The Keysight Technologies (Agilent HP) 85053A is a 3.5 mm verification kit designed to verify the error-corrected measurement performance of Vector Network Analyzer systems, specifically the HP 8510 and other compatible high-performance VNAs. The kit includes precision standards with traceable measurement data for verifying system accuracy.

Specifications

Physical

ParameterValue
Operating Temperature20 °C to 26 °C

General Specifications

ParameterValue
Frequency Range300 kHz to 26.5 GHz
Connector TypePrecision 3.5 mm
System Impedance50 Ω
Compatible Vector Network AnalyzersHP 8510 series, HP 8720 series, Keysight PNA series
Compatible Calibration KitsHP/Keysight 85052B, 85052D (3.5 mm)
Recommended Calibration Interval12 months

Kit Contents and Standards

ParameterValue
Attenuators Included20 dB and 40 dB precision attenuators
Airlines Included7.5 cm beadless coaxial airline
Mismatch Airlines Included7.5 cm beadless stepped impedance airline
Data Media ProvidedData disk / storage media containing calibration and measurement data
Tools IncludedNone (uses tools from calibration kit)

Standard Device Details

ParameterValue
20 dB Attenuator Part Number08505-60113
40 dB Attenuator Part Number08505-60114
50 Ω Airline Part Number85053-60005
25 Ω Stepped Airline Part Number85053-60006

Electrical Performance

ParameterValue
Nominal Impedance of Standards50 Ω (except mismatch airline)
Stepped Airline Nominal Impedance25 Ω
Maximum Input Power1 W (average) for attenuators

Mechanical and Dimensions

ParameterValue
Airline Length7.5 cm
Airline Center Conductor Diameter1.52 mm (nominal)
Airline Outer Conductor Inside Diameter3.50 mm (nominal)
Connector Gender ConfigurationPrecision male to female on attenuators, sexless outer/mating inner on airlines

Environmental & Metrology

ParameterValue
Calibration Reference Temperature23 °C ± 1 °C
Storage Temperature Range-40 °C to +75 °C
TraceabilityTraceable to National Metrology Institutes (NIST)
Relative Humidity Range (Operating)Non-condensing

Technology

Coaxial RF & Microwave Calibration

Applications

Vector Network Analyzer error-corrected performance verification