The Keysight 8482B (formerly Agilent HP) is a high-power thermocouple power sensor designed for precise RF and microwave power measurements over a frequency range of 100 kHz to 4.2 GHz. Integrating a 30 dB attenuator, it accurately measures average power from 0 dBm to +44 dBm (1 mW to 25 W) with exceptional SWR and linearity, making it a standard choice for high-power transmitter testing and system calibration.

Specifications

General

ParameterValue
InterfacesProprietary 12-pin multi-contact connector (compatible with Keysight 11730 series cables)
DisplayNone (results displayed on connected power meter)

Physical

ParameterValue
Power SupplySupplied by host power meter via sensor cable
Operating Temperature0 to 55 °C
Dimensions (W×H×D)137 mm × 83 mm × 127 mm
Weight0.8

RF Performance Specifications

ParameterValue
Frequency Range100 kHz to 4.2 GHz
Power Measurement Range0 to +44 dBm (1 mW to 25 W)
Maximum Continuous Input Power25 W average
Maximum Peak Input Power500 W (for pulse width ≤ 5 µs)
Nominal Impedance50 Ω
Maximum SWR (100 kHz to 2 MHz)1.10
Maximum SWR (2 MHz to 30 MHz)1.18
Maximum SWR (30 MHz to 2 GHz)1.10
Maximum SWR (2 GHz to 4.2 GHz)1.18

Calibration and Accuracy

ParameterValue
Reference Calibration Factor100% (at 50 MHz)
Power Linearity (0 to +35 dBm)±3%
Power Linearity (+35 to +44 dBm)±3%
Zero Set±50 µW
Zero Drift< ±10 µW (within 1 hour at constant temperature)
Measurement Noise< 110 µW

Physical and Connector Interfaces

ParameterValue
RF Input Connector TypeType-N (male)
Integrated Attenuator Value30 dB

System Compatibility

ParameterValue
Compatible Power MetersKeysight/Agilent EPM (E4418A/B, E4419A/B), EPM-P (E4416A, E4417A), P-Series (N1911A, N1912A), E1416A VXI, 435A/B, 436A, 437A/B, 438A, 70100A
Required Sensor Cable ModelKeysight 11730 Series

Environmental Conditions

ParameterValue
Storage Temperature Range-40 to +75 °C
Operating Humidity Range95% R.H. at 40 °C (non-condensing)
Operating Altitude Limit4,572 m (15,000 ft)
Storage Altitude Limit15,240 m (50,000 ft)

Physical Dimensions

ParameterValue
Overall Width137 mm
Overall Height83 mm
Overall Depth127 mm
Net Weight0.8 kg
Shipping Weight1.3 kg

Compliance and Quality Standards

ParameterValue
Recommended Calibration Cycle1 year

Operating Characteristics

ParameterValue
Sensor TechnologyThermocouple
Input CouplingAC coupled

Power Limits

ParameterValue
Maximum Energy per Pulse100 W·µs

Inputs & Outputs

ParameterValue
RF Connector GenderMale

Connectivity

ParameterValue
Maximum Sensor Cable Length61 m (200 ft)

Accuracy

ParameterValue
Calibration Factor RSS Uncertainty (100 kHz to 10 MHz)2.3%
Calibration Factor RSS Uncertainty (10 MHz to 30 MHz)2.3%
Calibration Factor RSS Uncertainty (30 MHz to 2 GHz)2.3%
Calibration Factor RSS Uncertainty (2 GHz to 4.2 GHz)2.4%
Calibration Factor Worst-Case Uncertainty (100 kHz to 10 MHz)4.3%
Calibration Factor Worst-Case Uncertainty (10 MHz to 30 MHz)4.3%
Calibration Factor Worst-Case Uncertainty (30 MHz to 2 GHz)4.3%
Calibration Factor Worst-Case Uncertainty (2 GHz to 4.2 GHz)4.6%

Environmental

ParameterValue
Vibration and Shock ComplianceMeets MIL-T-28800C, Class 5

Technology

Thermocouple

Applications

RF and Microwave Power Measurement