The Keysight Technologies 6811B is a highly flexible, single-phase AC power source and power analyzer rated at 375 VA. Engineered for applications like power supply testing, UPS evaluation, and avionics ATE, it integrates the functionality of an AC source, a DC source, an arbitrary waveform generator, and a comprehensive power analyzer within a compact 3U rackmountable form factor.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232, Remote Inhibit, Discrete Fault
DisplayVFD (Vacuum Fluorescent Display)
Recommended Calibration Cycle1 year

Physical

ParameterValue
Power Supply99 to 135 VAC / 198 to 264 VAC (user switchable), 47 to 63 Hz
Operating Temperature0 to 40 °C
Dimensions (W×H×D)425.5 mm × 132.6 mm × 571.5 mm
Weight23.0

AC Output Ratings

ParameterValue
Maximum AC Power Rating375 VA
Output Phase ConfigurationSingle-phase
AC Voltage Ranges135 Vrms / 270 Vrms
Maximum AC RMS Voltage300 Vrms
Maximum AC RMS Current (135 V range)3.25 Arms
Maximum AC RMS Current (270 V range)1.625 Arms
Maximum Peak Repetitive Current (135 V range)12 A
Maximum Peak Repetitive Current (270 V range)6 A
Peak Current Crest Factor Capability3.6
Frequency Range45 Hz to 1000 Hz
Frequency Resolution10 mHz
Frequency Programming Accuracy0.01%

DC Output Ratings

ParameterValue
Maximum DC Power Rating285 W
DC Voltage Ranges190 VDC / 380 VDC
Maximum DC Voltage425 VDC
Maximum DC Current (190 V range)2.5 A
Maximum DC Current (380 V range)1.25 A

Output Distortion and Regulation

ParameterValue
Total Harmonic Distortion (THD) - Full Load (50/60 Hz)< 0.5%
Total Harmonic Distortion (THD) - Full Load (45 Hz to 1 kHz)< 1.0%
AC Load Regulation0.5% of full scale
AC Line Regulation0.1% of full scale
DC Load Regulation0.5% of full scale
DC Line Regulation0.1% of full scale
Programming Voltage Accuracy (AC, 135 V range)0.15% of output + 0.3 V
Programming Voltage Accuracy (AC, 270 V range)0.15% of output + 0.6 V
Programming Voltage Accuracy (DC, 190 V range)0.1% of output + 0.5 V
Programming Voltage Accuracy (DC, 380 V range)0.1% of output + 1.0 V

Measurement and Analyzer Capabilities

ParameterValue
Voltage RMS Measurement Range0 to 300 V
Voltage RMS Measurement Accuracy (45 Hz to 100 Hz, 135 V range)0.15% of reading + 0.3 V
Voltage RMS Measurement Accuracy (45 Hz to 100 Hz, 270 V range)0.15% of reading + 0.6 V
Current RMS Measurement Accuracy (45 Hz to 100 Hz, Low range)0.5% of reading + 1.5 mA
Current RMS Measurement Accuracy (45 Hz to 100 Hz, High range)0.5% of reading + 15 mA
Active Power Measurement Accuracy (45 Hz to 100 Hz, Low range)1.5% of reading + 1 W
Active Power Measurement Accuracy (45 Hz to 100 Hz, High range)1.5% of reading + 3 W
Apparent Power Measurement Accuracy (45 Hz to 100 Hz, Low range)1.5% of reading + 1 VA
Apparent Power Measurement Accuracy (45 Hz to 100 Hz, High range)1.5% of reading + 3 VA
Power Factor Measurement Range0.00 to 1.00
Crest Factor Measurement Range1.0 to 10.0
Harmonic Analysis Frequency RangeFundamental (45 Hz to 1000 Hz) up to 50th harmonic

Waveform and Transient Generation

ParameterValue
Standard Waveform TypesSine, Square
Arbitrary Waveform Point Capacity1024 points
Transient Programming ModesStep, Pulse, List
Phase Angle Control Range0 to 360 degrees
Phase Angle Programming Accuracy1 degree
External Analog Modulation InputYes

System Connectivity and Control

ParameterValue
Command Language SupportSCPI

Protection and Safeguards

ParameterValue
Over-Voltage Protection (OVP)Yes
Over-Current Protection (OCP)Yes
Over-Temperature Protection (OTP)Yes

Physical and Environmental

ParameterValue
Storage Temperature Range-40 to 75 °C
Cooling MechanismForced-air with rear exhaust
Rack Unit Height3U

Power Input

ParameterValue
AC Input Power Consumption1000 VA maximum
AC Input Frequency Range47 Hz to 63 Hz
AC Input Voltage Option 100 Range87 VAC to 106 VAC
AC Input Voltage Option 120 Range104 VAC to 127 VAC
AC Input Voltage Option 200 Range174 VAC to 220 VAC
AC Input Voltage Option 230 Range191 VAC to 254 VAC

Outputs

ParameterValue
Output Isolation Voltage300 VAC / 425 VDC maximum relative to chassis ground
Remote Sensing CapabilityUp to 2 V drop per output lead

Programming Accuracy

ParameterValue
Voltage Programming Temperature Coefficient0.04% + 0.02 V per °C
Frequency Programming Temperature Coefficient10 ppm per °C

Measurement Resolution

ParameterValue
Voltage RMS Measurement Resolution0.01 V
Current RMS Measurement Resolution (Low Range)1 mA
Current RMS Measurement Resolution (High Range)10 mA

Measurement Performance

ParameterValue
Maximum Harmonic Measurement Order50th harmonic (for fundamental frequencies <= 250 Hz)

Environmental

ParameterValue
Maximum Operating Altitude2000 m

Measurements

ParameterValue
DC Voltage Measurement Range0 to ±190 VDC / 0 to ±380 VDC
Frequency Measurement Resolution0.01 Hz
Phase Angle Measurement Resolution0.1°
Active Power Measurement Resolution0.01 W (low range) / 0.1 W (high range)
Apparent Power Measurement Resolution0.01 VA (low range) / 0.1 VA (high range)
Power Factor Measurement Resolution0.01
Crest Factor Measurement Resolution0.01

Output Characteristics

ParameterValue
Programmable Resistance Range (Option 005)0 to 1.0 Ω
Programmable Inductance Range (Option 005)0 to 2 mH

Inputs & Outputs

ParameterValue
Remote Inhibit (RI) and Discrete Fault Indicator (DFI)Available on rear panel terminal block

Compliance

ParameterValue
Safety Standards ComplianceCertified to CSA C22.2 No. 1010.1, EN 61010-1, UL 3111-1
EMC Standards ComplianceComplies with EN 55011, EN 50082-1

Technology

Linear / Switching Hybrid

Applications