The Keysight (Agilent) E3649A is a high-performance, dual-output, 97W programmable DC power supply designed for both benchtop use and system integration. Part of the popular E3640 series, it provides clean, low-ripple power with dual output ranges up to 35V/1.4A or 60V/0.8A per channel. Complete with GPIB and RS-232 interfaces, it features Standard Commands for Programmable Instruments (SCPI) support, excellent regulation, and dual-line vacuum fluorescent displays, making it ideal for general-purpose electronic test and R&D environments.

Specifications

General

ParameterValue
InterfacesGPIB, RS-232
DisplayVacuum fluorescent display (VFD), dual line

Output Specifications

ParameterValue
Number of Outputs2
Output Voltage Range (Range 1)0 to 35 V
Output Current Range (Range 1)0 to 1.4 A
Output Voltage Range (Range 2)0 to 60 V
Output Current Range (Range 2)0 to 0.8 A
Maximum Power Rating97 W
Load Regulation (Voltage)< 0.01% + 3 mV
Load Regulation (Current)< 0.01% + 250 µA
Line Regulation (Voltage)< 0.01% + 3 mV
Line Regulation (Current)< 0.01% + 250 µA
Normal Mode Voltage Ripple & Noise (rms)< 1 mVrms
Normal Mode Voltage Ripple & Noise (peak-to-peak)< 8 mVpp
Normal Mode Current Ripple & Noise (rms)< 4 mArms
Common Mode Noise Current< 1.5 µArms

Programming Accuracy

ParameterValue
Voltage Programming Accuracy (Output 1)< 0.05% + 10 mV
Voltage Programming Accuracy (Output 2)< 0.05% + 10 mV
Current Programming Accuracy (Outputs 1 & 2)< 0.2% + 10 mA

Readback Accuracy

ParameterValue
Voltage Readback Accuracy (Output 1)< 0.05% + 5 mV
Voltage Readback Accuracy (Output 2)< 0.05% + 5 mV
Current Readback Accuracy (Output 1)< 0.15% + 5 mA
Current Readback Accuracy (Output 2)< 0.15% + 5 mA

Transient Response and Settling

ParameterValue
Transient Response Recovery Time< 50 µs
Transient Response Settling Bandwithin 15 mV of final value following a 50% to 100% or 100% to 50% load change

Protection Features

ParameterValue
Overvoltage Protection (OVP) Accuracy< 0.5% + 0.5 V

Isolation and Connections

ParameterValue
Maximum Isolation Voltage (Output to Ground)240 Vdc
Remote Sensing CapabilityUp to 1 V drop per lead

System Control and Interface

ParameterValue
Command Language / Syntax StandardsSCPI (Standard Commands for Programmable Instruments)
Save/Recall State Storage Capacity5 states

Environmental Specifications

ParameterValue
Storage Temperature Range-20 to 70 °C
Relative Humidity (Non-condensing)Up to 80% RH at 40°C
Cooling System TypeFan-cooled

Physical Characteristics

ParameterValue
Net Weight9.2 kg
Rack Mounting Unit (RU) Height3U (half-rack width)

Compliance and Standards

ParameterValue
Safety Standards CertificationsCertified to CSA 22.2 No. 1010.1, IEC 1010-1 / EN 61010-1
Electromagnetic Compatibility (EMC) ComplianceEN 55011 Group 1 Class A, EN 50082-1

Power Requirements

ParameterValue
AC Input Frequency47 Hz to 63 Hz

Performance

ParameterValue
Maximum Sense Lead Compensation Voltage1 V per lead
Warm-up Time30 minutes

Environmental

ParameterValue
Operating AltitudeUp to 2000 meters

Compliance & Safety

ParameterValue
Installation CategoryIEC 1010-1 Overvoltage Category II
Pollution DegreePollution Degree 2

Maintenance

ParameterValue
Recommended Calibration Interval1 year

Inputs & Outputs

ParameterValue
Rear Panel ConnectionsScrew-terminal blocks for Output 1 and Output 2 (including remote sense), GPIB (IEEE-488) connector, RS-232 (9-pin female) connector

Interfaces & Control

ParameterValue
Default GPIB Address5
RS-232 Baud Rates300, 600, 1200, 2400, 4800, 9600 baud
RS-232 Flow ControlDTR/DSR (hardware) or None
SCPI Compliance Version1997.0

Technology

Linear power supply

Applications

Test and Measurement, R&D Benchtop, and Automated Test Systems (ATE)