The Keysight Technologies (formerly Agilent, HP) E3600 Series 30-60W DC Bench Power Supplies are a family of compact, single-output instruments providing solid performance for a variety of laboratory and production test applications.
100/120/220/240 VAC (selectable), 47-63 Hz (47-440 Hz for some models)
Operating Temperature
0 to 40 °C
Dimensions (W×H×D)
212.3 W × 132.6 H × 348.3 D mm
Weight
6.8 to 7.7 kg
General Specifications
Parameter
Value
Output Channels
1
Output Power
30 to 60 W (model dependent)
Input Type
AC
Output Type
DC
Output Ratings (Per Channel)
Parameter
Value
Output Voltage Range
0 to 120 V (model dependent)
Output Current Range
0 to 3 A (model dependent)
Maximum Output Power
30 W (E3610A, E3612A), 35 W (E3611A), 40 W (E3614A), 52.5 W (E3616A), 60 W (E3615A, E3617A)
Voltage Programming Resolution
10 mV
Current Programming Resolution
1 mA
Programming Accuracy (25°C ±5°C)
Parameter
Value
Voltage Programming Accuracy
0.05% + 10 mV (model dependent)
Current Programming Accuracy
0.2% + 10 mA (model dependent)
Ripple & Noise (20 Hz to 20 MHz)
Parameter
Value
Voltage RMS Ripple
< 0.5 mVrms
Voltage Peak-to-Peak Noise
< 10 mVp-p
Current RMS Noise
< 1 mArms
Load and Line Regulation
Parameter
Value
Voltage Load Regulation
< 0.01% + 2 mV
Current Load Regulation
< 0.01% + 2 mA
Voltage Line Regulation
< 0.01% + 2 mV
Current Line Regulation
< 0.01% + 2 mA
Transient Response
Parameter
Value
Voltage Transient Response Time
< 50 µs
Measurement Accuracy (25°C ±5°C)
Parameter
Value
Voltage Measurement Accuracy
0.05% + 10 mV
Current Measurement Accuracy
0.2% + 10 mA
Voltage Measurement Resolution
10 mV
Current Measurement Resolution
10 mA
Output Characteristics
Parameter
Value
Output Isolation Voltage
240 VDC (output to chassis/ground)
Remote Sensing Capability
Yes (E3614A-E3617A)
Output Stability
0.01% + 2 mV (over 8 hours)
Primary Operating Modes
Constant Voltage (CV), Constant Current (CC)
Remote Sense Lead Current
<1 mA
Protection Features
Parameter
Value
Over Voltage Protection (OVP)
Yes (E3614A-E3617A)
Over Current Protection (OCP)
Yes (Constant Current mode)
Over Temperature Protection (OTP)
Yes
Reverse Polarity Protection
Yes
Remote Control & Interfaces
Parameter
Value
Standard Interfaces
GPIB (IEEE-488.2)
Optional Interfaces
RS-232 (available on some models as option or standard)
Programming Language
SCPI (Standard Commands for Programmable Instruments)
Driver Support
IVI-COM, LabVIEW
Front Panel Operation
Parameter
Value
Controls
Keypad, Rotary knob
Output Connectors
Safety binding posts
Environmental Conditions
Parameter
Value
Storage Temperature Range
-40 to 70 °C
Operating Humidity
Up to 80% RH (non-condensing) to 30°C, 50% RH at 40°C
Operating Altitude
Up to 2000 m
Pollution Degree
Pollution Degree 2
Safety & Compliance
Parameter
Value
Safety Standards
EN 61010-1 (IEC 61010-1)
EMC Standards
EN 61326-1 (IEC 61326-1)
Certifications
CE, CSA (or equivalent UL/cUL)
RoHS Compliance
Yes
Physical Specifications
Parameter
Value
Rack Mount Kit Availability
Yes (optional)
AC Input Requirements
Parameter
Value
Input Voltage Range
100/120/220/240 VAC, selectable
Input Frequency Range
47-63 Hz
Maximum Input Power Consumption
Up to 200 VA (model dependent)
Input Fuse Rating
Input fuse ratings vary by model: E361x models are 1.6 A (220/240V) and 2.5 A (100/120V); E3620A is 2.0 A (220/240V) and 3.15 A (100/120V); E3630A/E3631A are 2.5 A (220/240V) and 4.0 A (100/120V).