- Home
- Categories
- Test equipment
- Power Supplies
- Agilent HP
- 6610 Series - 40-50W
The Keysight (formerly Agilent/HP) 6610 Series includes the 6611C, 6612C, 6613C, and 6614C single-output programmable DC power supplies. Providing 40 W to 50 W of power, these compact, high-performance units offer a one-box solution with low-noise outputs and built-in measurement capabilities. They feature fast output programming with active downprogramming, front-panel control, and standard GPIB and RS-232 interfaces utilizing the SCPI command set for flexible automated test system integration.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB, RS-232 |
| Warm-up Time | 30 minutes (to meet precision specifications) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 VAC ±10%, 47-63 Hz, 250 VA max |
| Operating Temperature | 0 to 55 |
| Dimensions (W×H×D) | 212.8 x 88.1 x 348.3 |
| Weight | 8.85 |
| Form Factor | 2U, half-rack width |
| Cooling | Fan cooled |
Output Ratings
| Parameter | Value |
|---|---|
| Number of Outputs | 1 |
| Maximum Output Power | 40 W / 50 W (model dependent) |
| Output Voltage Range | 0-8 V / 0-20 V / 0-50 V / 0-100 V (model dependent) |
| Output Current Range | 0-0.5 A / 0-1 A / 0-2 A / 0-5 A (model dependent) |
Dynamic Response and Timing
| Parameter | Value |
|---|---|
| Output Programming Response | Fast output programming with active downprogramming |
Supplemental Characteristics
| Parameter | Value |
|---|---|
| Series Operation Capability | Supported |
| Parallel Operation Capability | Supported |
Interfaces and Control
| Parameter | Value |
|---|---|
| Standard Control Interfaces | GPIB, RS-232 |
| Command Routing/Protocol | SCPI command set |
Power Requirements
| Parameter | Value |
|---|---|
| Maximum Input Power | 250 VA |
| Line Frequency | 47 to 63 Hz |
Safety & Isolation
| Parameter | Value |
|---|---|
| Isolation Voltage | ±240 VDC maximum between any output terminal and chassis ground |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature | -40 °C to 70 °C |
| Operating Altitude | Up to 2,000 m (6,560 ft) |
Performance
| Parameter | Value |
|---|---|
| Transient Response Time | < 50 µs (to within 15 mV for a 50% load change) |
| Remote Sense Capability | Compensates for up to 2 V load-lead drop per lead |
Power
| Parameter | Value |
|---|---|
| AC Input Voltage | 100 / 120 / 220 / 240 VAC ±10% (configuration/option dependent) |
Applications
Test and Measurement