The Keysight Agilent HP 6296A is a classic, highly reliable, convection-cooled linear DC power supply. Designed for bench and rack use, it operates in both constant-voltage (CV) and constant-current (CC) modes. It features a single-output rated at 180 Watts (0-60 V, 0-3 A), built-in remote sensing, analog programming, auto-series, auto-parallel, and auto-tracking capabilities with front and rear output terminals.

Specifications

General

ParameterValue
InterfacesAnalog programming (voltage/resistance), remote sensing
DisplayDual front-panel analog meters (voltage and current)

Physical

ParameterValue
Power Supply115 VAC ±10%, single-phase, 57-63 Hz (230 VAC optional)
Operating Temperature0 to 55 °C
Weight10.9

Output Ratings and Capabilities

ParameterValue
Number of Channels1
Rated Output Voltage0 to 60 V
Rated Output Current0 to 3 A
Rated Output Power180 W
Output ConfigurationConstant Voltage (CV) / Constant Current (CC)
Output Voltage IsolationUp to 300 VDC above/below ground
Output TerminalsFront and rear output terminals

Load and Line Regulation

ParameterValue
Voltage Line Regulation0.01% + 1 mV
Current Line Regulation0.05% + 1 mA
Voltage Load Regulation0.01% + 1 mV
Current Load Regulation0.05% + 1 mA

Ripple and Noise

ParameterValue
Constant Voltage Ripple and Noise (Vrms)500 uV rms (20 Hz to 20 MHz)
Constant Voltage Ripple and Noise (Vp-p)25 mV peak-to-peak (20 Hz to 20 MHz)
Constant Current Ripple and Noise (Irms)3 mA rms (20 Hz to 20 MHz)

Design & Features

ParameterValue
Voltage Control TypeCoarse and fine controls
Current Control TypeCoarse and fine controls
Remote SensingSupported
Auto-Series CapabilitySupported
Auto-Parallel CapabilitySupported
Auto-Tracking CapabilitySupported

Physical and Mechanical

ParameterValue
Cooling System DesignConvection cooled
Form Factor5-1/4 inch height, half rack width

Performance

ParameterValue
Transient Recovery Time< 50 µs to within 15 mV following full to half (or half to full) load change
Voltage Temperature Coefficient0.01% + 1 mV per °C
Voltage Drift (8 hours)0.1% + 5 mV
Current Drift (8 hours)0.1% + 3 mA
Output Impedance (DC to 100 Hz)1.2 mΩ (typical)
Output Impedance (100 Hz to 1 kHz)12 mΩ (typical)
Output Impedance (1 kHz to 100 kHz)0.15 Ω (typical)
Output Impedance (100 kHz to 1 MHz)1.5 Ω (typical)
Remote Programming Coefficient (Current)333 Ω/A
Current Temperature Coefficient0.02% + 1 mA/°C

Controls

ParameterValue
Voltage Control Resolution15 mV
Current Control Resolution10 mA

Display & Meters

ParameterValue
Meter Accuracy±2% of full scale

Protection

ParameterValue
Overvoltage Protection Crowbar Range10 V to 65 V (Option 011 dependent)

Programming & Control

ParameterValue
Remote Programming Coefficient (Voltage)200 Ω/V

Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C

Power

ParameterValue
AC Input Voltage115 VAC ±10% (standard) / 230 VAC ±10% (with Option 28)
AC Input Frequency Range48 Hz to 63 Hz
AC Input Current3.1 A (at 115 VAC)

Technology

Linear

Applications

Precision DC laboratory power and analog system integration