The Keysight E9325A is a high-performance peak and average diode power sensor designed for use with EPM-P and P-Series power meters. It covers a frequency range of 50 MHz to 18 GHz and features a 300 kHz video bandwidth, making it ideal for time-gated measurements on TDMA wireless systems including GSM, EDGE, and NADC.

Specifications

General

ParameterValue
InterfacesProprietary sensor cable interface to host power meter

Physical

ParameterValue
Power SupplyPowered by host power meter via sensor cable
Operating Temperature0 to 55 °C
Dimensions (W×H×D)150 mm x 38 mm x 30 mm
Weight0.18 kg
Shipping Weight0.55 kg (1.2 lb)

Frequency and Power Specifications

ParameterValue
Frequency Range50 MHz to 18 GHz
Minimum Operating Frequency50 MHz
Maximum Operating Frequency18 GHz
Power Measurement Range (Average Mode)-65 dBm to +20 dBm
Power Measurement Range (Normal Mode)-50 dBm to +20 dBm
Maximum Continuous Power+23 dBm (200 mW) average
Maximum Peak Power+30 dBm (1 W) peak for < 10 µs
Maximum Allowed DC Voltage5 V

RF Impedance and Impedance Matching

ParameterValue
Nominal Impedance50 Ω
SWR (50 MHz to 2 GHz)1.12
SWR (2 GHz to 10 GHz)1.16
SWR (10 GHz to 16 GHz)1.23
SWR (16 GHz to 18 GHz)1.28
RF Connector TypePrecision Type-N (male)

Dynamic and Pulse Characteristics

ParameterValue
Video Bandwidth (High Setting)300 kHz
Video Bandwidth (Medium Setting)100 kHz
Video Bandwidth (Low Setting)30 kHz
Rise/Fall Time (High Bandwidth Setting)< 0.9 µs
Rise/Fall Time (Medium Bandwidth Setting)< 3.2 µs
Rise/Fall Time (Low Bandwidth Setting)< 10 µs
Measurement SpeedUp to 1000 corrected readings per second (via GPIB)

Measurement Accuracy and Uncertainty

ParameterValue
Zero Drift (Normal Mode)1.5 nW
Zero Drift (Average Only Mode)< ±60 pW
Measurement Noise (Normal Mode)< 6 nW (free run acquisition)
Measurement Noise (Average Only Mode)< 165 pW

Calibration and Compatibility

ParameterValue
EEPROM Calibration MemoryStores calibration factors, linearity, temperature, and bandwidth correction
Compatible Power MetersKeysight EPM-P series (E4416A, E4417A) and P-Series (N1911A, N1912A)

Environmental Conditions

ParameterValue
Storage Temperature Range-40 °C to +75 °C
Operating HumidityUp to 95% RH at 40 °C (non-condensing)

Accuracy & Calibration

ParameterValue
Calibration Factor Uncertainty (50 MHz)1.8% (at 25 °C ± 5 °C)
Calibration Factor Uncertainty (0.1 GHz to 2 GHz)1.8% (at 25 °C ± 5 °C)
Calibration Factor Uncertainty (2 GHz to 10 GHz)2.4% (at 25 °C ± 5 °C)
Calibration Factor Uncertainty (10 GHz to 16 GHz)2.4% (at 25 °C ± 5 °C)
Calibration Factor Uncertainty (16 GHz to 18 GHz)2.6% (at 25 °C ± 5 °C)
Linearity (Average Only Mode, -65 to -10 dBm)±3.0% (at 25 °C ± 5 °C)
Linearity (Average Only Mode, -10 to +20 dBm)±4.5% (at 25 °C ± 5 °C)
Linearity (Normal Mode, -50 to -10 dBm, Peak)±4.0% (at 25 °C ± 5 °C)
Linearity (Normal Mode, -10 to +20 dBm, Average)±5.0% (at 25 °C ± 5 °C)
Linearity (Normal Mode, -10 to +20 dBm, Peak)±4.5% (at 25 °C ± 5 °C)

System Connectivity

ParameterValue
Compatible Sensor CablesE9288A (1.5 m), E9288B (3 m), E9288C (10 m)

Power Specifications

ParameterValue
Power Measurement Range (Normal Mode, Low Bandwidth)-50 dBm to +20 dBm
Power Measurement Range (Normal Mode, Medium Bandwidth)-45 dBm to +20 dBm
Power Measurement Range (Normal Mode, High Bandwidth)-40 dBm to +20 dBm
Maximum Peak Power Duration< 1 µs

Accuracy & Noise

ParameterValue
Zero Set (Normal Mode, Low Bandwidth)5.0 nW
Zero Set (Normal Mode, Medium Bandwidth)12.5 nW
Zero Set (Normal Mode, High Bandwidth)25.0 nW
Zero Set (Average Only Mode)0.5 nW

Operation

ParameterValue
Sensor Warm-up Time30 minutes

Environmental

ParameterValue
Maximum Operating Altitude4,600 m (15,000 ft)
Maximum Non-Operating Altitude15,240 m (50,000 ft)
Storage Humidity RangeUp to 90% RH at 65 °C (non-condensing)

Compliance

ParameterValue
EMC ComplianceIEC 61326-1 / EN 61326-1, CISPR 11 / EN 55011 Group 1 Class A
Safety StandardsIEC 61010-1 / EN 61010-1, CSA C22.2 No. 61010-1

Technology

Diode Power Sensor

Applications

RF & Microwave Power Measurement