The Keysight Agilent E3634A is a high-performance, single-output, dual-range programmable DC power supply offering up to 200 W of clean power. Designed for both benchtop and automated test environments, it features built-in GPIB and RS-232 interfaces, low output ripple and noise, high programming and readback accuracy, and excellent regulation capabilities.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB, RS-232 |
| Display | Vacuum Fluorescent Display (VFD) |
| State Storage | 3 user-defined states |
| Warm-up Time | 1 hour |
| Recommended Calibration Cycle | 1 year |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 Vac / 115 Vac / 230 Vac (±10%), 47 Hz to 63 Hz, 500 VA max |
| Operating Temperature | 0 to 40 °C |
| Dimensions (W×H×D) | 213 mm × 133 mm × 348 mm |
| Weight | 9.5 kg |
Output Ratings
| Parameter | Value |
|---|---|
| Number of Outputs | 1 |
| Output Ranges | Range 1: 0 to 25 V, 7 A / Range 2: 0 to 50 V, 4 A |
| Range 1 Output Voltage | 0 to 25 V |
| Range 1 Output Current | 0 to 7 A |
| Range 2 Output Voltage | 0 to 50 V |
| Range 2 Output Current | 0 to 4 A |
| Maximum Output Power | 200 W |
| Isolation Voltage | 240 Vdc |
Load Regulation
| Parameter | Value |
|---|---|
| Voltage Load Regulation | < 0.01% + 2 mV |
| Current Load Regulation | < 0.01% + 250 µA |
Line Regulation
| Parameter | Value |
|---|---|
| Voltage Line Regulation | < 0.01% + 2 mV |
| Current Line Regulation | < 0.01% + 250 µA |
Ripple & Noise
| Parameter | Value |
|---|---|
| Normal Mode Voltage Ripple (RMS) | < 500 µVrms |
| Normal Mode Voltage Ripple (Peak-to-Peak) | < 3 mVpp |
| Normal Mode Current Ripple (RMS) | < 2 mA rms |
| Common Mode Current Noise | < 1.5 µA rms |
Programming Accuracy
| Parameter | Value |
|---|---|
| Voltage Programming Accuracy | 0.05% + 10 mV |
| Current Programming Accuracy | 0.2% + 10 mA |
Readback Accuracy
| Parameter | Value |
|---|---|
| Voltage Readback Accuracy | 0.05% + 5 mV |
| Current Readback Accuracy | 0.15% + 5 mA |
Resolution
| Parameter | Value |
|---|---|
| Voltage Programming Resolution | 1.5 mV |
| Current Programming Resolution | 0.2 mA |
| Voltage Readback Resolution | 1.5 mV |
| Current Readback Resolution | 0.2 mA |
| Front Panel Meter Voltage Resolution | 10 mV |
| Front Panel Meter Current Resolution | 1 mA |
Transient Response
| Parameter | Value |
|---|---|
| Transient Recovery Time | < 50 µs |
| Voltage Settling Band | within 15 mV |
| Load Change Specification | 50% to 100% or 100% to 50% load change |
Stability & Temperature Coefficient
| Parameter | Value |
|---|---|
| Voltage Temperature Coefficient | < 0.01% + 3 mV per °C |
| Current Temperature Coefficient | < 0.02% + 3 mA per °C |
Connectivity & System Interfaces
| Parameter | Value |
|---|---|
| Programming Languages Supported | SCPI (Standard Commands for Programmable Instruments) |
Environmental Specifications
| Parameter | Value |
|---|---|
| Cooling Method | Fan cooled |
Power Requirements
| Parameter | Value |
|---|---|
| Maximum Input Apparent Power | 700 VA |
| AC Input Frequency | 47 Hz - 63 Hz |
Environmental
| Parameter | Value |
|---|---|
| Operating Altitude | Up to 2,000 meters |
Compliance
| Parameter | Value |
|---|---|
| Pollution Degree | Pollution Degree 2 |
| Installation Category | Category II |
Software & Programming
| Parameter | Value |
|---|---|
| SCPI Version | 1995.0 |
Technology
Linear
Applications
Benchtop R&D, automated production testing, and education