The Keysight E9322A (formerly Agilent) is a high-performance peak and average power sensor operating from 50 MHz to 6 GHz. Featuring a 1.5 MHz video bandwidth and a wide dynamic range, the E9322A is optimized for time-gated measurements on TDMA and CDMA wireless formats such as Bluetooth and cdmaOne. It contains a built-in EEPROM for storing calibration data and works seamlessly with the EPM-P and P-Series power meters.

Specifications

General

ParameterValue
InterfacesProprietary multi-pin connector for E-series power meters
DisplayNone (uses power meter display)
Warm-up Time30 minutes

Physical

ParameterValue
Power SupplyPowered via sensor cable from connected power meter
Operating Temperature0 to 55 °C
Dimensions (W×H×D)38 mm × 30 mm × 150 mm
Recommended RF Connector Torque1.35 N-m (12 in-lb)

RF Performance

ParameterValue
Frequency Range50 MHz to 6 GHz
Power Measurement Range (Normal Mode)-45 dBm to +20 dBm
Power Measurement Range (Average Only Mode)-60 dBm to +20 dBm
Maximum Continuous Average Power+23 dBm
Maximum Peak Power+30 dBm (< 10 µs duration)
Video Bandwidth (High Setting)1.5 MHz
Video Bandwidth (Medium Setting)300 kHz
Video Bandwidth (Low Setting)100 kHz
Rise Time (High Bandwidth Setting)< 0.4 µs
Fall Time (High Bandwidth Setting)< 0.4 µs
Rise Time (Medium Bandwidth Setting)< 1.5 µs
Fall Time (Medium Bandwidth Setting)< 1.5 µs
Rise Time (Low Bandwidth Setting)< 5.0 µs
Fall Time (Low Bandwidth Setting)< 5.0 µs
Minimum Pulse Width1 µs (with High BW setting)
Measurement SpeedUp to 1000 corrected readings per second (via GPIB)

Impedance & Matching

ParameterValue
Nominal Impedance50 Ω
Maximum VSWR (50 MHz to 2 GHz)1.12
Maximum VSWR (2 GHz to 6 GHz)1.15
RF Connector TypePrecision Type-N (male)

Measurement Accuracy & Calibration

ParameterValue
Internal EEPROM StorageYes (calibration factors, linearity, temperature, and bandwidth correction data)
System Calibration Cycle1 year
Zero Set (Normal Mode, High BW)< ±12 nW
Zero Set (Normal Mode, Medium BW)< ±4 nW
Zero Set (Normal Mode, Low BW)< ±2 nW
Zero Set (Average Only Mode)< ±500 pW
Zero Drift (Normal Mode, High BW)< ±5 nW
Zero Drift (Normal Mode, Medium BW)< ±1.5 nW
Zero Drift (Normal Mode, Low BW)< ±0.8 nW
Zero Drift (Average Only Mode)< ±100 pW
Measurement Noise (Normal Mode, High BW, Free Run)< 12 nW
Measurement Noise (Normal Mode, Medium BW, Free Run)< 4 nW
Measurement Noise (Normal Mode, Low BW, Free Run)< 2.5 nW
Measurement Noise (Average Only Mode)< 330 pW

System Compatibility & Connectivity

ParameterValue
Compatible Power MetersKeysight E4416A, E4417A
Compatible Sensor CablesE9288A (1.5 m), E9288B (3 m), E9288C (10 m)

Environmental & Compliance

ParameterValue
Storage Temperature Range-70 to +85 °C
Relative Humidity Limit95% RH at 40 °C (non-condensing)
Operating Altitude Limit4600 m (15000 ft)
Storage Altitude Limit15000 m (50000 ft)

Calibration & Uncertainty

ParameterValue
Sensor Calibration Factor Uncertainty (50 MHz to 2 GHz)1.8% (at 25 ± 5 °C)
Sensor Calibration Factor Uncertainty (2 GHz to 6 GHz)2.1% (at 25 ± 5 °C)

Power Linearity

ParameterValue
Power Linearity (Normal Mode, -32 to -20 dBm)±4.0% (at 25 ± 5 °C)
Power Linearity (Normal Mode, -20 to -10 dBm)±3.5% (at 25 ± 5 °C)
Power Linearity (Normal Mode, -10 to 0 dBm)±2.5% (at 25 ± 5 °C)
Power Linearity (Normal Mode, 0 to +10 dBm)±2.0% (at 25 ± 5 °C)
Power Linearity (Normal Mode, +10 to +20 dBm)±2.5% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, -60 to -10 dBm)±3.0% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, -10 to 0 dBm)±2.5% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, 0 to +10 dBm)±2.0% (at 25 ± 5 °C)
Power Linearity (Average-Only Mode, +10 to +20 dBm)±2.5% (at 25 ± 5 °C)

Performance

ParameterValue
Temperature Coefficient (Normal Mode, -20 to +20 dBm)0.15% per °C (0 to 55 °C)
Temperature Coefficient (Average-Only Mode, -60 to +20 dBm)0.25% per °C (-60 to -10 dBm), 0.15% per °C (-10 to +20 dBm)
Maximum Peak Power Duration<10 µs (for +30 dBm peak power)

Calibration

ParameterValue
EEPROM Calibration PointsAt 50 MHz, 100 MHz intervals (50 MHz to 2 GHz), and 200 MHz intervals (2 GHz to 6 GHz)

Compliance & Safety

ParameterValue
EMC StandardsIEC 61326-1:1997+A1:1998+A2:2000, EN 61326-1:1997+A1:1998+A2:2001, CISPR 11:1997+A1:1999+A2:2002 Class A, AS/NZS CISPR 11:2002, ICES/NMB-001:2004
Safety StandardsIEC 61010-1:2001, EN 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1:2004
Pollution DegreePollution Degree 2

Environmental

ParameterValue
Operating LocationIndoor use only

Technology

Diode Power Sensor

Applications

RF and microwave power measurements, Bluetooth and cdmaOne testing

Looking for the Keysight Technologies (Agilent HP) E9322A?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.