The Keysight N3301A (formerly Agilent N3301A) is a compact, 2-slot programmable DC electronic load mainframe designed for high-density, high-speed test systems. Operating with up to two single-slot modules or one double-slot module, it provides a flexible and modular load solution with built-in GPIB and RS-232 control, making it ideal for high-volume automated testing of power supplies and power components.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232
DisplayVacuum Fluorescent Display (VFD)
Warm-Up Time30 minutes
Recommended Calibration Cycle1 year
Store/Recall States10 non-volatile states

Physical

ParameterValue
Power Supply100 - 240 VAC, 48 - 63 Hz, 230 VA max
Operating Temperature0 °C to 55 °C
Dimensions (W×H×D)213 mm × 177 mm × 404 mm
Weight7.3

Mainframe Capacity

ParameterValue
Number of Slots2
Maximum Module Capacity2 single-slot modules or 1 double-slot module
Supported Load ModulesN3302A, N3303A, N3304A, N3305A, N3306A, N3307A

Input Ratings and Operating Ranges

ParameterValue
Total Power CapacityUp to 600 W (module dependent)
Maximum Input Power150 W to 600 W (module dependent)
Maximum Input Voltage60 V to 240 V (module dependent)
Maximum Input Current10 A to 120 A (module dependent)

Operating Modes & Accuracy

ParameterValue
Constant Power (CP) Range LowNot supported
Constant Power (CP) Range HighNot supported
Constant Power (CP) Programming AccuracyNot supported

Transient and Dynamic Characteristics

ParameterValue
Transient Mode Frequency Range0.25 Hz to 10 kHz (module dependent)
Transient Mode Duty Cycle Range3% to 97% (module dependent)
List Mode Step Duration Range0.5 ms to 262143 ms
Maximum Number of List Mode Steps100 steps (module dependent)

Measurement and Readback Accuracy

ParameterValue
Power Readback AccuracyNot supported

Protection Features

ParameterValue
Over-Temperature Protection (OTP)Yes, thermal sensor shuts down the overloaded module
Reverse Voltage Protection Maximum CurrentEqual to maximum rated current of module

Connectivity & Control Interfaces

ParameterValue
Analog Programming Input Voltage Range0 V to 10 V (module dependent)
Current Monitor Output Scale Factor0 V to 10 V for 0 to full scale current (module dependent)
External Trigger InputBNC connector, active-low, TTL level
External Trigger OutputBNC connector, active-low, TTL level
Supported Command Sets / EmulationsSCPI (Standard Commands for Programmable Instruments)

Power Input Requirements

ParameterValue
Line Voltage Input Range100 to 240 VAC
Line Frequency Range48 to 63 Hz
Maximum VA Power Consumption230 VA

Physical and Mechanical Specifications

ParameterValue
Form FactorBenchtop or Rackmount (4U half-rack size)
Rack Height Units4U
Net Weight7.3 kg (16 lb) empty mainframe
Shipping Weight9.1 kg (20 lb) empty mainframe
Cooling ConfigurationForced air cooling with internal fans

Environmental and Safety Compliance

ParameterValue
Storage Temperature Range-40 °C to 75 °C
Safety Standards CertificationsCSA C22.2 No. 1010.1, IEC 1010-1, UL 3111-1
Electromagnetic Compatibility (EMC) StandardsConforms to EN 55011 (Group 1, Class A), EN 50082-1

Environmental

ParameterValue
Operating AltitudeUp to 2,000 m
Storage AltitudeUp to 15,240 m
Operating HumidityUp to 95% RH (non-condensing)
Pollution Degree2
Overvoltage CategoryII
Acoustic Noise Level55 dBA (sound pressure level at 1 meter, typical)

Connectivity

ParameterValue
RS-232 Baud Rates1200, 2400, 4800, 9600, 19200, 38400 bps
GPIB Interface StandardIEEE-488.2
GPIB Address Range0 to 30
Digital I/O Connector14-pin quick-disconnect terminal block

Power

ParameterValue
AC Input Fuse (100-120 VAC)3.15 A, 250 V, Slow-Blow
AC Input Fuse (220-240 VAC)2.0 A, 250 V, Slow-Blow

Software & Control

ParameterValue
Command LanguageSCPI