- Home
- Categories
- Test equipment
- Power Supplies
- Agilent HP
- 6620 Series - 25-105W
The Keysight Technologies (formerly Agilent HP) 6620 Series - 25-105W represents a family of programmable DC system power supplies designed for automated test environments. These multi-output units offer precise voltage and current control, rapid programming, and comprehensive protection features, making them suitable for a wide range of applications requiring stable and reliable power delivery.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE 488.2) |
| Display | Segmented LED (Voltage/Current per channel) |
| Recommended Calibration Interval | 1 year |
| Warm-up Time | 30 minutes to rated specifications |
| Self-Test | Power-on self-test |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100-120 VAC / 220-240 VAC (selectable), 47-63 Hz |
| Operating Temperature | 0 to 55 °C |
| Dimensions (W×H×D) | 426 mm × 133 mm × 497 mm |
| Weight | 18.1 kg |
| Cooling Type | Fan-cooled |
| Cooling Airflow | Front intake, rear exhaust |
General Specifications
| Parameter | Value |
|---|---|
| Product Series | 6620 Series |
| Output Power Range (Total) | 25 W to 105 W (model dependent) |
Output Specifications
| Parameter | Value |
|---|---|
| Number of Outputs | 2, 3, or 4 (model dependent) |
| Output Voltage Range | Up to 50 V (channel dependent, e.g., 0-7V, 0-16V, 0-20V, 0-50V) |
| Output Current Range | Up to 10 A (channel dependent, e.g., 0-0.8A, 0-2A, 0-5A, 0-10A) |
| Voltage Programming Accuracy | ±(0.05% of output + 10 mV) (<20V outputs), ±(0.05% of output + 30 mV) (>20V outputs) |
| Voltage Programming Resolution | 10 mV (<20V outputs), 30 mV (>20V outputs) |
| Current Programming Accuracy | ±(0.1% of output + 10 mA) (<2A outputs), ±(0.1% of output + 20 mA) (<5A outputs), ±(0.1% of output + 50 mA) (<10A outputs) |
| Current Programming Resolution | 10 mA (<2A outputs), 20 mA (<5A outputs), 50 mA (<10A outputs) |
| Voltage Load Regulation | <0.01% + 2 mV (<20V outputs), <0.01% + 3 mV (>20V outputs) |
| Voltage Line Regulation | <0.01% + 2 mV (<20V outputs), <0.01% + 3 mV (>20V outputs) |
| Current Load Regulation | <0.01% + 2 mA (<2A outputs), <0.01% + 5 mA (<5A outputs), <0.01% + 10 mA (<10A outputs) |
| Current Line Regulation | <0.01% + 2 mA (<2A outputs), <0.01% + 5 mA (<5A outputs), <0.01% + 10 mA (<10A outputs) |
| PARD (Ripple and Noise) Voltage Peak-to-Peak | <15 mVpp (typically for <20V outputs) |
| PARD (Ripple and Noise) Voltage RMS | <0.5 mVrms (typically for <20V outputs) |
| PARD (Ripple and Noise) Current RMS | <1 mA rms (typically for <5A outputs) |
| Transient Response Time | <3 ms (to within 50mV of final value for 50% load change) |
Measurement (Readback) Specifications
| Parameter | Value |
|---|---|
| Voltage Readback Accuracy | ±(0.05% of output + 10 mV) |
| Voltage Readback Resolution | 10 mV |
| Current Readback Accuracy | ±(0.1% of output + 10 mA) |
| Current Readback Resolution | 10 mA |
Remote Sensing
| Parameter | Value |
|---|---|
| Remote Sense Compensation | Up to 0.5 V per lead |
Protection Features
| Parameter | Value |
|---|---|
| Over Voltage Protection (OVP) | Yes, programmable (e.g., 10% to 110% of Vset) |
| Over Current Protection (OCP) | Yes, programmable (current limit) |
| Over Temperature Protection (OTP) | Yes |
| Foldback Protection | Yes (current limiting) |
Programming and Interface
| Parameter | Value |
|---|---|
| Programming Language Compatibility | SCPI (Standard Commands for Programmable Instruments) |
| GPIB Compatibility | IEEE 488.2 |
| Triggering Capability | Yes (via GPIB) |
| Output Isolation Voltage (Output to Chassis) | 240 VDC |
| Channel Isolation Voltage (Channel to Channel) | 500 VDC (for multi-output models) |
Input Power Requirements
| Parameter | Value |
|---|---|
| AC Input Voltage Range | 100-120 VAC / 220-240 VAC |
| AC Input Frequency Range | 47-63 Hz |
| Maximum Input Power Consumption | 250 VA (for 105W models) |
| Input Current Rating | 3.2 A @ 100VAC, 1.6 A @ 200VAC (for 105W models) |
Environmental Conditions
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 to 70 °C |
| Operating Humidity Range | 15% to 95% RH, non-condensing |
| Storage Humidity Range | 5% to 90% RH, non-condensing |
| Operating Altitude | Up to 2000 m |
| Non-Operating Altitude | Up to 4600 m |
Physical Characteristics
| Parameter | Value |
|---|---|
| Form Factor | Benchtop, Rack Mount |
| Rack Units (U) | 3U |
Safety and Regulatory Compliance
| Parameter | Value |
|---|---|
| Safety Standards Compliance | IEC 61010-1, CSA 22.2 No. 61010-1 |
| EMC Standards Compliance | EN 61326-1, CISPR 11 Group 1 Class A |
| CE Mark | Yes |
| CSA Mark | Yes |
| IEC Compliance | Yes |
Included Accessories
| Parameter | Value |
|---|---|
| Power Cord | Yes |
| User Manual | Yes |
| Programming Manual | Yes |
Optional Accessories
| Parameter | Value |
|---|---|
| Rack Mount Kit | No (Standard accessory) |
| GPIB Cable | Yes (Optional accessory) |
Connectivity
| Parameter | Value |
|---|---|
| Remote Interface | GPIB (Standard), RS-232 (Optional via Option 002) |
| RS-232 Interface | Optional (via Option 002) |
Output Characteristics
| Parameter | Value |
|---|---|
| Voltage Programming Rise Time (no load) | < 40 ms |
| Voltage Programming Fall Time (no load) | < 100 ms |
| Voltage Programming Rise Time (full resistive load) | < 100 ms |
| Voltage Programming Fall Time (full resistive load) | < 100 ms |
| Current Programming Rise Time | < 50 ms |
| Current Programming Fall Time | < 50 ms |
| Output Voltage Stability (8 hours, CV mode) | < 0.01% + 200 µV |
| Output Current Stability (8 hours, CC mode) | < 0.05% + 500 µA |
| Remote Sense Lead Drop Compensation | 0.5 V per lead |
| Output Series Operation | Supported (external connection) |
| Output Parallel Operation | Supported (external connection) |
| Output Control | Individual channel ON/OFF |
Environmental
| Parameter | Value |
|---|---|
| Acoustic Noise | < 45 dBA |
Display
| Parameter | Value |
|---|---|
| Display Digits | Not universally 4 digits; some voltage ranges display 3 digits (e.g., 50.0V), while others display 4 digits. |
Technology
Linear Regulated DC Power Supply
Applications
Automated Test Equipment (ATE), Benchtop R&D, System Integration