The Keysight Technologies (formerly Agilent HP) 6060B is a 300 W, single-output DC electronic load designed for test systems and bench evaluation of DC power sources and components. Equipped with built-in measurements and GPIB programmability using standard SCPI commands, it offers flexible operating modes including Constant Current, Constant Voltage, and Constant Resistance along with comprehensive protection features.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488)
DC Input Isolation Voltage±240 VDC
Recommended Calibration Interval1 year

Physical

ParameterValue
Operating Temperature0 to 55 °C

Input Ratings

ParameterValue
Maximum Input Power300 W
Maximum Input Current60 A
Maximum Input Voltage60 V
Minimum Operating Voltage3 V (at 60 A)

Constant Current (CC) Mode

ParameterValue
CC Operating Ranges0 to 6 A, 0 to 60 A
CC Programming Accuracy0.1% ±75 mA
CC Regulation10 mA

Constant Voltage (CV) Mode

ParameterValue
CV Operating Range0 to 60 V
CV Programming Accuracy0.1% ±50 mV
CV Regulation10 mV (with remote sense)

Constant Resistance (CR) Mode

ParameterValue
CR Operating Ranges0.033 to 1 Ω / 1 to 1,000 Ω / 10 to 10,000 Ω
CR Programming Accuracy (1 kΩ range)0.3% ±8 mS (with ≥6V at input)

Transient Operation

ParameterValue
Transient Frequency Range0.25 Hz to 10 kHz

Protection Features

ParameterValue
Overvoltage Protection (OVP)Included
Overcurrent Protection (OCP)Included
Overtemperature Protection (OTP)Included
Overpower Protection (OPP)Included
Reverse Polarity ProtectionIncluded

Connectivity and System Control

ParameterValue
Command LanguageSCPI (Standard Commands for Programmable Instruments)
Driver SupportVXI Plug & Play
Remote Sensing TerminalsYes

Readback Accuracy

ParameterValue
Voltage Readback Accuracy±(0.05% + 45 mV)
Current Readback Accuracy (60 A Range)±(0.12% + 65 mA)

Resolution

ParameterValue
Voltage Programming Resolution16 mV
Current Programming Resolution (60 A Range)16 mA
Voltage Readback Resolution16 mV
Current Readback Resolution (60 A Range)16 mA

Analog Outputs

ParameterValue
Voltage Monitor Output0 to 10 V
Current Monitor Output0 to 10 V

Isolation & Grounding

ParameterValue
Maximum Common-Mode Voltage±240 VDC

Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C

Performance

ParameterValue
Ripple and Noise (Current)4 mA rms / 40 mA p-p

Inputs & Outputs

ParameterValue
External Analog Programming Voltage Range0 to 10 V
External Analog Programming Input Impedance10 kΩ nominal

Power

ParameterValue
AC Input Voltage Selection100 / 120 / 220 / 240 VAC (switchable)

Technology

Solid-state DC Electronic Load

Applications

Power source evaluation, DC-DC converter testing, battery capacity testing, and general bench test applications