The Agilent Keysight 86145B is a portable, high-performance optical spectrum analyzer designed for 50 GHz and wider Wavelength Division Multiplexing (WDM) systems. It delivers identical measurement capabilities to the benchtop Agilent 86142B but in a smaller footprint and lighter weight, featuring 10 pm wavelength accuracy, fast dual-sweep measurement speeds, and built-in WDM, DFB-LD, and EDFA test applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB, RS-232, Parallel (Centronics), VGA out |
| Display | 6.5-inch active matrix color LCD |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 VAC, 50/60 Hz, 150 W max |
| Operating Temperature | 0 to 55 °C |
| Dimensions (W×H×D) | 325 mm × 222 mm × 427 mm |
Wavelength Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 600 to 1700 nm |
| Wavelength Linearity | ±10 pm (1525 to 1570 nm) |
| Wavelength Repeatability | ±2 pm |
| Wavelength Span Range | 0.2 nm to 1100 nm, plus Zero Span |
Optical Power and Amplitude
| Parameter | Value |
|---|---|
| Power Measurement Range (Sensitivity) | -90 to +15 dBm (1250 to 1610 nm) |
| Maximum Total Safe Input Power | +20 dBm (total power) |
| Power Level Accuracy | ±0.35 dB |
| Power Flatness | ±0.2 dB (1250 to 1610 nm) |
| Polarization Dependency (PDL) | ±0.12 dB (at 1550 nm) |
| Level Linearity | ±0.05 dB |
| Optical Return Loss | ≥ 35 dB (with straight connector) |
Resolution and Bandwidth
| Parameter | Value |
|---|---|
| Resolution Bandwidth (FWHM) | 0.06 to 10 nm |
| Minimum Resolution Bandwidth | 0.06 nm |
| Maximum Resolution Bandwidth | 10 nm |
| Resolution Bandwidth Accuracy | ±3% (0.1, 0.2, 0.5, 1, 2, 5, 10 nm) |
Measurement Speed and Sweep
| Parameter | Value |
|---|---|
| Sweep Time (100nm Span) | 180 ms (chop off) |
| Sampling Points per Sweep | 51 to 10001 (user-selectable) |
Analysis and Application Software
| Parameter | Value |
|---|---|
| WDM Analysis Support | OSNR, Channel Power, Channel Wavelength |
| DFB-LD Analysis Support | Supported |
| FP-LD Analysis Support | Supported |
| LED Analysis Support | Supported |
| EDFA Analysis Support | Gain and Noise Figure |
| Spectral Width Analysis | Envelope, RMS, and Threshold modes |
External Connectivity and Control
| Parameter | Value |
|---|---|
| GPIB (IEEE-488) Interface | 2 ports (one HP-IB/GPIB, one Systems GPIB) |
| Floppy Disk Drive | 3.5-inch 1.44 MB (MS-DOS format) |
| External Expansion Slot | PCMCIA Type II |
Power and Electrical
| Parameter | Value |
|---|---|
| AC Input Voltage Range | 100 to 240 VAC |
| AC Input Frequency Range | 50 to 60 Hz |
| Power Consumption | 150 W max |
Physical Characteristics
| Parameter | Value |
|---|---|
| Footprint | 12.8 in × 16.8 in |
Environmental and Regulatory
| Parameter | Value |
|---|---|
| Warm-up Time | 40 minutes |
Inputs & Outputs
| Parameter | Value |
|---|---|
| External Display Port | VGA, 15-pin D-sub |
| External Trigger Input | BNC female, TTL compatible |
| Keyboard and Mouse Interfaces | Two PS/2 mini-DIN ports |
| Parallel Printer Port | Centronics 25-pin D-sub |
| Optical Input Connector Type | Interchangeable connector adapters (FC, SC, ST, DIN, HMS-10) |
Hardware Features
| Parameter | Value |
|---|---|
| Internal Printer | Built-in thermal printer |
Options
| Parameter | Value |
|---|---|
| Built-in EELED Source (Option 002) | Dual-wavelength edge-emitting LED source (1310 nm / 1550 nm) |
| Current Source Output Range (Option 001) | 0 to ±100 mA |
Measurement & Control
| Parameter | Value |
|---|---|
| Remote Control Command Languages | SCPI, HP 71450B emulation |
| Sweep Trigger Modes | Free Run, External, Single |
Optical Input
| Parameter | Value |
|---|---|
| Fiber Type Compatibility | 9/125 µm single-mode fiber |
Display
| Parameter | Value |
|---|---|
| Display Size | 6.5 in (165 mm) diagonal |
| Display Type | Active-matrix color LCD |
Bandwidth
| Parameter | Value |
|---|---|
| Resolution Bandwidth Settings | 0.07, 0.1, 0.2, 0.5, 1, 2, 5, 10 nm |
Wavelength
| Parameter | Value |
|---|---|
| Wavelength Display Resolution | 0.001 nm |
Power & Amplitude
| Parameter | Value |
|---|---|
| Amplitude Display Resolution | 0.01 dB |
| Amplitude Scale | 0.1 to 20 dB/div, Linear |
Measurement
| Parameter | Value |
|---|---|
| Sweep Trace Points | 51 to 10001 (user-selectable) |
Technology
Monochromator-based Optical Spectrum Analyzer
Applications
WDM system and component characterization