The Keysight N1922A is a P-Series wideband power sensor designed for high-performance peak, average, and peak-to-average ratio measurements over a broad frequency range of 50 MHz to 40 GHz. Operating in conjunction with the N1911A and N1912A P-Series power meters, the N1922A provides internal zeroing and calibration to streamline test setups in aerospace, defense, and wireless communications applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Proprietary multi-pin sensor cable interface to P-Series power meters |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered directly by compatible P-Series power meter via sensor cable |
| Operating Temperature | 0 to 55 °C |
| Dimensions (W×H×D) | 127 mm × 40 mm × 27 mm |
| Weight | 0.226 kg |
RF Performance
| Parameter | Value |
|---|---|
| Frequency Range | 50 MHz to 40 GHz |
| Power Measurement Range (Normal Mode) | -35 dBm to +20 dBm |
| Power Measurement Range (Average-Only Mode) | -30 dBm to +20 dBm |
| Maximum Continuous Power (Average) | +23 dBm |
| Maximum Peak Power | +30 dBm (< 1 µs) |
| Maximum DC Input Voltage | 5 VDC |
| Video Bandwidth (VBW) | 30 MHz |
| Rise Time | ≤ 13 ns (for power levels ≥ 0 dBm) |
| Fall Time | ≤ 13 ns (for power levels ≥ 0 dBm) |
| Minimum Pulse Width | 50 ns |
| Maximum Sampling Rate | 100 MSamples/s (continuous sampling) |
| VSWR (50 MHz to 10 GHz) | ≤ 1.25 |
| VSWR (10 GHz to 20 GHz) | ≤ 1.30 |
| VSWR (20 GHz to 26.5 GHz) | ≤ 1.50 |
| VSWR (26.5 GHz to 40 GHz) | ≤ 1.70 |
| RF Input Impedance | 50 Ω nominal |
| RF Input Connector | 2.4 mm (male) |
Measurement Accuracy & Uncertainty
| Parameter | Value |
|---|---|
| Power Linearity (-10 dBm to +20 dBm) | ±4.0% |
| Sensor Calibration Type | Internal zeroing and calibration using built-in electronic RF switch |
Measurement Capabilities
| Parameter | Value |
|---|---|
| Measurement Types | Peak, average, peak-to-average ratio (PAR), rise time, fall time, pulse width, pulse period, pulse repetition frequency (PRF), duty cycle |
| Statistical Measurements | CCDF (Complementary Cumulative Distribution Function), PDF (Probability Density Function) |
| Time-Gated Measurements | Up to 4 independent gates |
| Internal Trigger Level Range | -20 dBm to +20 dBm |
| Internal Trigger Level Resolution | 0.1 dB |
System Compatibility
| Parameter | Value |
|---|---|
| Compatible Power Meters | Keysight N1911A (single-channel) and N1912A (dual-channel) P-Series Power Meters |
| Sensor Cable Length Options | 1.5 m (standard), 3.0 m (Option 105), 10.0 m (Option 106) |
Environmental & Standards
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +75 °C |
| Operating Relative Humidity | Up to 95% RH at 40 °C (non-condensing) |
| Storage Relative Humidity | Up to 90% RH at 65 °C |
| Operating Altitude | Up to 3,000 meters (9,840 feet) |
| Storage Altitude | Up to 15,240 meters (50,000 feet) |
| EMC Standards Compliance | Complies with EN 61326-1, ICES-001 |
| Safety Standards Compliance | IEC 61010-1, CSA C22.2 No. 61010-1 |
| Vibration Resistance Rating | MIL-PRF-28800F Class 3 (random vibration) |
| Shock Resistance Rating | MIL-PRF-28800F Class 3 (half-sine shock) |
Power & Calibration Requirements
| Parameter | Value |
|---|---|
| Calibration Cycle | 1 year |
| Onboard Calibration Storage | Internal EEPROM contains calibration factors and 3D linearity correction data |
Performance
| Parameter | Value |
|---|---|
| Zero Set (Normal Mode) | 250 nW |
| Zero Set (Average-Only Mode) | 150 pW |
| Zero Drift (Normal Mode) | 100 nW |
| Zero Drift (Average-Only Mode) | 60 pW |
| Measurement Noise (Normal Mode) | 50 nW |
| Measurement Noise (Average-Only Mode) | 150 pW |
| Video Bandwidth Settings | High (30 MHz), Medium (1.5 MHz), Low (300 kHz) |
| Rise Time (Medium Bandwidth Setting) | 230 ns |
| Rise Time (Low Bandwidth Setting) | 1.2 µs |
Calibration
| Parameter | Value |
|---|---|
| Internal Zeroing Time | < 10 s (typical) |
| Internal Calibration Time | < 10 s (typical) |
Triggering
| Parameter | Value |
|---|---|
| Internal Trigger Latency | 160 ns ± 10 ns |
| Internal Trigger Jitter | 1 ns RMS |
| Trigger Delay Range (Pre-Trigger) | -40 ms to 0 s |
| Trigger Delay Range (Post-Trigger) | 0 s to 1 s |
| Trigger Delay Resolution | 1% of delay setting or 10 ns (whichever is greater) |
Technology
Diode Power Sensor
Applications
RF and Microwave Power Measurement