The Keysight Technologies (Agilent HP) 8922S is a stand-alone GSM mobile station test set designed for incoming inspection and mobile repair. Featuring a built-in GSM base station emulator and extensive signalling capabilities, it enables comprehensive mobile device testing without auxiliary hardware. Its band coverage can be extended to DCS1800 and PCS1900 when paired with the Agilent 83220A/E frequency extension.

Specifications

General

ParameterValue
InterfacesGPIB
DisplayCRT

Physical

ParameterValue
Power Supply100 to 240 VAC, 48 to 440 Hz, 500 VA max
Operating Temperature0 to +55 °C
Dimensions (W×H×D)426 mm × 177 mm × 574 mm
Weight31 kg

RF Signal Generator

ParameterValue
RF Generator Frequency Range10 MHz to 1 GHz
RF Generator Frequency Resolution1 Hz
RF Generator Output Level Range-14 to -127 dBm
RF Generator Output Level Resolution0.1 dB

RF Analyzer

ParameterValue
RF Analyzer Frequency Range10 MHz to 1 GHz
RF Analyzer Frequency Resolution1 Hz
RF Analyzer Hop Mode Resolution100 kHz
CW RF Frequency Measurement Range10 MHz to 1 GHz
CW RF Frequency Measurement Level Range-6 to +41 dBm
CW RF Power Measurement Range10 MHz to 1 GHz
CW RF Power Measurement Level Range-5 to +41 dBm

Protocol & Call Processing

ParameterValue
Base Station EmulationGSM Base Station Emulator
Bit/Frame Error Rate MeasurementsClass Ia, Ib, and Class II bits in raw and residual form
MS Power Output Level Control0 to 15 with RF analyzer auto adjust
Broadcast Channel CapabilityBCCH + CCCH or BCCH + CCCH + SDCCH/4
Control ChannelsBCCH + CCCH, BCCH + CCCH + SDCCH/4, SDCCH/8 (nonhopped), SACCH/FACCH
Call Control CapabilitiesBS originated call (FS/EFS), MS originated call (FS/EFS), MS camp-on, BS call disconnect, MS call disconnect
Traffic ChannelsTCH (FS/EFS)-HSCSD
Timing ModesAuto, manual, uplink-downlink and offset measurement
RF Hopping CapabilitiesTwo independent, user-definable MA tables with offsets
Speech Encoding/DecodingFull rate speech
Speech Echo ModeUser selectable delay 0 to 5 seconds
Measurement CoordinationFlexible control of burst type, ARFCN and timeslot
SACCH Measurement ResultsRXLEV, RXQUAL and timing advance
GSM Phase CompatibilityGSM Phase II compatible (including fast burst BER and Enhanced Full-rate Speech)
High Speed Data SupportHigh Speed Circuit Switched Data (HSCSD)

Audio & Baseband Generator/Analyzer

ParameterValue
Audio Source Frequency RangeDC to 25 kHz
Audio Analyzer Frequency Measurement Range20 Hz to 400 kHz
Oscilloscope Frequency Range (-3 dB)2 Hz to 50 kHz

Features & Functionality

ParameterValue
Easy To Use ArchitectureOne-button call set-up from power on
Dual-Band Assignment CapabilityE-GSM and Dual-band assignment capability (requires external Agilent 83220A/E for DCS1800/PCS1900)
RF Attenuator TechnologyElectronic attenuators for increased reliability
Integrated Instruments ToolkitIncludes audio analyzer
Automation Software SupportAgilent 83212D software provides full automatic testing

Options

ParameterValue
High-Stability Timebase OptionAvailable
Transit Protection OptionProvides front panel cover and extended rear feet
Test SIM Card OptionAvailable
Micro Test SIM Card OptionAvailable
Mobile Station Test Software OptionAvailable
Rack Mount Flange Kit OptionAvailable

Reference Oscillator

ParameterValue
Standard Reference Oscillator Frequency10 MHz
Standard Reference Oscillator Aging Rate< 2 ppm/year
Standard Reference Oscillator Temperature Stability< 1 ppm (0 °C to +55 °C)

Audio Generator

ParameterValue
Audio Generator Output Level Range0.1 mV to 4 Vrms
Audio Generator Distortion< 0.12% (at 1 kHz, 1 Vrms)

Audio Analyzer

ParameterValue
Audio Distortion Measurement Frequency1 kHz
Audio Distortion Measurement Input Level Range100 mVrms to 30 Vrms
Audio Distortion Measurement Range0.1% to 100%
Audio SINAD Measurement Frequency1 kHz
Audio SINAD Measurement Range0 to 46 dB
Audio Voltmeter Frequency Range20 Hz to 100 kHz
DC Voltmeter Measurement Range±42 V

RF Generator

ParameterValue
RF Generator SSB Phase Noise<-114 dBc/Hz (at 20 kHz offset, 900 MHz)
GMSK Modulation FilterBT = 0.3
Pulse Modulation On/Off Ratio> 80 dB
Pulse Modulation Rise/Fall Time< 1.5 µs

Transmitter Measurements

ParameterValue
Transmitter Burst Power Measurement Accuracy±0.6 dB
Transmitter Burst Power Measurement Range-15 dBm to +43 dBm
Phase Error RMS Measurement Accuracy< 1.5° RMS
Phase Error Peak Measurement Accuracy< 4.0° peak
Frequency Error Measurement Accuracy±(15 Hz + timebase error)

Inputs & Outputs

ParameterValue
RF IN/OUT Port SWR< 1.25:1

Technology

GSM

Applications

GSM Mobile Station Testing & Repair