The Keysight N1923A is a P-Series wideband power sensor operating from 50 MHz to 18 GHz. Featuring an integrated 3.0-meter (10 ft) cable, it is designed for use with Keysight N1911A and N1912A P-Series power meters to perform high-speed peak, average, and time-gated power measurements. The sensor supports internal zeroing and calibration to eliminate the need for an external reference source.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Integrated P-Series sensor cable (3.0 m) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered directly from the host P-Series power meter |
| Operating Temperature | 0 to 55 °C |
| Dimensions (W×H×D) | 40 mm × 27 mm × 135 mm |
| Weight | 0.226 |
| Cable Configuration | Fixed/integrated cable |
Measurement Ranges & Power Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 50 MHz to 18 GHz |
| Power Measurement Range (Average Mode) | -35 dBm to +20 dBm |
| Power Measurement Range (Normal Mode) | -35 dBm to +20 dBm |
| Maximum Continuous Input Power | +23 dBm (200 mW) |
| Maximum Peak Input Power | +30 dBm (1 W) for < 10 µs duration |
| Maximum DC Input Voltage | 5 VDC |
Sensor Technology & Architecture
| Parameter | Value |
|---|---|
| Sensor Detection Type | Diode-based |
| Measurement Modes Supported | Normal (Peak, Average, Time-gated, Rise/Fall time, Pulse width) and Average-only |
| Video Bandwidth | ≥ 30 MHz |
| Rise Time | ≤ 13 ns (for power levels > -15 dBm) |
| Fall Time | ≤ 13 ns (for power levels > -15 dBm) |
| Minimum Pulse Width | 50 ns |
RF Port Characteristics
| Parameter | Value |
|---|---|
| Input Impedance | 50 Ω nominal |
| RF Connector Type | Type-N |
| Connector Sex | Male |
| Maximum SWR (50 MHz to 2 GHz) | 1.12 |
| Maximum SWR (2 GHz to 10 GHz) | 1.16 |
| Maximum SWR (10 GHz to 16 GHz) | 1.21 |
| Maximum SWR (16 GHz to 18 GHz) | 1.29 |
Triggering & Sampling
| Parameter | Value |
|---|---|
| Maximum Sampling Rate | 100 MSamples/s continuous sampling |
| Trigger Delay Range | -1.0 s to +1.0 s |
| Time Resolution | 100 ps |
Meter Compatibility & System Integration
| Parameter | Value |
|---|---|
| Compatible Power Meters | Keysight N1911A (Single Channel) and N1912A (Dual Channel) P-Series Power Meters |
| Calibration Factor Storage Type | Internal EEPROM |
| Internal Calibration Support | Yes (Internal Zero and Calibration, no external reference needed) |
Environmental Parameters
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +70 °C |
| Operating Humidity Range | 95% RH at 40 °C (non-condensing) |
| Storage Humidity Range | 90% RH at 65 °C (non-condensing) |
| Maximum Operating Altitude | 4,600 meters (15,000 feet) |
| Maximum Storage Altitude | 15,000 meters (50,000 feet) |
Physical Dimensions & Weight
| Parameter | Value |
|---|---|
| Cable Length | 3.0 meters (10 feet) |
Accuracy & Calibration
| Parameter | Value |
|---|---|
| Zero Set (Average Mode) | ±0.17 µW |
| Zero Set (Normal Mode) | ±1.4 µW |
| Zero Drift (Average Mode) | ±0.035 µW |
| Zero Drift (Normal Mode) | ±0.15 µW |
Performance
| Parameter | Value |
|---|---|
| Measurement Noise (Average Mode) | ≤ 240 nW |
| Measurement Noise (Normal Mode) | ≤ 1.6 µW |
| Video Bandwidth Settings | High (30 MHz), Medium (15 MHz), Low (5 MHz), Off (300 kHz) |
| System Rise Time | ≤ 13 ns (with N1911A/N1912A) |
| System Fall Time | ≤ 13 ns (with N1911A/N1912A) |
| Trigger Jitter | 1 ns rms |
| Power Linearity (Normal Mode, -32 to -20 dBm) | ±4.5% (Normal Mode, -35 to -20 dBm) |
| Power Linearity (Normal Mode, -20 to +20 dBm) | ±4.0% |
| Power Linearity (Average Mode, -35 to -20 dBm) | ±4.5% |
| Power Linearity (Average Mode, -20 to +20 dBm) | ±4.0% |
Triggering
| Parameter | Value |
|---|---|
| Internal Trigger Level Range | -20 dBm to +20 dBm |
Calibration
| Parameter | Value |
|---|---|
| Internal Zero and Cal Support | Yes, without disconnecting RF input (internal RF switch) |
| Zeroing Time | 11 s (typical) |
| Calibration Time | 19 s (typical) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Connector to Power Meter | P-Series multi-pin sensor connector |
EMC & Safety
| Parameter | Value |
|---|---|
| EMC Compliance | IEC 61326-1 / EN 61326-1, CISPR 11 / EN 55011 Class A |
| Safety Compliance | IEC 61010-1 / EN 61010-1 |
Technology
Diode-based wideband power sensor
Applications
RF and microwave pulse profiling, peak and average power measurements