Keysight Technologies (Agilent HP) — Fiber Optics Communications Test Equipment

18 models with full specifications in Fiber Optics Comm.

This page summarizes Keysight (formerly Agilent / HP) instruments commonly used for fiber‑optic communications testing: optical spectrum analyzers (OSAs), programmable attenuators and signal conditioners, lightwave switches, polarization controllers/analyzers, and a modular OTDR mainframe. It is written for engineers and technical buyers who must match instrument capabilities to measurement tasks (WDM channel analysis, source characterization, automated test racks, power‑controlled measurements, polarization work or field OTDR testing). The copy explains how the product range is organized, what technical tradeoffs to expect between models, and which model lines are typically considered when specifying for R&D, manufacturing, or automated test environments.

Representative OSA wavelength range
600 nm to 1700 nm (86140A / 86140B / 86146B / 86145A / 86145B)
Highest OSA resolution (minimum RBW)
60 pm (86146B); 0.06 nm on several 86140-series models
Standard programmable attenuator range
60 dB (8156A, 8158B)
Integrated power‑monitored attenuator
N7751A/52A power meter: -80 dBm to +10 dBm (1250–1650 nm), max +16 dBm

All models

ModelTypeInterfacesOperating TemperaturePower Supply
8156AFiberoptic AttenuatorsGPIB0 to +55 °C100-240 Vrms ±10%, 50/60 Hz, 60 VA max
8163AIntensity, Power Meters——100 to 240 VAC, 50/60 Hz
86142BSpectrum AnalyzersGPIB (IEEE‑488.2)——
86142ASpectrum Analyzers———
86060CFiberoptic Switches———
8158BFiberoptic AttenuatorsHP-IB (GPIB)0 to 55 °C100 / 120 / 220 / 240 V AC (+10% / -15%), 48 - 66 Hz, 55 VA max
8169APolarization ControllersGPIB (HP-IB)0 to 55 °C100 / 115 / 230 / 240 V AC (+10% / -15%), 48 to 66 Hz, 45 VA max
86062CFiberoptic SwitchesGPIB, RS-2320 to 55 °C100 / 120 / 220 / 240 V AC (±10%), 48 to 66 Hz, 45 VA max
86140ASpectrum AnalyzersGPIB (HP-IB), RS-232, Parallel Printer Port, VGA Output, Analog Output, Trigger Input/Output, PS/2 Keyboard/Mouse Ports0 to +55 °C90 V to 260 V AC, 47 Hz to 63 Hz, < 230 VA
86140BSpectrum AnalyzersDual GPIB (System and Controller), RS-232, Centronics Parallel, PS/2 Keyboard/Mouse, VGA Output0 to 55—
86146BSpectrum AnalyzersGPIB (Dual ports), RS-232, Parallel (Centronics), VGA output0 °C to 55 °C100 V to 240 V AC, 50/60 Hz, 230 VA max
N7751A/52AFiberoptic AttenuatorsLAN, USB, GPIB5 to 40100 - 240 V AC ±10%, 50/60 Hz
11896APolarization ControllersGPIB (HP-IB)0°C to +55°C90 - 132 VAC or 198 - 264 VAC, 47 - 66 Hz, 45 VA max
8509BPolarization AnalyzersAnalog output (±10 V)——
E6000COTDRRS-232C (serial up to 115.2 kbps), Centronics (parallel printer port), Mini-DIN (external keyboard), PCMCIA Type II slot0 to +50 &#176;C100 to 240 V AC (&#177;10%), 50 to 60 Hz input via external AC adapter; 18 V DC, 2.0 A output
83437ALight Source - Broadband—0 °C to +55 °CPowered via mainframe slot
86145ASpectrum AnalyzersDual GPIB (HP-IB) ports, Parallel (Centronics) printer port, VGA video output, two PCMCIA Type II slots, External Trigger input, Analog output0 °C to +55 °C90 V AC to 260 V AC, 47 Hz to 440 Hz, power consumption maximum 230 VA
86145BSpectrum AnalyzersGPIB, RS-232, Parallel (Centronics), VGA out0 to 55 °C100 to 240 VAC, 50/60 Hz, 150 W max

Looking for Keysight Technologies (Agilent HP) — Fiber Optics Communications Test Equipment?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us

How the range is structured

Keysight’s fiber‑optic test instruments in this set are grouped by measurement function rather than a single continuous product family. The 86140-series covers optical spectrum analyzers in benchtop and portable variants (examples include the benchtop 86140A/B and 86142B, and the portable 86145A/B and 86145B which offer benchtop‑grade capability in a smaller footprint). The 86146B is an OSA variant that uses a double‑pass monochromator and specifies tighter wavelength accuracy for narrow‑line sources. Attenuation and signal conditioning are represented by dedicated programmable attenuators (8156A, 8158B) and integrated power‑monitored units (N7751A/N7752A). Switching and routing are handled by half‑ and full‑width lightwave switches (86060C, 86062C). Polarization instruments include motorized controllers (8169A, 11896A) and a legacy analyzer (8509B). The E6000C is a modular OTDR mainframe for field and installation work.

Optical spectrum analyzers — pick by accuracy, sensitivity and form factor

Use an OSA when you need wavelength, spectral shape, channel power or OSNR measurements. Models in this set span bench and portable form factors and different tradeoffs:

  • For benchtop WDM and component characterization with options for high accuracy and fast sweep rates, consider the 86142B or the 86146B; the 86146B documents wavelength accuracy to ±10 pm (with an internal calibrator option) and a minimum resolution bandwidth as small as 60 pm, with a noise floor down to about -90 dBm in the 1250–1610 nm region.
  • The 86140A/B and the portable 86145A/B provide a broad 600–1700 nm coverage and practical RBW choices (many 86140-series models list minimum RBW around 0.06–0.07 nm) and documented power ranges and level linearity suitable for lab or manufacturing throughput.
  • Watch maximum safe input power: some 86140-series portable/benchtop models list +20 dBm or +10 dBm limits, while the 86146B specifies up to +30 dBm total safe input power.

Attenuators, signal conditioning and automated test

For precise optical power control the 8156A and 8158B are programmable attenuators with a 60 dB range; the 8156A documents very fine resolution down to 0.001 dB and continuous attenuation ranging, while the 8158B lists 0.01 dB display resolution and 0.04 dB repeatability. When you need closed‑loop power setting and meter readout in one instrument, the N7751A/N7752A combine a power‑monitored attenuator with extra power meter channels, with a power‑meter range of -80 dBm to +10 dBm (1250–1650 nm), low polarization dependence and a maximum safe input for the meter channel of +16 dBm. These instruments expose LAN/USB/GPIB for automation and are half‑width 1U instruments suitable for rack integration.

Switching, polarization and OTDR choices

Choose switching hardware by channel count and routing topology: the 86060C family covers half‑rack lightwave switches in smaller 1/2‑rack configurations (1 or 2 inputs, 4–8 outputs), while the full‑width 86062C addresses large 1xN channel counts from 20 to 100 with documented crosstalk below -80 dB and a long switching lifetime. For polarization control and measurement, motorized controllers such as the 8169A (operating near 1475–1575 nm) and the four‑loop 11896A (1250–1600 nm, low insertion‑loss variation and PDL) provide repeatable SOP synthesis; the 8509B is a calibrated polarization analyzer covering 1200–1600 nm if you need Poincaré/DOP measurements. For OTDR tasks and fiber‑plant commissioning, the E6000C serves as a modular mainframe that accepts optical modules for different wavelengths and dynamic ranges.

How to choose between these instruments

Match instrument selection to the measurement objective and the key specs: wavelength coverage and minimum RBW for spectral resolution (86146B and 86140‑series values), sensitivity/noise floor and maximum safe input for handling strong sources, and attenuation range and resolution where power control is required (8156A/8158B). For automated test racks choose half‑width or full‑width form factors and confirmed remote interfaces (GPIB, LAN, USB, RS‑232 are documented across these models). When polarization fidelity is critical, compare insertion‑loss variation, PDL and provided wavelength range (for example, 11896A lists ±0.002 dB insertion‑loss variation and PDL < 0.01 dB). For field OTDR work select a modular mainframe such as the E6000C to adapt pulse widths and distance ranges via optical modules.

Popular models

Frequently asked questions

Which Keysight OSA is appropriate for narrow‑line DWDM lasers and tight wavelength accuracy?
Look to the 86146B or the high‑accuracy 86142B variants. The 86146B uses a double‑pass monochromator, specifies wavelength accuracy to about ±10 pm with the internal calibrator option, and offers minimum RBW down to 60 pm for narrow‑line source characterization.
When should I use an integrated power‑monitored attenuator instead of a stand‑alone programmable attenuator?
If you need closed‑loop setting of optical power and built‑in power readout for automation, the N7751A/N7752A combine an attenuator with InGaAs power meter channels (1250–1650 nm) with meter range -80 dBm to +10 dBm and low polarization dependence. Stand‑alone attenuators such as the 8156A provide higher attenuation resolution (0.001 dB) where a separate meter is acceptable.
Which switch should I choose for very large channel counts in automated production?
The 86062C is a full‑width 1xN lightwave switch designed for 20 to 100 output channels, with bidirectional operation, documented single‑mode wavelength range and crosstalk below -80 dB.
Which instrument is appropriate for polarization control with minimal insertion‑loss variation?
The 11896A motorized polarization controller documents insertion‑loss variation of ±0.002 dB and PDL < 0.01 dB across its operating wavelength range, making it suitable where low PDL and repeatable SOP control are required.