The Keysight E9322A (formerly Agilent) is a high-performance peak and average power sensor operating from 50 MHz to 6 GHz. Featuring a 1.5 MHz video bandwidth and a wide dynamic range, the E9322A is optimized for time-gated measurements on TDMA and CDMA wireless formats such as Bluetooth and cdmaOne. It contains a built-in EEPROM for storing calibration data and works seamlessly with the EPM-P and P-Series power meters.

Specifications

General

ParameterValue
InterfacesProprietary multi-pin connector for E-series power meters
DisplayNone (uses power meter display)
Warm-up Time30 minutes

Physical

ParameterValue
Power SupplyPowered via sensor cable from connected power meter
Operating Temperature0 to 55 °C
Dimensions (W×H×D)38 mm × 30 mm × 150 mm
Recommended RF Connector Torque1.35 N-m (12 in-lb)

RF Performance

ParameterValue
Frequency Range50 MHz to 6 GHz
Power Measurement Range (Normal Mode)-45 dBm to +20 dBm
Power Measurement Range (Average Only Mode)-60 dBm to +20 dBm
Maximum Continuous Average Power+23 dBm
Maximum Peak Power+30 dBm (< 10 µs duration)
Video Bandwidth (High Setting)1.5 MHz
Video Bandwidth (Medium Setting)300 kHz
Video Bandwidth (Low Setting)100 kHz
Rise Time (High Bandwidth Setting)< 0.4 µs
Fall Time (High Bandwidth Setting)< 0.4 µs
Rise Time (Medium Bandwidth Setting)< 1.5 µs
Fall Time (Medium Bandwidth Setting)< 1.5 µs
Rise Time (Low Bandwidth Setting)< 5.0 µs
Fall Time (Low Bandwidth Setting)< 5.0 µs
Minimum Pulse Width1 µs (with High BW setting)
Measurement SpeedUp to 1000 corrected readings per second (via GPIB)

Impedance & Matching

ParameterValue
Nominal Impedance50 Ω
Maximum VSWR (50 MHz to 2 GHz)1.12
Maximum VSWR (2 GHz to 6 GHz)1.15
RF Connector TypePrecision Type-N (male)

Measurement Accuracy & Calibration

ParameterValue
Internal EEPROM StorageYes (calibration factors, linearity, temperature, and bandwidth correction data)
System Calibration Cycle1 year
Zero Set (Normal Mode, High BW)< ±12 nW
Zero Set (Normal Mode, Medium BW)< ±4 nW
Zero Set (Normal Mode, Low BW)< ±2 nW
Zero Set (Average Only Mode)< ±500 pW
Zero Drift (Normal Mode, High BW)< ±5 nW
Zero Drift (Normal Mode, Medium BW)< ±1.5 nW
Zero Drift (Normal Mode, Low BW)< ±0.8 nW
Zero Drift (Average Only Mode)< ±100 pW
Measurement Noise (Normal Mode, High BW, Free Run)< 12 nW
Measurement Noise (Normal Mode, Medium BW, Free Run)< 4 nW
Measurement Noise (Normal Mode, Low BW, Free Run)< 2.5 nW
Measurement Noise (Average Only Mode)< 330 pW

System Compatibility & Connectivity

ParameterValue
Compatible Power MetersKeysight E4416A, E4417A
Compatible Sensor CablesE9288A (1.5 m), E9288B (3 m), E9288C (10 m)

Environmental & Compliance

ParameterValue
Storage Temperature Range-70 to +85 °C
Relative Humidity Limit95% RH at 40 °C (non-condensing)
Operating Altitude Limit4600 m (15000 ft)
Storage Altitude Limit15000 m (50000 ft)

Calibration & Uncertainty

ParameterValue
Sensor Calibration Factor Uncertainty (50 MHz to 2 GHz)1.8% (at 25 ± 5 °C)
Sensor Calibration Factor Uncertainty (2 GHz to 6 GHz)2.1% (at 25 ± 5 °C)

Power Linearity

ParameterValue
Power Linearity (Normal Mode, -32 to -20 dBm)±4.0% (at 25 ± 5 °C)
Power Linearity (Normal Mode, -20 to -10 dBm)±3.5% (at 25 ± 5 °C)
Power Linearity (Normal Mode, -10 to 0 dBm)±2.5% (at 25 ± 5 °C)
Power Linearity (Normal Mode, 0 to +10 dBm)±2.0% (at 25 ± 5 °C)
Power Linearity (Normal Mode, +10 to +20 dBm)±2.5% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, -60 to -10 dBm)±3.0% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, -10 to 0 dBm)±2.5% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, 0 to +10 dBm)±2.0% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, +10 to +20 dBm)±2.5% (at 25 ± 5 °C)

Performance

ParameterValue
Temperature Coefficient (Normal Mode, -20 to +20 dBm)0.15% per °C (0 to 55 °C)
Temperature Coefficient (Average-Only Mode, -60 to +20 dBm)0.25% per °C (-60 to -10 dBm), 0.15% per °C (-10 to +20 dBm)
Maximum Peak Power Duration<10 µs (for +30 dBm peak power)

Calibration

ParameterValue
EEPROM Calibration PointsAt 50 MHz, 100 MHz intervals (50 MHz to 2 GHz), and 200 MHz intervals (2 GHz to 6 GHz)

Compliance & Safety

ParameterValue
EMC StandardsIEC 61326-1:1997+A1:1998+A2:2000, EN 61326-1:1997+A1:1998+A2:2001, CISPR 11:1997+A1:1999+A2:2002 Class A, AS/NZS CISPR 11:2002, ICES/NMB-001:2004
Safety StandardsIEC 61010-1:2001, EN 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1:2004
Pollution DegreePollution Degree 2

Environmental

ParameterValue
Operating LocationIndoor use only

Technology

Diode Power Sensor

Applications

RF and microwave power measurements, Bluetooth and cdmaOne testing