The Keysight (Agilent) Q8486A is a high-performance thermocouple waveguide power sensor designed for precise average power measurements in the Q-band frequency range from 33 GHz to 50 GHz. Operating over a power level of -30 dBm to +20 dBm, it offers excellent SWR to minimize measurement uncertainty due to mismatch. The sensor features a 50 MHz calibration port with NIST-traceable calibration and is compatible with a wide array of legacy and modern Keysight/Agilent power meters.

Specifications

General

ParameterValue
DisplayNone (uses host power meter display)

Physical

ParameterValue
Power SupplyPowered by host power meter
Operating Temperature0 to 55 °C
Dimensions (W×H×D)44 mm × 39 mm × 79 mm
Weight0.16 kg

Frequency and Power Performance

ParameterValue
Frequency Range33 GHz to 50 GHz
Power Measurement Range-30 dBm to +20 dBm (1 µW to 100 mW)
Maximum SWR1.50
Maximum Average Power200 mW (+23 dBm)
Maximum Peak Power15 W peak

Sensor Element and Dynamic Characteristics

ParameterValue
Sensing Element TypeThermocouple

RF and Waveguide Interface Specifications

ParameterValue
EIA Waveguide SizeWR-22
Flange TypeUG-383/U
Frequency BandQ-band

Measurement Accuracy and Uncertainty

ParameterValue
Calibration Port Reference Frequency50 MHz
Calibration TraceabilityNIST (National Institute of Standards and Technology)

Connectivity and Compatibility

ParameterValue
Compatible Power Meter ModelsEPM Series (E4418A/B, E4419A/B), EPM-P Series (E4416A, E4417A), P-Series (N1911A, N1912A), E1416A VXI, 70100A, 435B, 436A, 437B, 438A
Sensor Cable Interface Type12-pin multi-contact connector (8480 Series Cable)

Environmental Specifications

ParameterValue
Storage Temperature Range-40 °C to +75 °C

Inputs & Outputs

ParameterValue
Calibration Port ConnectorAPC-3.5 (male)

Environmental

ParameterValue
Operating Humidity< 95% RH at 40 °C (non-condensing)
Storage Humidity< 95% RH at 65 °C
Maximum Operating Altitude4570 m
Maximum Storage Altitude15240 m

Waveguide Specifications

ParameterValue
IEC Waveguide DesignationR400

Performance

ParameterValue
Calibration Factor RSS Uncertainty (33.0 to 40.0 GHz)±3.1%
Calibration Factor RSS Uncertainty (40.0 to 50.0 GHz)±3.7%
Calibration Factor Worst-Case Uncertainty (33.0 to 40.0 GHz)±5.8%
Calibration Factor Worst-Case Uncertainty (40.0 to 50.0 GHz)±7.0%
Calibration Port SWR< 1.10 @ 50 MHz
Maximum Energy per Pulse30 W·μs
Zero Set±50 nW (with E4418B / E4419B / 437B)
Zero Drift< ±10 nW (within 1 hour at constant temperature)
Measurement Noise< 110 nW (within 1 minute, with E4418B / 437B)
Calibration Factor Frequency Step Size1 GHz intervals

Technology

Thermocouple

Applications

RF and Microwave Power Measurement