| Parameter | Value |
|---|---|
| Deviation Measurement Modes | Δ (difference) and Δ% (percent difference) from stored reference |
Technology
LF Impedance Analyzer
Applications