The Agilent Keysight 8991A Peak Power Analyzer provides complete and accurate characterization of complex pulsed RF and microwave signals. It features two dedicated sensor channels and two oscilloscope/video channels, allowing for the simultaneous measurement of modulating signals and detector power envelopes.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488.2)
DisplayMonochrome CRT

Physical

ParameterValue
Power Supply115 Vac (90 to 127 Vac) / 230 Vac (195 to 253 Vac), 48 - 66 Hz (or 360 - 440 Hz at 115 Vac), 250 VA max
Operating Temperature0 to +55 °C
Dimensions (W×H×D)425 mm × 194 mm × 366 mm
Weight11 kg

Frequency and Power Measurement Specifications

ParameterValue
Frequency Range50 MHz to 40 GHz (sensor dependent)
Power Measurement Range-32 to +20 dBm (sensor dependent)
Dynamic Range52 dB (sensor dependent)
Minimum Pulse Width20 ns
Rise Time<5 ns (with 84812A/13A/14A sensors)

Sensor Compatibility and Inputs

ParameterValue
Number of Channels4 (2 sensor channels, 2 oscilloscope/video channels)
Sensor Input ChannelsCh 1 and Ch 4
Oscilloscope/Video ChannelsCh 2 and Ch 3
Compatible Sensor ModelsAgilent 84812A, 84813A, 84814A, 84815A
Internal Calibrator Frequency1.00 GHz
Internal Calibrator Power Level10.0 mW (+10 dBm)
Internal Calibrator ConnectorType-N (female) on front panel
Calibrator Power Uncertainty±1.5% (at 10 mW)
Video Channel Bandwidth100 MHz (Channels 2 and 3)
Video Channel Input Impedance1 MOhm (nominal)

Display and Automated Measurements

ParameterValue
Automatic Timing MeasurementsRise time, fall time, pulse width, PRI, PRF, duty cycle, delay
Automatic Power MeasurementsPeak power, average power, pulse top amplitude, pulse base amplitude, overshoot
Marker FunctionsAmplitude and Time markers

Triggering and Timebase

ParameterValue
Timebase Range2 ns/div to 5 s/div
Trigger ModesAuto, Normal, Single
Trigger SourcesChannel 1, 2, 3, or 4

Measurement Accuracy and Resolution

ParameterValue
Vertical Resolution8 bits (12 bits with averaging)

Environmental and Safety

ParameterValue
Storage Temperature Range-40 to +75 °C

Timebase

ParameterValue
Timebase Accuracy±0.005%
Timebase Resolution100 ps

Video Channels

ParameterValue
Video Channel Vertical Sensitivity Range2 mV/div to 5 V/div

Power Requirements

ParameterValue
Line Voltage115 Vac (90 to 127 Vac) / 230 Vac (195 to 253 Vac)
Line Frequency48 to 66 Hz (at 115/230 Vac) or 360 to 440 Hz (at 115 Vac)
Power Consumption250 VA maximum

Environmental

ParameterValue
Operating HumidityUp to 95% RH at +40 °C (non-condensing)

Options

ParameterValue
Option 001Deletes channel 3 and channel 4 (reduces instrument to 1 sensor channel and 1 video channel)
Option 002Rear panel inputs (sensor/video inputs and calibrator output on the rear panel)
Option 908Rackmount kit without handles
Option 913Rackmount kit with handles
Option 003Rear panel sensor inputs

System & Control

ParameterValue
HP-IB Interface CapabilitiesSH1, AH1, T5, TE0, L3, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2

Inputs & Outputs

ParameterValue
Auxiliary Trigger ConnectorBNC (rear panel)
Auxiliary Trigger Input ImpedanceSelectable 1 M̤ (approx. 30 pF) or 50 Ω
Auxiliary Trigger Maximum Input Voltage±40 V (dc + peak ac) @ 1 M̤, 5 V rms @ 50 Ω
Recorder Output ConnectorBNC (rear panel)
Recorder Output Voltage Range0 to 1 V

Internal Calibrator

ParameterValue
Internal Calibrator SWR1.05 typical, 1.10 maximum

Technology

RF/Microwave Peak Power Analysis

Applications

Characterization of pulsed RF signals, radar testing, and peak power measurements