The Keysight (formerly Agilent / HP) E4407B is a 26.5 GHz mid-performance spectrum analyzer belonging to the ESA-E Series. Designed for general-purpose RF and microwave measurements in both laboratory and field environments, it offers a standard frequency range of 9 kHz to 26.5 GHz (extendable down to 100 Hz with optional configuration). Key features include an overall amplitude accuracy of ±0.4 dB, a 10 MHz analysis bandwidth, and a highly sensitive displayed average noise level (DANL) of -167 dBm when utilizing the internal preamplifier.

Specifications

General

ParameterValue
Calibration Cycle1 year
EMC/EMI ComplianceCISPR Pub 11 Group 1 Class A

Physical

ParameterValue
Power Supply90 to 132 V (47-440 Hz) / 195 to 250 V (47-66 Hz); < 300 W
Operating Temperature0 to 55
Dimensions (W×H×D)416 × 222 × 409
Weight17.1

Frequency Specifications

ParameterValue
Frequency Range9 kHz to 26.5 GHz (100 Hz to 26.5 GHz with Option UKB)
Phase Noise @ 10 kHz Offset≤ -90 dBc/Hz (1 GHz carrier)
Sweep Time Range1 ms to 4000 s (Span > 0 Hz), 50 µs to 4000 s (Zero Span)

Amplitude Specifications

ParameterValue
Absolute Amplitude Accuracy±0.4 dB overall
Maximum Safe Continuous CW Power+30 dBm (1 W)
Input Attenuator Range0 to 65 dB in 5 dB steps

Inputs and Outputs

ParameterValue
Front Panel RF Input ConnectorAPC-3.5 mm (male)
RF Input Impedance50 Ω

General and Environmental

ParameterValue
Warm-up Time5 minutes to guaranteed measurement accuracy
Storage Temperature Range-40 to +75 °C

Frequency

ParameterValue
Span Range0 Hz (zero span), 100 Hz to 26.5 GHz
Timebase Aging Rate (Standard)± 2 × 10^-6 / year
Timebase Aging Rate (Option 1D5)± 1 × 10^-7 / year

Inputs & Outputs

ParameterValue
Timebase Reference Input/Output10 MHz

Amplitude

ParameterValue
Amplitude Reference Signal50 MHz, -20 dBm
Reference Level Range-149.9 dBm to +30 dBm
Amplitude UnitsdBm, dBmV, dBµV, V, W
Standard DetectorsNormal, positive peak, sample, negative peak
Log Scale Fidelity± 0.3 dB (0 to -60 dB from reference level)
Residual Responses<-90 dBm (0 dB attenuation, 50 Ω load)

Options & Enhancements

ParameterValue
Tracking Generator Frequency (Option 1DN)9 kHz to 3.0 GHz

Frequency Performance

ParameterValue
Phase Noise @ 100 kHz Offset≤ -113 dBc/Hz (typical ≤ -117 dBc/Hz)
Phase Noise @ 1 MHz Offset≤ -119 dBc/Hz (typical ≤ -127 dBc/Hz)

Sweep Specifications

ParameterValue
Sweep Points401

Options & Configurations

ParameterValue
Time-Gated Spectrum AnalysisAvailable via Option 1D6
DSP and Fast ADCAvailable via Option B7D

Technology

Superheterodyne Spectrum Analyzer

Applications

RF & Microwave test, general purpose spectrum analysis