The Keysight (Agilent HP) 81536A is a high-sensitivity InGaAs optical power sensor module designed for the HP 8153A Lightwave Multimeter mainframe. It provides high-accuracy optical power measurements over a wavelength range of 800 to 1700 nm, supporting both single-mode and multimode fiber types.

Specifications

General

ParameterValue
InterfacesInternal mainframe interface (GPIB and RS-232 available via host mainframe)
DisplayNone (utilizes host mainframe display)

Physical

ParameterValue
Power SupplyPowered directly by host mainframe
Operating Temperature0 to 55 °C
Dimensions (W×H×D)75 × 32 × 335
Weight0.5
Shipping Weight1.2 kg

Optical Performance Specifications

ParameterValue
Wavelength Range800 to 1700 nm
Detector TypeInGaAs
Power Measurement Range-70 to +3 dBm
Maximum Safe Input Power+10 dBm
Calibration Uncertainty±2.5% (1000 to 1600 nm, at reference conditions)
Power Linearity±0.05 dB (+3 to -60 dBm), ±0.15 dB (-60 to -70 dBm)
Return Loss> 40 dB

Measurement and Dynamic Characteristics

ParameterValue
Noise (Peak-to-Peak)< 50 pW (1200 to 1600 nm, 1 s averaging time)

Optical Interface and Compatibility

ParameterValue
Compatible Fiber Types9/125 µm to 100/140 µm
Maximum Numerical Aperture0.3
Connector Adapter InterfaceHP 81000xI series (e.g., FC, SC, ST, DIN adapters)

Mainframe Integration and Interface

ParameterValue
Compatible Host MainframesHP / Agilent 8153A Lightwave Multimeter
Required Slot Width1 slot

Environmental

ParameterValue
Storage Temperature-40 °C to +70 °C
Operating Humidity< 95% R.H. (0 °C to 40 °C, non-condensing)

Operating Features

ParameterValue
Warm-up Time40 minutes

Optical Specs

ParameterValue
Reference Wavelengths1300 nm, 1550 nm

Performance

ParameterValue
Polarization Dependence< ±0.02 dB (1250 to 1580 nm)
Temperature Dependence< ±0.03 dB (1200 to 1600 nm, 0 to 55 °C), < ±0.15 dB (800 to 1700 nm, 0 to 55 °C)
Detector Active Diameter2 mm
Total Operating Uncertainty±4% (1000 to 1600 nm, +3 to -60 dBm), ±7% (800 to 1700 nm, +3 to -60 dBm)
Averaging Time Range20 ms to 10 s (selectable)
Range SelectionAutomatic or manual (7 ranges)

Analog Output

ParameterValue
Analog Output Voltage0 to 2 V (into ≥ 10 kΩ, proportional to power in current range)
Analog Output Bandwidth3 Hz (ranges 1 to 4), 300 Hz (ranges 5 to 7)

Technology

InGaAs Optical Sensor

Applications

Optical Power Measurement