The Keysight (formerly Agilent/HP) E3631A is an 80 W triple-output programmable DC power supply widely used in benchtop and automated test applications. It features one 6 V/5 A output and two ±25 V/1 A outputs, offering clean power with low ripple and noise. The unit includes built-in GPIB and RS-232 interfaces for remote programming and monitoring, alongside precise readback capabilities.

Specifications

General

ParameterValue
InterfacesGPIB, RS-232
DisplayVacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply115 Vac ± 10% (standard), 100 Vac or 230 Vac ± 10% (by option); 47-63 Hz; 250 VA maximum
Operating Temperature0 to 40 °C
Dimensions (W×H×D)212.6 × 132.6 × 348.3 mm
Weight8.2 kg
CoolingFan cooled

Output Ratings

ParameterValue
Number of Outputs3
Maximum Total Output Power80 W
Output 1 Voltage Range0 to 6 V
Output 1 Current Range0 to 5 A
Output 2 Voltage Range0 to +25 V
Output 2 Current Range0 to 1 A
Output 3 Voltage Range0 to -25 V
Output 3 Current Range0 to 1 A

Programming Accuracy

ParameterValue
Voltage Programming Accuracy0.05% + 20 mV (±25 V outputs), 0.1% + 5 mV (6 V output)
Current Programming Accuracy0.15% + 4 mA (±25 V outputs), 0.2% + 10 mA (6 V output)

Readback Accuracy

ParameterValue
Voltage Readback Accuracy0.05% + 10 mV (±25 V outputs), 0.1% + 5 mV (6 V output) @ 25°C ±5°C
Current Readback Accuracy0.15% + 4 mA (±25 V outputs), 0.2% + 10 mA (6 V output) @ 25°C ±5°C

Regulation

ParameterValue
Voltage Load Regulation< 0.01% + 2 mV
Current Load Regulation< 0.01% + 250 µA
Voltage Line Regulation< 0.01% + 2 mV
Current Line Regulation< 0.01% + 250 µA

Ripple and Noise

ParameterValue
Normal Mode Voltage RMS< 350 µV rms (20 Hz to 20 MHz)
Normal Mode Voltage Peak-to-Peak< 2 mV p-p (20 Hz to 20 MHz)
Normal Mode Current RMS< 500 µA rms (±25 V outputs), < 2 mA rms (6 V output)
Common Mode Current RMS< 1.5 µA rms

Performance

ParameterValue
Transient Response Time< 50 µs to recover within 15 mV for a 50% change in load current
Output Tracking Accuracy± (0.2% of output + 20 mV) for ±25V outputs
Warm-Up Time1 hour (for specified accuracy)
Remote SensingNot supported (local sensing only)

Safety & Protection

ParameterValue
Isolation to Ground±240 Vdc maximum (between any output terminal and earth ground)

Environmental

ParameterValue
Storage Temperature-40 °C to 70 °C
Maximum Altitude2000 meters
Pollution Degree2
Installation CategoryII

Power

ParameterValue
AC Input Voltage115 Vac ±10% (standard); 100 Vac or 230 Vac ±10% (by option)
AC Input Frequency47 Hz to 63 Hz
Maximum Input Power250 VA

Configuration

ParameterValue
Output TopologyAll three outputs share a common COM terminal

Software & Control

ParameterValue
State Storage Memory3 user-configurable operating states (Store/Recall)
Programming LanguageSCPI

Technology

Programmable DC Power Supply

Applications

Benchtop and Automated Test Equipment (ATE)