The Keysight (formerly Agilent / HP) 8481D is a high-accuracy diode RF and microwave power sensor. Operating across a frequency range of 10 MHz to 18 GHz, it is designed for measuring low power levels from -70 dBm to -20 dBm. To facilitate calibration with standard 1 mW (0 dBm) 50 MHz power meter reference outputs, the sensor is supplied with a precision 30 dB reference attenuator (Model 11708A). It provides exceptional SWR for minimized mismatch uncertainty and is compatible with a wide range of Keysight EPM, EPM-P, P-series, and legacy power meters.

Specifications

General

ParameterValue
InterfacesProprietary multi-pin connector for power meter sensor cables
DisplayNone (uses connected power meter display)

Physical

ParameterValue
Power SupplySupplied by connected power meter via sensor cable
Operating Temperature0 to 55 °C

Measurement & Performance

ParameterValue
Frequency Range10 MHz to 18 GHz
Power Measurement Range-70 to -20 dBm (100 pW to 10 µW)
Sensor Technology TypeDiode power sensing element
Sensor Dynamic Range50 dB
Zero Set±20 pW
Zero Drift±4 pW
Measurement Noise (Zero Noise)45 pW

Maximum Input Ratings

ParameterValue
Maximum Continuous Average Power100 mW (+20 dBm)
Maximum Peak Power100 mW (+20 dBm)

RF & Impedance Characteristics

ParameterValue
RF Input Impedance50 Ω nominal

Calibration & Traceability

ParameterValue
Reference Calibration FactorTraceable to US National Institute of Standards and Technology (NIST)

Compatibility & Interface

ParameterValue
RF Input Connector TypeType-N (male)
Compatible Power Meter ModelsEPM (E4418A/B, E4419A/B), EPM-P (E4416A, E4417A), P-series (N1911A, N1912A), E1416A VXI, and discontinued 435B, 436A, 437B, 438A, 70100A
Compatible Interface CablesKeysight 11730 Series (11730A, 11730B, 11730C, 11730D, 11730E, 11730F)

Accessories & Configuration Options

ParameterValue
Included Reference Attenuator ModelAgilent 11708A 30 dB Attenuator

Environmental & Reliability

ParameterValue
Storage Temperature Range-40 °C to +75 °C

Environmental

ParameterValue
Operating HumidityUp to 95% relative humidity at 40 °C (non-condensing)
Operating Altitude4572 m (15000 ft)
Non-Operating Altitude15240 m (50000 ft)

Power Limits with Attenuator

ParameterValue
Maximum Peak Power (with Attenuator)250 mW (+24 dBm) peak, pulse width <= 10 µs

Reference Attenuator

ParameterValue
Reference Attenuator Attenuation Accuracy (50 MHz)±0.05 dB
Reference Attenuator SWR (DC to 12.4 GHz)1.20
Reference Attenuator SWR (12.4 to 18 GHz)1.25

Technology

Diode

Applications

RF and microwave power measurement