The Keysight U2000A is a high-performance USB average power sensor designed to operate over a wide frequency range of 10 MHz to 18 GHz. By eliminating the need for a separate power meter mainframe, it provides a compact, portable, and cost-effective solution for RF and microwave power measurements directly via a PC or compatible Keysight instrument.

Specifications

General

ParameterValue
InterfacesUSB 2.0 (via proprietary 5-pin to USB-A cable)
DisplayNone (operates via PC host software or compatible instrument)
Warm-up Time30 minutes
Pollution Degree2

Physical

ParameterValue
Operating Temperature0 to 55 °C
Dimensions (W×H×D)163.7 mm × 46.0 mm × 35.9 mm
Weight0.262

RF Performance and Ranges

ParameterValue
Frequency Range10 MHz to 18 GHz
Power Range-60 dBm to +20 dBm
Maximum Average Power+25 dBm (300 mW)
Maximum Peak Power+33 dBm (2 W) for < 10 µs
Maximum DC Voltage5 VDC
Impedance50 Ω
RF Connector TypeN-type (male)

Connectivity and Interfaces

ParameterValue
Sensor Connector TypeProprietary 5-pin
USB ClassUSBTMC (USB Test and Measurement Class)

Software and Programmability

ParameterValue
Command LanguageSCPI (Standard Commands for Programmable Instruments)
Compatible Software ApplicationsKeysight BenchVue (Power Meter App), N1918A Power Analysis Manager
Driver SupportIVI-COM, IVI-C, LabVIEW

Physical and Environmental

ParameterValue
Storage Temperature Range-40 to +70 °C
Operating Humidity RangeUp to 95% RH at 40 °C (non-condensing)

Compliance and Standards

ParameterValue
EMC Compliance StandardsIEC 61326-1 / EN 61326-1
Safety StandardsIEC 61010-1 / EN 61010-1
Recommended Calibration Cycle1 year

Ordering and Configuration Options

ParameterValue
Optional Cable (U2031A)1.5 m (5 ft)
Optional Cable (U2031B)3.0 m (10 ft)
Optional Cable (U2031C)5.0 m (16.4 ft)

Environmental

ParameterValue
Operating AltitudeUp to 3,000 m (9,840 ft)
Non-operating AltitudeUp to 15,240 m (50,000 ft)

Performance

ParameterValue
Calibration TypeInternal calibration (no external source required)
Dynamic Range80 dB
Detector TypeDiode

Electrical

ParameterValue
Maximum Damage Level (Average)+25 dBm

Calibration

ParameterValue
Internal CalibrationCalibrator-free (built-in internal zeroing and calibration)

Technology

Diode Power Sensor

Applications

RF and Microwave Average Power Measurement