The Keysight Agilent HP 6030A is a 1200 W, single-output, autoranging programmable DC power supply designed for automatic test systems. It provides high efficiency, GPIB programmability, analog control of output voltage and current, and extensive built-in measurements in a compact 3U rack-mount form factor.

Specifications

General

ParameterValue
InterfacesGPIB (HP-IB), Analog Programming/Monitoring
DisplayDual 4-digit LED displays (voltage and current)
Recommended Calibration Interval1 year

Physical

ParameterValue
Power Supply120 VAC nominal (standard); 100 / 220 / 240 VAC options, 47-63 Hz, 2800 VA max
Operating Temperature0 to 55 °C
Dimensions (W×H×D)425.5 mm × 132.6 mm × 503.7 mm
Weight16.3

Output Ratings and Capabilities

ParameterValue
Number of Channels1
Rated Output Voltage0 to 200 V
Rated Output Current0 to 17 A
Rated Output Power1000 W

Load and Line Regulation

ParameterValue
Voltage Load Regulation0.01% + 15 mV
Current Load Regulation0.01% + 10 mA
Voltage Line Regulation0.01% + 10 mV
Current Line Regulation0.01% + 10 mA

Ripple and Noise

ParameterValue
Constant Voltage Ripple and Noise (Vrms)22 mV
Constant Voltage Ripple and Noise (Vp-p)50 mV
Constant Current Ripple and Noise (Irms)15 mA

Programming Accuracy and Resolution

ParameterValue
Voltage Programming Accuracy0.035% + 145 mV
Current Programming Accuracy0.2% + 25 mA
Voltage Programming Resolution50 mV
Current Programming Resolution4.25 mA
OVP Programming Accuracy1.6% + 1.2 V
OVP Programming Resolution1.2 V

Readback Accuracy and Resolution

ParameterValue
Voltage Readback Accuracy0.08% + 80 mV
Current Readback Accuracy0.36% + 15 mA
Voltage Readback Resolution50 mV
Current Readback Resolution5 mA

Protection Features

ParameterValue
Over-Voltage Protection (OVP) Range2.4 to 230 V

System Connectivity and Interfaces

ParameterValue
GPIB (IEEE-488) InterfaceStandard (HP-IB)
Analog Programming and Monitoring0 to 5 V, 0 to 10 V, 0 to 5 kΩ, or 0 to 10 kΩ control options
Remote Sensing Compensation VoltageUp to 2 V drop per load lead

AC Input Specifications

ParameterValue
AC Input Voltage Options100 VAC / 120 VAC / 220 VAC / 240 VAC (factory-configured options)
AC Input Frequency Range47 to 63 Hz

Physical and Mechanical

ParameterValue
Form Factor3U rack-mount
Cooling System DesignForced air cooling with internal fan, side/bottom intake and rear exhaust
Shipping Weight21.8 kg

Environmental and Standards

ParameterValue
Storage Temperature Range-40 to +75 °C

Power Specifications

ParameterValue
AC Input Power (VA)2800 VA
AC Input Current24 A @ 120 VAC, 14 A @ 200 VAC, 13 A @ 220 VAC, 12 A @ 240 VAC

Performance

ParameterValue
Temperature Coefficient (Voltage)0.01% + 3 mV per °C
Temperature Coefficient (Current)0.02% + 0.5 mA per °C
Command Processing Time15 ms (average time for output to change after receipt of GPIB command)
Output Rise Time (Full Load)100 ms
Output Rise Time (No Load)100 ms
Output Fall Time (Full Load)250 ms
Output Fall Time (No Load)6 s
Transient Recovery Time1 ms to recover within 150 mV of nominal output voltage following a load change

Safety & Isolation

ParameterValue
Maximum Isolation Voltage±550 VDC

Environmental

ParameterValue
Operating AltitudeUp to 2000 m
Operating HumidityUp to 80% RH (non-condensing)

Stability

ParameterValue
Output Voltage Drift0.03% + 5 mV (over 8 hours)
Output Current Drift0.03% + 3 mA (over 8 hours)

Capability

ParameterValue
Series Connection LimitUp to 2 units
Parallel Connection LimitUp to 3 units

Inputs & Outputs

ParameterValue
Analog Control Input Ranges0 to 5 V (voltage control) or 0 to 4 kΩ (resistance control)
Rear Panel ConnectionsGPIB (IEEE-488) connector, 15-pin analog control terminal block (J1), output bus bars

Technology

Switch-mode / Autoranging

Applications

Automated test equipment (ATE), system power, industrial testing