Specifications
Physical
| Parameter | Value |
|---|
| Operating Temperature | 0 °C to 55 °C |
Output Ratings
| Parameter | Value |
|---|
| Maximum Output Power | 50 W (N6761A); 100 W (N6762A) |
| Output Voltage Range | 0 to 50 V |
| Output Current Range | 0 to 1.5 A (N6761A); 0 to 3 A (N6762A) |
| Voltage Low Range | 5.5 V |
| Current Low Range | 100 mA |
Programming Accuracy
| Parameter | Value |
|---|
| Voltage Programming Accuracy | 0.016% + 6 mV |
| Current Programming Accuracy | 0.04% + 200 µA |
| Voltage Low Range Accuracy | 0.016% + 1.5 mV |
| Current Low Range Accuracy | 0.04% + 15 µA |
Measurement/Readback Accuracy
| Parameter | Value |
|---|
| Voltage Readback Accuracy | 0.016% + 6 mV |
| Current Readback Accuracy | 0.04% + 160 µA |
| Voltage Low Range Readback Accuracy | 0.016% + 1.5 mV |
| Current Low Range Readback Accuracy | 0.03% + 15 µA |
| Microamp Range Measurement Accuracy | 0.5% + 100 nA (with Option 2UA) |
Ripple and Noise
| Parameter | Value |
|---|
| Voltage Peak-to-Peak Ripple (20 Hz to 20 MHz) | 4.5 mV |
| Voltage RMS Ripple | 0.35 mV |
| Current RMS Ripple | 2 mA |
Dynamic Characteristics
| Parameter | Value |
|---|
| Output Turn-on Delay Range | 0 to 3600 s |
| Output Turn-off Delay Range | 0 to 3600 s |
Output Protections
| Parameter | Value |
|---|
| Overvoltage Protection (OVP) Range | 0 to 55 V |
Features and Options
| Parameter | Value |
|---|
| Current Measurement Low Range Option | Option 2UA (200 µA range) |
| Output Disconnect Relays Option | Option 761 |
| Polarity Reversal Relays Option | Option 760 (includes disconnect relays) |
| Galvanic Isolation | Yes (Output to chassis and between outputs) |
Physical and Environmental
| Parameter | Value |
|---|
| Module Slot Width | 1 slot |
| Mainframe Compatibility | Keysight N6700, N6701, N6702, N6705 Series |
Technology
Precision Programmable DC Power Supply
Applications
Automated Test Equipment (ATE), low-power device testing, battery emulation, simultaneous V/I digitizing