The Keysight Technologies (formerly Agilent, HP) E3600 Series 30-60W DC Bench Power Supplies are a family of compact, single-output instruments providing solid performance for a variety of laboratory and production test applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2) standard; RS-232 (some models) |
| Display | Dual 4-digit LED displays |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100/120/220/240 VAC (selectable), 47-63 Hz (47-440 Hz for some models) |
| Operating Temperature | 0 to 40 °C |
| Dimensions (W×H×D) | 212.3 W × 132.6 H × 348.3 D mm |
| Weight | 6.8 to 7.7 kg |
General Specifications
| Parameter | Value |
|---|---|
| Output Channels | 1 |
| Output Power | 30 to 60 W (model dependent) |
| Input Type | AC |
| Output Type | DC |
Output Ratings (Per Channel)
| Parameter | Value |
|---|---|
| Output Voltage Range | 0 to 120 V (model dependent) |
| Output Current Range | 0 to 3 A (model dependent) |
| Maximum Output Power | 30 W (E3610A, E3612A), 35 W (E3611A), 40 W (E3614A), 52.5 W (E3616A), 60 W (E3615A, E3617A) |
| Voltage Programming Resolution | 10 mV |
| Current Programming Resolution | 1 mA |
Programming Accuracy (25°C ±5°C)
| Parameter | Value |
|---|---|
| Voltage Programming Accuracy | 0.05% + 10 mV (model dependent) |
| Current Programming Accuracy | 0.2% + 10 mA (model dependent) |
Ripple & Noise (20 Hz to 20 MHz)
| Parameter | Value |
|---|---|
| Voltage RMS Ripple | < 0.5 mVrms |
| Voltage Peak-to-Peak Noise | < 10 mVp-p |
| Current RMS Noise | < 1 mArms |
Load and Line Regulation
| Parameter | Value |
|---|---|
| Voltage Load Regulation | < 0.01% + 2 mV |
| Current Load Regulation | < 0.01% + 2 mA |
| Voltage Line Regulation | < 0.01% + 2 mV |
| Current Line Regulation | < 0.01% + 2 mA |
Transient Response
| Parameter | Value |
|---|---|
| Voltage Transient Response Time | < 50 µs |
Measurement Accuracy (25°C ±5°C)
| Parameter | Value |
|---|---|
| Voltage Measurement Accuracy | 0.05% + 10 mV |
| Current Measurement Accuracy | 0.2% + 10 mA |
| Voltage Measurement Resolution | 10 mV |
| Current Measurement Resolution | 10 mA |
Output Characteristics
| Parameter | Value |
|---|---|
| Output Isolation Voltage | 240 VDC (output to chassis/ground) |
| Remote Sensing Capability | Yes (E3614A-E3617A) |
| Output Stability | 0.01% + 2 mV (over 8 hours) |
| Primary Operating Modes | Constant Voltage (CV), Constant Current (CC) |
| Remote Sense Lead Current | <1 mA |
Protection Features
| Parameter | Value |
|---|---|
| Over Voltage Protection (OVP) | Yes (E3614A-E3617A) |
| Over Current Protection (OCP) | Yes (Constant Current mode) |
| Over Temperature Protection (OTP) | Yes |
| Reverse Polarity Protection | Yes |
Remote Control & Interfaces
| Parameter | Value |
|---|---|
| Standard Interfaces | GPIB (IEEE-488.2) |
| Optional Interfaces | RS-232 (available on some models as option or standard) |
| Programming Language | SCPI (Standard Commands for Programmable Instruments) |
| Driver Support | IVI-COM, LabVIEW |
Front Panel Operation
| Parameter | Value |
|---|---|
| Controls | Keypad, Rotary knob |
| Output Connectors | Safety binding posts |
Environmental Conditions
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 to 70 °C |
| Operating Humidity | Up to 80% RH (non-condensing) to 30°C, 50% RH at 40°C |
| Operating Altitude | Up to 2000 m |
| Pollution Degree | Pollution Degree 2 |
Safety & Compliance
| Parameter | Value |
|---|---|
| Safety Standards | EN 61010-1 (IEC 61010-1) |
| EMC Standards | EN 61326-1 (IEC 61326-1) |
| Certifications | CE, CSA (or equivalent UL/cUL) |
| RoHS Compliance | Yes |
Physical Specifications
| Parameter | Value |
|---|---|
| Rack Mount Kit Availability | Yes (optional) |
AC Input Requirements
| Parameter | Value |
|---|---|
| Input Voltage Range | 100/120/220/240 VAC, selectable |
| Input Frequency Range | 47-63 Hz |
| Maximum Input Power Consumption | Up to 200 VA (model dependent) |
| Input Fuse Rating | Input fuse ratings vary by model: E361x models are 1.6 A (220/240V) and 2.5 A (100/120V); E3620A is 2.0 A (220/240V) and 3.15 A (100/120V); E3630A/E3631A are 2.5 A (220/240V) and 4.0 A (100/120V). |
Calibration Interval
| Parameter | Value |
|---|---|
| Recommended Calibration Interval | 1 year |
Output Performance
| Parameter | Value |
|---|---|
| Ripple & Noise Measurement Bandwidth | 20 Hz to 20 MHz |
| Transient Response Test Condition | 50% load change |
Programming Characteristics
| Parameter | Value |
|---|---|
| Voltage Programming Slew Rate (Full Scale) | <50 ms (typical) |
| Current Programming Slew Rate (Full Scale) | <50 ms (typical) |
User Interface
| Parameter | Value |
|---|---|
| Voltage Display Digits | 3 digits (LED) |
| Current Display Digits | 3 digits (LED) |
Protection
| Parameter | Value |
|---|---|
| Remote Sense Protection Feature | Automatic shutdown upon disconnect |
Power Input
| Parameter | Value |
|---|---|
| Input Fuse Location | Rear panel accessible |
Environmental
| Parameter | Value |
|---|---|
| Typical Acoustic Noise | <40 dBA (typical) |
General Characteristics
| Parameter | Value |
|---|---|
| Required Warm-up Time | 30 minutes |
Connectivity
| Parameter | Value |
|---|---|
| Common Standard Digital Interface | GPIB (IEEE-488.2, SCPI-compatible) (model option dependent) |
| Common Optional Digital Interfaces | RS-232 (model option dependent) |
Display
| Parameter | Value |
|---|---|
| Display Type | 7-segment LED |
Performance
| Parameter | Value |
|---|---|
| Voltage Temperature Coefficient | <= (0.02% of V_out + 2 mV) / C (typical over 0-55 C) |
| Current Temperature Coefficient | <= (0.02% of I_out + 2 mA) / C (typical over 0-55 C) |
Power
| Parameter | Value |
|---|---|
| Hold-up Time | >10 ms (at full load, min AC input) |
Technology
Linear Regulated Power Supply
Applications
Bench Testing, Education, Production Test, R&D