The Keysight (formerly Agilent) N9030A is the first-generation PXA signal analyzer, representing the highest-performance member of the X-Series family. Designed to replace high-performance legacy analyzers like the PSA, it delivers exceptional phase noise, dynamic range, and measurement sensitivity for advanced RF and microwave test applications. The N9030A is highly configurable, with frequency options up to 50 GHz and analysis bandwidths up to 160 MHz, and supports Real-Time Spectrum Analysis (RTSA) and a wide range of measurement applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0, 1000Base-T LAN, GPIB |
| Display | 8.4 inch (213 mm) diagonal XGA color LCD |
| Warm-up Time | 5 minutes |
| Calibration Cycle | 1 year |
| Display Size | 8.4 inch (213 mm) diagonal |
| Display Resolution | 1024 x 768 pixels (XGA) |
| Operating System | Windows Embedded Standard 7 (Windows XP on older units) |
| Data Storage | Removable solid-state drive (SSD) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 120 V / 220 to 240 V, 50/60 Hz; 100 to 120 V, 400 Hz |
| Operating Temperature | 0 to 55 |
| Dimensions (W×H×D) | 426 x 177 x 556 |
| Weight | 22 kg |
Frequency Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 3 Hz to 3.6, 8.4, 13.6, 26.5, 43, 44, or 50 GHz (configuration dependent) |
| Internal Reference Frequency | 10 MHz |
| Resolution Bandwidth (RBW) Range | 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz |
| Analysis Bandwidth (Standard) | 10 MHz |
Phase Noise Specifications
| Parameter | Value |
|---|---|
| Phase Noise at 10 kHz Offset | -129 dBc/Hz (1 GHz center frequency) |
Amplitude and Dynamic Range
| Parameter | Value |
|---|---|
| Maximum Safe Input Level (Average Total Power) | +30 dBm (1 W) |
| Input Attenuator Range | 0 to 70 dB |
| Input Attenuator Step Size | 2 dB steps |
Sweep Specifications
| Parameter | Value |
|---|---|
| Sweep Time Range (Span > 0 Hz) | 1 µs to 6000 s |
| Sweep Time Range (Zero Span) | 1 µs to 6000 s |
Front Panel Inputs and Outputs
| Parameter | Value |
|---|---|
| RF Input Impedance | 50 Ω |
| USB Ports (Front Panel) | USB 2.0 (Type-A) |
Rear Panel Inputs and Outputs
| Parameter | Value |
|---|---|
| 10 MHz Reference Input | BNC female |
| 10 MHz Reference Output | BNC female |
Connectivity and Interfaces
| Parameter | Value |
|---|---|
| Gigabit LAN Interface | 1000Base-T (RJ-45) |
| GPIB Interface | IEEE-488 bus connector |
| LXI Class Compliance | LXI Class C |
Software and Measurement Capabilities
| Parameter | Value |
|---|---|
| Real-Time Spectrum Analysis (RTSA) Support | Supported (Option RT1/RT2) |
| Vector Signal Analysis (VSA) Software Compatibility | Supported (89600 VSA Software) |
Environmental and Regulatory
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to 70 °C |
Measurement & Sweep
| Parameter | Value |
|---|---|
| Detector Types | Normal, Peak, Sample, Negative Peak, Log power average, RMS average, Voltage average |
| Trace Functions | Clear/write, max hold, min hold, view, blank, trace average |
| Trigger Sources | Free run, line, video, external 1, external 2, RF burst, periodic timer |
Performance
| Parameter | Value |
|---|---|
| Absolute Amplitude Accuracy | ±0.19 dB (20 Hz to 3.6 GHz, 95th percentile) |
| Video Bandwidth (VBW) Range | 1 Hz to 3 MHz (10% steps), plus 4, 5, 6, 8 MHz |
| Noise Floor Extension (NFE) | Standard feature (characterizes and removes analyzer noise) |
| Residual Responses | -137 dBm (nominal, 200 kHz to 8.4 GHz) |
Power
| Parameter | Value |
|---|---|
| Power Consumption (Maximum) | 695 W (fully loaded) |
| Power Consumption (Standby) | 70 W |
Compliance
| Parameter | Value |
|---|---|
| EMC Compliance | CISPR Pub 11 Group 1, class A; AS/NZS CISPR 11; ICES/NMB-001 |
| Safety Compliance | IEC 61010-1, CSA C22.2 No. 61010-1, UL 61010-1 |
Options & Configuration
| Parameter | Value |
|---|---|
| Preamplifier Options | 3.6 GHz, 8.4 GHz, 13.6 GHz, 26.5 GHz, 43 GHz, 44 GHz, or 50 GHz (dependent on maximum frequency option) |
Frequency
| Parameter | Value |
|---|---|
| Available Frequency Options | 3.6 / 8.4 / 13.6 / 26.5 / 43 / 44 / 50 GHz (Options 503, 508, 513, 526, 543, 544, 550) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Probe Power Connector | +15 VDC, -12.6 VDC at 150 mA maximum (front panel) |
Technology
Signal/Spectrum Analysis
Applications
Aerospace, Defense, Commercial Communications, and General Purpose RF/Microwave Test